儿童可疑异物摄入的X线摄影:准直与曝光策略的剂量优化
结果显示腹部投照剂量最高,而胸片参数联合限制性准直可显著降低辐射剂量,提示在保证异物检出的前提下可通过优化准直与曝光策略减少儿童辐射暴露。
要点
- 颈部至中上腹投照联合胸片低剂量参数即可满足多数异物定位需求
- 腹部投照剂量明显更高,报告时应关注其曝光参数是否必要
- 硬币、纽扣电池、磁铁是最常见异物类型,胃内检出率最高
核心结论
- 准直范围和曝光参数对儿童可疑异物摄入的辐射剂量有显著影响,低剂量胸片参数联合限制性准直可使辐射暴露最小化。
- 腹部投照辐射剂量最高,主要归因于其所采用的腹部曝光参数而非准直范围本身。
- 标准颈部至中上腹投照与扩展至骨盆的准直在剂量面积乘积上无显著差异,但后者有效剂量更高。
研究设计与样本
- 患儿中位年龄2.8岁(四分位距1.5–5.5岁),男性112例,女性71例。
- 按投照范围将X线片分为颈部至中上腹组、扩展准直含骨盆组和仅腹部组三类。
- 颈部至中上腹组和扩展准直含骨盆组均采用胸片曝光参数,仅腹部组采用腹部曝光参数。
- 记录指标包括患者人口学信息、异物类型与位置、投照方案、剂量面积乘积和估算有效剂量。
关键影像表现/征象
- 55.7%的X线片检出异物,其中59.4%位于胃内。
- 硬币最常见(31.7%),其次为纽扣电池(13.8%)和磁铁(12.2%)。
- 仅腹部投照组的剂量面积乘积显著高于其他两组(P<0.001)。
诊断标准与关键数值
- 标准颈部至中上腹投照与扩展至骨盆的准直在剂量面积乘积上无显著差异。
- 扩展准直含骨盆组的有效剂量显著高于标准颈部至中上腹组(P<0.05)。
- 仅腹部投照组的有效剂量最高,中位数为0.053 mSv(0.034–0.086 mSv),与其他组比较P<0.001。
- 患者年龄和重复投照率在不同投照方案间无显著差异。
临床应用与阅片要点
- 对可疑异物摄入患儿可优先采用颈部至中上腹投照,联合胸片低剂量曝光参数以降低辐射剂量。
- 应避免不必要地将投照范围扩展至骨盆,因其增加有效剂量而并未提高剂量面积乘积的利用效率。
- 腹部投照仅应在明确需要时使用,并注意其曝光参数带来的较高辐射负担。
局限性
- 单中心回顾性设计,投照方案的选择可能存在适应证偏倚。
- 未比较不同投照方案间的异物检出敏感度或诊断准确性差异。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Foreign body ingestion constitutes a common emergency in children. Radiography is essential for confirming presence, type, and location of ingested objects. International guidelines recommend broad anatomical coverage from the neck to the pelvis; however, the impact of radiographic exposure parameters and collimation on radiation dose remains underexplored.
OBJECTIVE: To evaluate the effects of different scan protocols on radiation dose in pediatric foreign body ingestion.
MATERIALS AND METHODS: Two-hundred-twenty-one radiographs from 183 pediatric patients (median age 2.8 years [interquartile range 1.5-5.5 years]; 112 males, 71 females) performed between September 2019 and April 2025 were retrospectively analyzed. Radiographs were categorized as neck-upper abdomen, extended collimation including pelvis, or abdomen only. Notably, the first two protocols used chest radiograph exposure parameters. Data included patient demographics, foreign body type and location, radiographic protocol, dose-area product, and estimated effective dose.
RESULTS: A foreign body was detected in 55.7% of radiographs, most of them in the stomach (59.4%). Coins were most common (31.7%), followed by button batteries (13.8%) and magnets (12.2%). Dose-area products were comparable between standard neck-to-upper-abdomen and extended neck-to-pelvis collimation but were markedly higher for abdomen-only imaging (P<0.001). Effective dose increased with extended collimation (P<0.05) and was highest for abdomen-only radiographs (0.053 mSv [0.034-0.086 mSv], P<0.001). No differences were observed in patient age or repeat imaging rates.
CONCLUSION: In suspected pediatric foreign body ingestion, collimation and exposure parameters substantially influence radiation dose. Low-dose chest parameters combined with restrictive collimation minimize radiation exposure, whereas abdominal radiographs carry higher dose, attributable predominantly to the abdominal exposure parameters employed.
引用
- Sturm, Benoit, Grunz, et al. Radiographic imaging in pediatric suspected foreign body ingestion: potential for dose reduction through collimation and exposure strategies. Pediatr Radiol. 2026. PMID: 42622820. doi:10.1007/s00247-026-06751-7
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。