文献速递胸部介入CT

改良微弹簧圈装置用于肺结节术前定位的前瞻性研究

The utility of a modified microcoil device for pulmonary nodule localization: a single-centre prospective study.
PubMed 文献 · Eur J Radiol · 2026-08-01

本研究为单中心前瞻性单臂研究,纳入178例患者共231枚≤10 mm、cN0的肺小结节,评估新型CT引导微弹簧圈定位系统在VATS术前定位中的效能与安全性。技术成功定义为弹簧圈置入距病灶10 mm内且无断裂、移位或脱落,成功率为97.0%。并发症轻微,无症状气胸发生率5.2%,轻微出血发生率4.8%。

要点

  • 微弹簧圈定位技术成功率97.0%,定位时间平均4.4分钟,适合非可触及肺小结节VATS术前定位
  • 无症状气胸5.2%及轻微出血4.8%,均无需处理,并发症发生率优于传统hook-wire
  • 3.0%结节发生定位失败,原因为导丝断裂、脱落或移位,阅片报告时需关注弹簧圈与病灶关系

核心结论

  • 该新型微弹簧圈系统对肺小结节VATS术前定位具有高效能、操作效率高且安全性良好
  • 技术成功率为97.0%(95% CI: 93.7-98.6%),所有结节均首次切除且切缘阴性

研究设计与样本

  • 单中心前瞻性单臂研究,纳入178例患者共231枚≤10 mm、cN0的肺小结节
  • 均采用CT引导下新型微弹簧圈系统行术前定位,随后行VATS切除
  • 主要终点为技术成功,定义为弹簧圈置入距病灶10 mm内且无断裂、移位或脱落

关键影像表现/征象

  • 新型微弹簧圈系统含同轴针及带长可视远端尾端的可回收弹簧圈,以改善稳定性与术中识别
  • 失败病例表现为导丝断裂、弹簧圈脱落或移位,发生于3.0%的结节

诊断标准与关键数值

  • 技术成功标准:弹簧圈置入距病灶10 mm内且无断裂、移位或脱落,成功率为97.0%(95% CI: 93.7-98.6%)
  • 平均定位时间为4.4 min,弹簧圈至结节距离中位数为4.9 mm
  • 无症状气胸发生于12例(5.2%),轻微出血发生于11例(4.8%),均无需处理
  • 97.4%的弹簧圈随标本一并取出

临床应用与阅片要点

  • 术前CT定位报告应描述弹簧圈与目标结节的距离及有无断裂、移位或脱落征象
  • 弹簧圈尾端较长且可视的设计有利于VATS术中识别,阅片时需确认尾端位置
  • 出现少量气胸或肺实质出血时,若无症状通常无需干预

局限性

  • 单中心研究设计,结果外推需谨慎
  • 单臂设计缺乏与传统定位方法的同期随机对照

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: Video-assisted thoracoscopic surgery (VATS) is the standard treatment for small pulmonary nodules (SPNs), yet accurate localization of non-palpable lesions remains challenging. This study evaluated the efficacy, feasibility, and safety of a novel computed tomography (CT)-guided microcoil localization system for preoperative SPN management.

METHODS: We developed a novel CT-guided microcoil system featuring a coaxial needle and a retractable coil with a long, visible distal tail to improve stability and intraoperative identification. In this prospective single-arm study, 178 patients with 231 SPNs (≤10 mm, cN0) underwent preoperative localization followed by VATS. The primary endpoint was technical success, defined as deployment within 10 mm of the lesion without fracture, displacement, or dislodgement.

RESULTS: Technical success was achieved in 224 of 231 nodules, yielding a rate of 97.0% (95% CI: 93.7-98.6%). The mean localization time was 4.4 min, and the median coil-to-nodule distance was 4.9 mm. All nodules were completely resected with negative margins on the first attempt and were successfully identified for definitive pathological diagnosis. Complications were infrequent and mild: asymptomatic pneumothorax occurred in 12 cases (5.2%) and minor hemorrhage in 11 (4.8%), none of which required treatment; these rates compare favorably to those of conventional hook-wire systems. Patient pain was negligible, and 97.4% of coils were retrieved with the specimen. Failures occurred in 3.0% of nodules due to wire fracture, dislodgement, or displacement.

CONCLUSIONS: This novel microcoil system demonstrated high efficacy, procedural efficiency, and excellent safety, addressing key limitations of existing localization methods.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Xu, Ren, Lv, et al. The utility of a modified microcoil device for pulmonary nodule localization: a single-centre prospective study. Eur J Radiol. 2026. PMID: 42551306. doi:10.1016/j.ejrad.2026.113122

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。