EID-CT与PCD-CT冠状动脉斑块定量可重复性及样本量模型

Coronary plaque quantification on energy-integrating and photon-counting detector CT: reproducibility and power modeling for mixed-platform trials.
PubMed 文献 · Eur Radiol · 2026-07-22

优化重建组合Qr40EID-CT与Bv36PCD-CT时,血管水平标准差0.14、患者水平0.12。检测斑块总体积5%和10%变化时,每组需116和29支血管或92和23例患者。

要点

  • Qr40EID-CT与Bv36PCD-CT配对重建时斑块体积跨机型一致性最高。
  • 钙化斑块跨机型一致性最强,低衰减斑块相对最弱但仍为强相关。
  • 混合平台纵向试验检测斑块体积10%变化时每组仅需29支血管或23例患者。

核心结论

  • 优化重建条件下定量冠脉斑块体积在EID-CT与PCD-CT之间具有高度跨机型一致性。

研究设计与样本

  • 同一患者30天内先后接受双源EID-CT冠脉CTA与PCD-CT检查,EID-CT用Qr40和Bv40核重建。
  • PCD-CT用Bv36/40/44与Qr36/40/44核、量子迭代重建强度2–4重建。
  • 采用深度学习自动化斑块定量平台,按固定与自适应亨氏单位阈值计算斑块成分体积并空间配准。

诊断标准与关键数值

  • 各斑块成分跨机型相关系数:总体积r=0.85–0.95,纤维斑块r=0.74–0.94,钙化斑块r=0.92–0.98。
  • 低衰减斑块体积相关系数r=0.71–0.85,总体积平均偏倚范围8.2–164.4 mm³。
  • 最优重建组合Qr40EID-CT与Bv36PCD-CT的跨机型标准差为0.14(血管水平)和0.12(患者水平)。
  • 80%把握度、α=0.05时,检测总体积5%变化需116支血管或92例患者每组,10%变化需29支血管或23例患者每组。

临床应用与阅片要点

  • 报告冠脉斑块定量结果时应注明重建核组合,优先采用Qr40EID-CT与Bv36PCD-CT配对以保证可比性。
  • 钙化斑块跨机型差异最小,低衰减斑块差异最大,随访低衰减斑块变化时需更谨慎解读。

局限性

  • EID-CT与PCD-CT检查间隔30天内,期间斑块本身可能存在演进影响一致性评估。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To evaluate the reproducibility of quantitative plaque characterization between energy-integrating detector (EID)-CT and photon-counting detector (PCD)-CT, identify reconstruction settings yielding the lowest variability, and model sample-size requirements for trial planning.

MATERIALS AND METHODS: Patients who underwent coronary CT angiography on dual-source EID-CT and PCD-CT within 30 days were screened retrospectively. EID-CT data were reconstructed using quantitative (Qr40) and vascular (Bv40) kernels, while PCD-CT data were reconstructed with Bv36/40/44 and Qr36/40/44 kernels and quantum iterative reconstruction strengths 2-4. For each plaque component (total, low-attenuation, fibrotic, and calcified), volumes were computed using fixed and adaptive Hounsfield-unit thresholds using an automated deep-learning-based plaque-quantification platform and spatially co-registered. Inter-scanner standard deviation (SD) was calculated, and optimal reconstruction pairs were used for power modeling.

RESULTS: Thirty-eight patients (age 68.0 [64.0-72.5] years, 30 men) and 77 vessels were included. Inter-scanner correlations were very strong for total (r = 0.85-0.95), fibrotic (r = 0.74-0.94), and calcified plaque (r = 0.92-0.98), and strong for low-attenuation plaque volumes (r = 0.71-0.85). Mean bias ranged from 8.2 to 164.4 mm³ for total plaque volume. Applying the optimal reconstruction pair (Qr40EID-CT vs. Bv36PCD-CT), inter-scanner SDs were 0.14 (vessel-based) and 0.12 (patient-based). At 80% power and α = 0.05, the estimated sample sizes to detect 5% and 10% changes in total plaque volume were 116 and 29 vessels or 92 and 23 patients per group.

CONCLUSION: Quantitative coronary plaque volumes demonstrated high inter-scanner consistency between EID-CT and PCD-CT under optimized conditions. The dual-level power-modeling framework translates inter-scanner variability into actionable sample-size estimates for study design.

KEY POINTS: Question Quantitative coronary plaque volumes are increasingly used as surrogate biomarkers, yet variability between energy-integrating and photon-counting CT remains a major concern. Findings Harmonized reconstruction settings achieved high inter-scanner consistency with low variability when optimized reconstruction settings and automated plaque quantification are applied. Clinical relevance Standardized coronary plaque quantification across CT technologies enables reliable use of quantitative plaque endpoints, supporting mixed-platform longitudinal trials.

来源:Eur Radiol,查看原文。版权归原作者所有。