三维超声与CT成像在腹主动脉瘤评估中的比较
结果显示三维超声技术可行,与CTA对应良好,加入造影剂后体积测量一致性提高。
要点
- 3D US与3D-CEUS均可完成AAA成像检查并支持离线回顾性评估
- 3D-CEUS管腔体积显著大于CTA,总AAA体积与CTA相当
- 3D US总AAA体积显著小于CTA,阅片时需注意体积系统性偏低
核心结论
- 3D US与3D-CEUS用于AAA成像具有良好的可行性,与CTA对应良好
- 加入超声造影剂后AAA总体积测量与CTA的一致性提高
- 3D US支持回顾性离线评估,相较于二维超声具有重要附加价值
研究设计与样本
- 所有患者均完成3D US、3D-CEUS及CTA三种模态的成像
- 对全部三种模态进行AAA整体分割,对3D-CEUS与CTA进行管腔分割
- 将CTA分割结果配准到对应三维超声分割上并计算评价指标
- 参考文献标准为增强CT血管成像
关键影像表现/征象
- 三维超声可生成AAA的三维整体概览,无需有害辐射及肾毒性对比剂
- 三维超声造影可用于管腔显示与分割
- 三维超声支持回顾性离线AAA评估
诊断标准与关键数值
- 3D-CEUS管腔体积中位数94 mL,CTA管腔体积中位数70 mL,差异有统计学意义(P<.001)
- 3D US总AAA体积中位数142 mL,CTA为165 mL,差异有统计学意义(P=.002)
- 3D-CEUS总AAA体积中位数176 mL,CTA为165 mL,差异无统计学意义(P=.433)
- Dice相似系数范围为0.82至0.88,平均表面距离为2.5 mm
临床应用与阅片要点
- 3D US可作为AAA影像评估的低成本、无辐射替代方法
- 对比剂增强三维超声可提高AAA体积测量的准确性
- 三维超声支持离线分析,可用于随访和科室质控场景
局限性
- 体积测量存在模态间系统性差异
- 作者指出需大规模研究进一步验证三维超声在AAA中的应用价值
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: Three-dimensional ultrasound (3D US) is emerging as a novel imaging method for abdominal aortic aneurysms (AAAs) as it creates a low-cost 3D AAA overview without harmful radiation and nephrotoxic contrast agents, as opposed to conventional imaging. 3D US can be used with a non-nephrotoxic contrast agent (3D-CEUS) for lumen delineation. This study aims to compare 3D US and 3D-CEUS using a commercially available 3D US system against contrast-enhanced computed tomography (CTA).
METHODS: For twenty AAA patients, the total AAA was segmented for all 3 modalities, and the lumen was segmented in the 3D-CEUS and CTA scans. The CTA segmentations were registered onto each 3D (CE)US segmentation, and evaluation metrics were computed, including the volumes, average surface distance, sensitivity, precision, and Dice score.
RESULTS: 3D US and 3D-CEUS scans were feasible in all patients. Lumen volumes were significantly larger on 3D-CEUS compared to CTA (94 versus 70 mL, p < .001). The total AAA volumes were significantly smaller on 3D US compared to CTA (142 versus 165 mL, p = .002), and comparable on 3D-CEUS (176 versus 165 mL, p = .433). The Dice scores were overall acceptable, ranging from 0.82 to 0.88, and the average surface distances were 2.5 mm.
CONCLUSIONS: This study demonstrated the feasibility of 3D US and 3D-CEUS for AAA imaging with good correspondence to CTA. Addition of the contrast agent increased agreement of volume measurements. 3D US facilitated retrospective offline AAA assessment, which is a major added value compared to 2D US. For now, 3D US already seems a promising imaging modality, which has to be proven in a large-scale study.