超声主导多模态影像机器人穿刺的体模与动物实验研究

Ultrasound-Led Multimodality Imaging-Guided Robotic Needle Insertion With Respiratory Phase Consistency in Phantom and Animal Models.
PubMed 文献 · Ultrasound Med Biol · 2026-08-12

本研究通过静态体模、呼吸体模、新西兰白兔肝脏及巴马小型猪肝脏实验,评估电磁导航超声主导多模态影像机器人系统TONGMAI的穿刺准确性。结果显示有经验医师组机器人辅助与徒手、导针架辅助的US-Error无显著差异;无经验医师组机器人辅助在体模及猪肝实验中优于徒手穿刺。探索性非肝靶器官穿刺首次尝试成功率因靶器官而异,无操作相关并发症。

要点

  • 机器人辅助对有经验医师提升不明显,但可缩小经验差异
  • 无经验医师采用机器人辅助可显著降低超声下穿刺误差
  • 呼吸运动一致的认知融合策略可用于变形胸腹部靶器官穿刺

核心结论

  • TONGMAI机器人在体模与动物条件下可维持有经验医师穿刺表现稳定,并提高无经验医师穿刺准确性
  • 结果支持超声主导多模态机器人工作流用于可变形胸腹部靶器官的进一步开发

研究设计与样本

  • 研究类型为体模与动物实验的穿刺准确性评价
  • 穿刺对象包含静态体模、呼吸体模、新西兰白兔肝脏、广西巴马小型猪肝脏
  • 操作者包含3名有经验医师与6名无经验医师,对比机器人辅助、徒手穿刺及导针架辅助三种方式
  • 主要结局指标为实时超声下穿刺准确性US-Error

关键影像表现/征象

  • 有经验医师组静态体模机器人辅助平均US-Error为0.35 mm
  • 有经验医师组呼吸体模机器人辅助平均US-Error为0.57 mm
  • 有经验医师组兔肝机器人辅助平均US-Error为1.28 mm,小型猪肝为1.32 mm
  • 探索性非肝靶器官穿刺涵盖脾、肾、胰腺、胆囊、膀胱及肺,无操作相关并发症

诊断标准与关键数值

  • 有经验医师组US-Error在静态体模、呼吸体模、兔肝及小型猪肝中三种方式间总体p值分别为0.115、0.131、0.124和0.397
  • 无经验医师组机器人辅助在两种体模条件下优于徒手及导针架辅助
  • 无经验医师组机器人辅助在兔肝与小型猪肝实验中较徒手穿刺显著降低US-Error,p值分别为0.003和0.002
  • 机器人辅助下无经验与有经验医师US-Error在各场景均无显著差异,p值分别为0.110、0.054、0.818和0.548

临床应用与阅片要点

  • 机器人辅助可帮助经验不足的操作者获得与有经验者相近的实时超声下穿刺精度
  • 呼吸位相一致性的认知融合策略适用于肝脏等随呼吸运动的可变形靶器官
  • 经皮穿刺非肝实质器官时首次尝试成功率受靶器官类型影响,需结合目标解剖位置规划入路

局限性

  • 本研究为体模与动物实验,结果尚不能直接外推至临床人体应用
  • 研究未报告样本量与中心数
  • 样本量、随机化及盲法方案在摘要中未予描述

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To evaluate the puncture accuracy and feasibility of TONGMAI, an electromagnetic-tracking-based, ultrasound (US)-led multimodality imaging-guided robotic system using consistency-aware respiratory-guided cognitive fusion, in phantom and animal models.

METHODS: Three experienced and six inexperienced physicians performed needle insertion in static phantom, respiratory phantom, New Zealand white rabbit liver and Guangxi Bama miniature-pig liver experiments using robot assistance, freehand puncture or needle-guide assistance. The primary outcome was real-time US-based puncture accuracy (US-Error). Exploratory robot-assisted punctures of non-hepatic targets (spleen, kidney, pancreas, gallbladder, bladder and lung) in miniature pigs were evaluated for first-attempt success, needle redirection, US-Error when applicable and complications.

RESULTS: Among experienced physicians, US-Error did not differ significantly among robot assistance, freehand puncture and needle-guide assistance in static phantom, respiratory phantom, rabbit liver or miniature-pig liver experiments (overall p = 0.115, 0.131, 0.124 and 0.397, respectively). With robot assistance, mean US-Error values were 0.35, 0.57, 1.28 and 1.32 mm, respectively. Among inexperienced physicians, robot assistance reduced US-Error versus freehand puncture or needle-guide assistance in both phantom settings and versus freehand puncture in rabbit and miniature-pig liver experiments (p = 0.003 and p = 0.002, respectively). Under robot assistance, US-Error did not differ significantly between experienced and inexperienced physicians across all settings (p = 0.110, 0.054, 0.818 and 0.548, respectively). In exploratory non-hepatic punctures, first-attempt success varied by target, with no procedure-related complications.

CONCLUSION: TONGMAI supported consistent puncture performance in experienced physicians and improved accuracy in inexperienced physicians under phantom and animal conditions, supporting further development of a US-led multimodality robotic workflow for deformable thoracoabdominal targets.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。

引用

  1. Xiao, Ding, Pang, et al. Ultrasound-Led Multimodality Imaging-Guided Robotic Needle Insertion With Respiratory Phase Consistency in Phantom and Animal Models. Ultrasound Med Biol. 2026. PMID: 42580952. doi:10.1016/j.ultrasmedbio.2026.07.016

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。