栓塞后颅内DAVF的MRI随访:3D TOF与增强T1评估
结果显示3D TOF检测残余瘘的阴性预测值为93%、准确率90%,增强序列评估静脉窦通畅性与DSA一致性良好。
要点
- 3D TOF阴性预测值高,适合排除有临床意义的残余瘘。
- 微小残余分流或低流量瘘可能在3D TOF上漏诊。
- 增强3D T1 GRE评估静脉窦通畅性与DSA一致性良好。
核心结论
- 3D TOF对排除有临床意义的残余DAVF具有较高诊断效能,阴性预测值达93%。
- 微小残余微分流或低流量瘘仍可能被3D TOF漏诊,部分病例需DSA确认。
研究设计与样本
- 回顾性单中心研究,纳入2004至2023年间连续收治的栓塞后颅内DAVF患者。
关键影像表现/征象
- 3D TOF像上可观察到残余瘘的高信号血管结构,用于判断治疗是否完全。
- 存在残余瘘时,3D TOF还用于判断是否伴有皮质静脉引流。
- 增强3D T1 GRE像上可见静脉窦内对比剂充盈情况,用于评估瘘口或近端静脉窦通畅性。
诊断标准与关键数值
- 3D TOF检测残余瘘的阳性预测值为79%(95%CI 54%-94%)。
- 3D TOF检测残余瘘的阴性预测值为93%(95%CI 84%-98%)。
- 3D TOF检测残余瘘的诊断准确率为90%(95%CI 81%-95%)。
- 3D TOF残余瘘评估的观察者间一致性κ=0.71(95%CI 0.52-0.90),提示一致性良好。
临床应用与阅片要点
- 3D TOF呈阴性时,对排除有临床意义的残余瘘可靠性较高,可减少不必要的DSA。
- 3D TOF呈阳性或可疑时,报告应提示需结合DSA进一步确认残余瘘及皮质静脉引流。
- 增强3D T1 GRE可用于无创评估静脉窦通畅性,阅片时应注意与DSA的对应关系。
局限性
- 回顾性单中心研究设计本身存在选择偏倚与验证偏倚风险。
- 原文指出微小残余微分流或低流量瘘在3D TOF上可能保持隐匿,需DSA确认。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Although digital subtraction angiography (DSA) remains the reference standard for post-treatment follow-up of intracranial dural arteriovenous fistulas (DAVFs), it is an invasive procedure associated with procedural risks and may not be justified for the follow-up of all DAVFs.
PURPOSE: This study aimed to evaluate the diagnostic performance of 3D time-of-flight (3D TOF) and contrast-enhanced 3D T1-weighted gradient-echo (3D T1 Gd) magnetic resonance imaging (MRI) sequences for monitoring embolized intracranial DAVFs.
MATERIALS AND METHODS: In this retrospective single-center study, consecutive patients with embolized DAVFs between 2004 and 2023, who underwent MRI and arteriography during follow-up, were evaluated. 3D TOF sequence images were independently reviewed by two readers to assess treatment completeness. In cases of residual fistula, cortical venous drainage (CVD) was evaluated. Patency of the fistulous or proximal venous sinus was assessed on contrast-enhanced 3D T1-weighted GRE imaging. The median follow-up duration was 3.6 months (interquartile range [IQR], 3.0-8.4 months). Imaging findings were systematically compared with DSA as the reference standard.
RESULTS: Seventy-six patients (mean age, 56 ± 12,8 years; 46 men) with 78 embolized DAVFs were included. The positive predictive value, negative predictive value, and diagnostic accuracy of 3D TOF sequence for the detection of residual fistulas were 79% [54% - 94%], 93% [84% - 98%], 90% [81% - 95%], respectively. Inter-rater agreement was good (κ = 0.71; [0.52-0.90]). Intermodality agreement was good for the detection of residual fistulas (κ = 0.72; [0.54-0.90]) and moderate for the detection of cortical venous drainage (κ = 0.54; [0.22-0.85]). Additionally, agreement between 3D T1 Gd and DSA for assessing venous sinus patency was good (κ = 0.80; [0.58-1.00]).
CONCLUSION: 3D TOF and contrast-enhanced 3D T1-weighted GRE imaging are effective non-invasive tools for the follow-up of embolized DAVFs. 3D TOF provides high diagnostic performance for excluding clinically significant residual fistulas, whereas small residual microshunts or low-flow fistulas may remain occult and still require DSA confirmation in selected cases.
引用
- Dji'ala, Vallée, Civelli, et al. MRI follow-up of embolized intracranial dural arteriovenous fistulas using 3D time-of-flight and contrast-enhanced 3D T1-weighted GRE imaging. Neuroradiology. 2026. PMID: 42635758. doi:10.1007/s00234-026-04160-z
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。