文献速递胸部儿科影像技术与安全CT

优化儿童胸部CT扫描范围以降低辐射剂量

Optimising chest computed tomography (CT) scan ranges in paediatric patients to reduce radiation exposure.
PubMed 文献 · Clin Radiol · 2026-06-25

该研究纳入3174例3-14岁接受平扫胸部CT的儿童,分两个队列确定并验证精准扫描范围方案。精准方案下肺实质漏检概率低于5%,有效剂量降低17.0%-19.2%。结论认为该方案可在满足诊断需求的同时降低辐射暴露。

要点

  • 儿童胸部CT下界可设为肋膈角下方35 mm(3-6岁)或40 mm(7-14岁)
  • 精准方案使z轴覆盖范围减少4.7%-6.1%,肺实质漏检概率低于5%
  • 报告可关注扫描范围是否按年龄调整,以避免不必要的辐射暴露

核心结论

  • 精准扫描范围方案可在保证肺实质覆盖的同时使有效剂量降低17.0%-19.2%
  • 优化方案中上下界肺实质漏检概率均低于5%
  • 该策略在诊断质量与患者安全间取得平衡,未来多中心验证及AI自动化可进一步推动临床应用

研究设计与样本

  • 两队列研究,共纳入3174例3-14岁接受平扫胸部CT的儿童
  • 队列1共1894例,回顾性分析肺边界变异以确定最优扫描范围
  • 队列2共1280例,与传统方案比较验证肺边界勾画准确性及辐射剂量降低效果
  • 辐射剂量以剂量长度乘积结合年龄特异性转换系数(k=0.018 mSv·mGy⁻¹·cm⁻¹)计算有效剂量

关键影像表现/征象

  • 传统方案上界为第一肋上缘,下界为膈肌以下由技师判断,覆盖肺尖至肺底
  • 精准方案上界与第一肋对齐(0 mm偏移),下界为肋膈角下方35 mm(3-6岁)或40 mm(7-14岁)
  • 优化后z轴覆盖范围在3-6岁组减少4.7%,在7-14岁组减少6.1%

诊断标准与关键数值

  • 3-6岁组精准方案下界为肋膈角下方35 mm;7-14岁组下界为肋膈角下方40 mm
  • 精准方案上界与第一肋对齐(0 mm偏移)
  • 3-6岁组有效剂量降低19.2%(0.09 mSv),剂量长度乘积降低17.7%(4.5 mGy·cm)
  • 7-14岁组有效剂量降低17.0%(0.21 mSv),剂量长度乘积降低17.0%(11.7 mGy·cm)
  • 上下界肺实质漏检概率均低于5%

临床应用与阅片要点

  • 儿童胸部平扫CT可依据年龄设定下界,替代技师依赖的膈肌下方随意定位
  • 阅片时应确认扫描范围上界是否与第一肋对齐、下界是否按年龄置于肋膈角下方35-40 mm
  • 注意3-6岁与7-14岁下界标准不同,避免将大龄儿童标准误用于低龄儿童

局限性

  • 本研究为单中心设计,未来需多中心验证
  • 精准方案依赖人工定位解剖标志,AI驱动自动化尚待开发

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

AIM: The aim of this study was to accurately position the scan range of unenhanced chest computed tomography (CT) scans for paediatric patients by clarifying the upper and lower scan boundary ranges to meet clinical needs while minimising radiation risks.

MATERIALS AND METHODS: This two-cohort study included 3174 children (aged 3-14 years) undergoing unenhanced chest CT scans. Cohort 1 (n=1894) was used to retrospectively analyse lung boundary variability to determine optimal scan ranges, while cohort 2 (n=1,280) was used to validate the accuracy in delineating lung boundaries and reducing radiation dose compared with traditional protocols. Traditional protocol: upper lung border at the first rib's upper edge; lower border below the diaphragm (technician dependent), covering apex to base. Precise protocol: upper border aligns with the first rib (0 mm offset); lower border 35 mm (3-6 years) or 40 mm (7-14 years) below costophrenic angle. Radiation dose calculation: effective dose (ED) derived from dose-length product (DLP) using age-specific conversion coefficients (k = 0.018 mSv·mGy-1·cm-1).

RESULTS: Scan range optimisation reduced z-axis coverage by 4.7% (3-6 years) and 6.1% (7-14 years). The probability of missed lung parenchyma was <5% for both upper and lower boundaries using the optimised protocol. Radiation dose reductions were as follows: 3-6 year: ED decreased by 19.2% (0.09 mSv) and DLP by 17.7% (4.5 mGy-1cm1); 7-14 years: ED decreased by 17.0% (0.21 mSv) and DLP by 17.0% (11.7 mGy-1cm1).

CONCLUSION: The proposed protocol achieves precise lung coverage while reducing radiation exposure by 17-19%. This strategy balances diagnostic quality and patient safety, offering a scalable solution for paediatric unenhanced chest CT imaging. Future multicentre validation and artificial intelligence (AI)-driven automation could further enhance clinical adoption.

来源:Clin Radiol,查看原文。版权归原作者所有。

引用

  1. Yang, Wang, Cai, et al. Optimising chest computed tomography (CT) scan ranges in paediatric patients to reduce radiation exposure. Clin Radiol. 2026. PMID: 42497578. doi:10.1016/j.crad.2026.107420

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。