低管电压联合深度学习重建在头颈CTA中的应用

Application value of combining low tube voltage and deep learning image reconstruction algorithm in head and neck CT angiography with reduced radiation and contrast doses.
PubMed 文献 · Eur J Radiol · 2026-07-10

前瞻性随机对照研究纳入100例拟行头颈CTA的血管病患者,低剂量组采用80 kVp、1.2gI/s对比剂及DLIR重建,常规组采用100 kVp、1.6gI/s及ASIR-V50%。结果显示80 kVp联合DLIR-H可使辐射剂量降低36%、对比剂用量降低26%、注射速率降低28%,同时图像质量更好,与DSA评估狭窄的一致性更佳。

要点

  • 80 kVp联合DLIR-H可在头颈CTA中显著降低辐射及对比剂剂量,同时改善图像质量。
  • DLIR-H重建的SNR、CNR高于常规方案,背景噪声最低,血管显示更清晰。
  • 以DSA为金标准,DLIR-H对>50%狭窄的定量评估相关性最高。

核心结论

  • 与100 kVp联合ASIR-V50%的标准方案相比,80 kVp联合DLIR-H使头颈CTA辐射剂量降低36%、对比剂用量降低26%、注射速率降低28%,同时图像质量显著改善,与DSA诊断一致性更优。

研究设计与样本

  • 前瞻性随机对照研究,100例因血管病拟行头颈CTA患者按1:1随机分为低剂量组(80 kVp)和常规组(100 kVp)。
  • 低剂量组对比剂速率为1.2gI/s持续10s,常规组为1.6gI/s持续10s,两组均采用自动管电流调制。
  • 低剂量组分别用DLIR-H、DLIR-M和ASIR-V50%重建,常规组用ASIR-V50%重建。
  • 以DSA作为评估血管狭窄程度的参考标准。

关键影像表现/征象

  • 低剂量组重建后血管CT值与常规组相近,四组背景噪声存在显著差异,A1组(DLIR-H)噪声最低。
  • A1组全部血管的SNR和CNR均显著高于常规组。
  • A1组主观图像质量评分最高,其次为A2组、常规组和A3组,两名阅片者一致性良好(κ=0.88-1)。

诊断标准与关键数值

  • 对比剂用量、注射速率和辐射剂量在低剂量组较常规组分别降低25.61%、28.38%和36.11%(均P<0.001)。
  • 对23个狭窄>50%的节段行DSA对照分析,A1组与DSA在量化血管狭窄方面相关性最强(Pearson r=0.97),MSE最低(14.83)。
  • A1组全部血管的SNR和CNR均显著高于常规组(均P<0.05),四组背景噪声差异有统计学意义(均P<0.05)。

临床应用与阅片要点

  • 低管电压80 kVp联合DLIR-H可作为头颈CTA降低辐射和对比剂剂量的优化扫描方案。
  • 阅片时DLIR-H重建图像噪声更低、血管锐利度更好,有利于提高图像诊断信心。
  • 对血管狭窄程度评估要求较高时,DLIR-H重建结果与DSA一致性更优。

局限性

  • 原文仅涉及单中心前瞻性研究,样本量为100例。
  • 本研究结论基于特定设备与重建算法组合,其他平台的可推广性需进一步验证。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To demonstrate the value of combining low tube voltage (80kVp) and deep learning image reconstruction (DLIR) in head and neck CT angiography (CTA) with reduced radiation and contrast doses, in comparison with routine dose scans with 100kVp and adaptive statistical iterative reconstruction V (ASIR-V).

METHODS: This prospective study included 100 patients scheduled for head and neck CTA due to vascular diseases. Patients were prospectively and randomly assigned in a 1:1 ratio to either the lower-dose group (group A, n = 50) with 80kVp, automatic tube current modulation (ATCM) scanning and contrast medium (CM) dose rate of 1.2gI/s for 10 s, and conventional group (group B, n = 50) with 100kVp, ATCM and CM dose rate of 1.6gI/s for 10 s. Group A used the high-level DLIR (DLIR-H) (subgroup A1), medium-level DLIR (DLIR-M) (subgroup A2), and 50% ASIR-V (A3) for image reconstruction; while Group B used ASIR-V50% for image reconstruction. The CT attenuation and standard deviation (SD) values of the sternocleidomastoid muscle (SCM) at the level of hypopharynx and epiglottis and the cerebral white matter (WM) at the level of splenium of corpus callosum were measured. For both quantitative and qualitative assessments, we evaluated image quality across the four groups. Objectively, we calculated the signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and edge rise slope (ERS) to quantify vessel sharpness. Subjectively, image quality was assessed using subjective image quality scores (SIQS). Finally, using digital subtraction angiography (DSA) as the reference standard, we compared the diagnostic accuracy of the four groups for determining the degree of vessel stenosis.

RESULTS: Group A showed significant reductions in contrast dosage, injection rate, and radiation dose by 25.61%, 28.38%, and 36.11%, respectively, compared to group B (all P < 0.001). After contrast dose reduction, the vessel attenuation values in group A were similar to that in group B. However, the four image groups differed significantly in terms of background noise (all P < 0.05) with group A1 having the lowest noise. Group A1 also had significantly higher SNR and CNR values compared to group B in all vessels (all P < 0.05). In addition, Group A1 had the highest SIQS, followed by A2, B, and A3 with good agreement between the two reviewers in all groups (κ values 0.88---1). With DSA as the reference standard, analysis of the 23 segments with > 50% stenosis showed that group A1 with DLIR-H exhibited a strong correlation with DSA in quantifying vascular stenosis (Pearson's r = 0.97) and achieved the lowest MSE (14.83) among the four reconstruction types.

CONCLUSIONS: Compared to the standard-dose protocol using 100 kVp and ASIR-V50%, a protocol using 80 kVp combined with DLIR-H significantly reduces the radiation dose (by 36%), contrast dose (26%), and injection rate (28%) in head and neck CTA while still significantly improving image quality and providing better diagnostic agreement with DSA.

CLINICAL RELEVANCE: Combination of low tube voltage and DLIR-H may be used to reduce patient doses while improving image quality and diagnostic accuracy in head and neck CT imaging.

来源:Eur J Radiol,查看原文。版权归原作者所有。