儿科1.5-T MRI新型柔性毯式躯干线圈初步评价

Evaluation of a novel flexible, lightweight blanket-like torso coil in pediatric MRI at 1.5-T - initial experience.
PubMed 文献 · Pediatr Radiol · 2026-08-22

本前瞻性研究纳入20例7–18岁接受门诊腹部MRI的儿科患者,在同一1.5-T系统上先后使用常规相控阵躯干线圈和新型柔性毯式躯干线圈成像。5名儿科放射科医生盲法评价图像质量,患者和技术人员填写问卷;结果显示两种线圈图像质量相当,柔性线圈在轴位DWI上肝脏SNR更高(校正P=0.01),患者和技术人员更偏好柔性线圈。

要点

  • 柔性毯式线圈腹部MRI图像质量与常规线圈相当,日常使用不必担心SNR损耗
  • 轴位DWI上柔性线圈肝脏SNR高于常规线圈(校正P=0.01),可放心用于弥散序列
  • 线圈相关伪影少见,全部序列评估中仅2%出现,不影响诊断

核心结论

  • 柔性毯式躯干线圈提供的图像质量与常规线圈相似,且普遍被患者和MRI技术人员偏好

研究设计与样本

  • 前瞻性机构审查委员会批准研究,20例7–18岁门诊腹部MRI儿科患者,同一1.5-T系统先后使用常规躯干线圈和柔性毯式躯干线圈成像
  • 5名儿科放射科医生独立评价配对图像集的主观信噪比和线圈相关伪影;客观图像质量采用器官和组织SNR估计及CNR测量
  • 常规临床成像后交换线圈重复选定序列;患者和MRI技术人员完成舒适度、易用性和检查体验问卷

关键影像表现/征象

  • 放射科医生评价新型线圈主观SNR在所有三条研究序列上与常规线圈相似
  • 线圈相关伪影少见,全部评估中仅2%出现(所有评估序列汇总)
  • 客观分析显示多数序列图像SNR无显著差异,柔性线圈在轴位DWI上肝脏SNR更高
  • 两种线圈之间CNR无差异

诊断标准与关键数值

  • 柔性线圈轴位DWI肝脏SNR高于常规线圈,校正P=0.01
  • 患者舒适度和焦虑程度:55%患者未来MRI检查倾向选择柔性线圈,20%倾向选择常规线圈
  • 技术人员总体偏好:70%表达对柔性线圈的总体偏好
  • 线圈相关伪影发生率:全部序列评估中为2%

临床应用与阅片要点

  • 柔性毯式躯干线圈可用于儿科腹部MRI,图像质量不劣于常规线圈,尤其适合体型较小或焦虑患儿
  • DWI序列使用柔性线圈时肝脏SNR可能更高,有利于病灶检出
  • 报告中无需因线圈更换而特别注明图像质量差异,常规诊断信度不受影响

局限性

  • 单中心样本量仅20例,年龄范围7–18岁
  • 研究为初步经验,未报告长期使用耐久性或不同体型亚组差异

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: Conventional phased-array torso coils used for abdominal MRI are relatively heavy and inflexible, potentially affecting patient comfort and technologist workflow.

OBJECTIVE: To compare image quality, patient experience, and technologist experience using a recently United States Food and Drug Administration-cleared flexible, lightweight, blanket-like torso coil versus a conventional torso coil for pediatric abdominal MRI.

MATERIALS AND METHODS: In this prospective institutional review board-approved study, 20 pediatric patients (7 years-old to 18 years-old) undergoing routine clinical outpatient abdominal MRI were imaged on the same 1.5-T system using both a conventional torso coil and a flexible blanket-like torso coil (Smart Fit TorsoCardiac; Philips Healthcare). After standard clinical imaging, selected sequences were repeated following coil exchange. Five pediatric radiologists independently evaluated paired image sets for subjective signal-to-noise ratio (SNR) and coil-related artifacts. Objective image quality was assessed using organ- and tissue-based SNR estimates and contrast-to-noise ratio (CNR) measurements. Patients and MRI technologists completed questionnaires regarding comfort, usability, and examination experience.

RESULTS: All 20 patients (median age, 15 years; 55% female) underwent imaging with both coils. Radiologists rated the novel coil as similar to the conventional coil for subjective SNR across all three sequences under study; coil-related artifacts were uncommon (2% of assessments for all sequences evaluated). Objective analysis showed no significant image-based SNR differences for most sequences, although the flexible coil demonstrated higher liver SNR on axial diffusion-weighted imaging (adjusted P=0.01). CNR did not differ between coils. Patients reported greater comfort and less anxiety with the flexible coil; 55% preferred it for future MRI examinations versus 20% for the conventional coil. Technologists also favored the flexible coil, with 70% expressing an overall preference.

CONCLUSION: The flexible blanket-like torso coil provided image quality similar to a conventional coil and was generally preferred by patients and MRI technologists.

来源:Pediatr Radiol,查看原文。版权归原作者所有。

引用

  1. Dillman, Tkach, Guillerman, et al. Evaluation of a novel flexible, lightweight blanket-like torso coil in pediatric MRI at 1.5-T - initial experience. Pediatr Radiol. 2026. PMID: 42631733. doi:10.1007/s00247-026-06761-5

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。