文献速递腹部技术与安全CT

光子计数CT采集参数对虚拟单能衰减值与有效原子序数稳定性的体模研究

Impact of acquisition parameters in photon-counting computed tomography on the stability of virtual monoenergetic attenuation values and effective atomic number: a phantom study.
PubMed 文献 · Eur J Radiol · 2026-08-05

本研究采用成年男性腹部仿真体模,评估管电压、图像质量等级和螺距对光子计数CT虚拟单能图像CT值及有效原子序数稳定性的影响。结果显示螺距对VMI CT值稳定性影响最大,管电压和图像质量等级影响相对较小;中高能级VMI(≥70 keV)在不同参数下相对稳定,推荐作为定量分析首选。Zeff测量在各采集条件下均保持高度稳定。

要点

  • 螺距改变对VMI CT值稳定性影响最大,40 keV图像上L1椎体偏差可达-41 HU,定量分析时需注意。
  • 中高能级VMI(≥70 keV)在各参数下相对稳定,推荐作为定量分析首选。
  • 有效原子序数在各采集条件下高度稳定(ICC≥0.999),适合定量应用。

核心结论

  • 螺距对VMI CT值稳定性影响最大,管电压和图像质量等级影响相对较小。
  • 中高能级VMI(≥70 keV)在不同参数下表现出相对稳定性,推荐作为定量分析的首选。
  • Zeff测量在各种采集条件下均保持高度稳定,支持其在PCCT定量应用中的可靠性。

研究设计与样本

  • 实验类型为体模研究,使用成年男性拟人化腹部体模(Kyoto Kagaku CTU-41)进行光子计数CT扫描。
  • 参考协议为120 kVp、图像质量等级(IQ level)145、螺距1。
  • 通过独立改变管电压(140 kVp、90 kVp、Sn100 kVp)、图像质量等级(115、85、55、25)和螺距(3)获得八组附加协议。
  • VMI重建于40、60、70和100 keV四个能级,并在多个腹部器官中测量CT值和有效原子序数。

关键影像表现/征象

  • 所有协议中VMI衰减值与参考协议呈极好相关(r≥0.996)。
  • 140 kVp协议下所有绝对偏差低于5 HU;90 kVp协议下40 keV前列腺偏差为-11.33 HU,另有五处测量值在5至10 HU之间。
  • Sn100 kVp协议下60 keV图像上L1椎体和前列腺绝对偏差超过10 HU,L1达到19.33 HU。
  • 螺距改变导致变异性最大,仅75%的测量绝对偏差在10 HU以内,67.5%在5 HU以内,40 keV图像上偏差显著,L1椎体达-41 HU。
  • 有效原子序数在140 kVp和IQ等级115、85、55下偏差均在0.2以内;IQ等级25时胰腺偏差为0.34,螺距3时L1最大偏差为0.37。

诊断标准与关键数值

  • VMI衰减值与参考协议的相关系数:所有协议r≥0.996。
  • IQ等级115、85、55下所有绝对CT值偏差均低于5 HU;IQ等级25时仅两处测量值位于5至10 HU范围。
  • 140 kVp协议:所有绝对偏差低于5 HU。
  • 90 kVp协议:40 keV前列腺偏差为-11.33 HU,另有五处测量值介于5至10 HU。
  • Sn100 kVp协议:60 keV时L1椎体和前列腺绝对偏差超过10 HU(L1达19.33 HU),另有三处测量值介于5至10 HU。

临床应用与阅片要点

  • 日常定量分析中优先使用中高能级VMI(≥70 keV),其CT值在不同采集参数下更稳定。
  • 当扫描协议涉及螺距改变时,应注意低能级VMI(如40 keV)CT值可能产生较大偏差,避免用于定量比较。
  • Zeff测量受管电压、图像质量等级和螺距影响较小,可在不同采集条件下可靠使用。
  • 低管电压协议(90 kVp、Sn100 kVp)在低能级VMI上可致部分器官CT值偏差超过5 HU,报告定量结果时需考虑采集条件。

局限性

  • 本研究为单中心体模实验,结果在人体组织中的适用性尚需临床验证。
  • 研究仅使用一种成年男性腹部体模,未覆盖不同体型和性别的影响。
  • 未评估其他可能影响定量稳定性的采集参数,如扫描视野、重建核等。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: Photon-counting CT (PCCT) can generate virtual monoenergetic images (VMI) and effective atomic number (Zeff), demonstrating considerable potential for tissue quantification. However, the influence of key acquisition parameters on the stability of these quantitative metrics has not been fully elucidated. This study aimed to evaluate the reliability of VMI CT numbers and Zeff in abdominal organs under varying tube voltage, image quality (IQ) level, and pitch.

METHODS: An adult male anthropomorphic abdominal phantom (Kyoto Kagaku CTU-41) was scanned using PCCT. The reference protocol was 120 kVp, image quality level (IQ level) 145, and pitch 1. Eight additional protocols were acquired by independently altering tube voltage (140 kVp, 90 kVp, Sn100 kVp), IQ level (115, 85, 55, 25), and pitch (3). VMI were reconstructed at 40, 60, 70, and 100 keV, and CT numbers and Zeff were measured in multiple organs. Differences from the reference protocol were calculated to assess the reliability of these quantitative indices under modified scan conditions.

RESULTS: Across all protocols, VMI attenuation showed excellent correlation with the reference protocol (r ≥ 0.996). For IQ levels of 115, 85, and 55, all absolute CT number deviations were below 5 HU; at IQ level 25, only two measurements fell within the 5-10 HU range. At 140 kVp, all absolute deviations were below 5 HU. For the 90 kVp protocol, the prostate deviation at 40 keV was -11.33 HU, and five other measurements ranged between 5 and 10 HU. With the Sn100 kVp protocol, absolute deviations of the L1 vertebra and prostate exceeded 10 HU at 60 keV (L1 reached 19.33 HU), and three additional measurements were between 5 and 10 HU. Pitch alteration caused the greatest variability: only 75% of measurements had absolute deviations within 10 HU and 67.5% within 5 HU; deviations were prominent on 40 keV images, with the L1 vertebra reaching -41 HU. The intraclass correlation coefficient for Zeff was ≥0.999 under all conditions. Zeff deviations were within 0.2 for 140 kVp and IQ levels of 115, 85, and 55; the pancreas deviation was 0.34 at IQ level 25; at pitch 3, the maximum deviation (L1) was 0.37, and several organs showed deviations close to or exceeding 0.2.

CONCLUSIONS: Pitch exerts the greatest influence on the stability of VMI CT numbers, whereas tube voltage and IQ level have relatively minor effects. VMI at medium-to-high energy levels (≥70 keV) exhibit relative stability across different parameters and are recommended as the first choice for quantitative analysis. Zeff measurements remain highly stable under various acquisition conditions, supporting their reliability in quantitative PCCT applications.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Feng, Yang, Yang, et al. Impact of acquisition parameters in photon-counting computed tomography on the stability of virtual monoenergetic attenuation values and effective atomic number: a phantom study. Eur J Radiol. 2026. PMID: 42580079. doi:10.1016/j.ejrad.2026.113145

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。