文献速递泌尿生殖技术与安全CTUS

便携式膀胱扫描仪在盆腔放疗膀胱管理中的验证与预测模型

Validation and predictive modeling of a portable bladder scanner for precision bladder management in pelvic radiotherapy.
PubMed 文献 · Br J Radiol · 2026-01-01

该研究回顾性分析114例使用Caresono PadScan HD5评估的患者与111例历史对照,验证HD5膀胱容积测量与定位CT的一致性。结果显示HD5与定位CT容积相关但系统性低估,且低估程度随膀胱容积增大而加重。引入HD5后初始CBCT重复率从48.7%降至22.8%。

要点

  • HD5测得的膀胱容积与定位CT中度相关(r=0.723,P<.001),但平均低估34.39 mL。
  • 膀胱容积≥300 mL时低估更明显(-71.46 mL vs -18.64 mL,P<.001),报告时需注意容积依赖性偏倚。
  • 年龄分层的中等充盈流程可减少初始CBCT重复率,建议按年龄分层解读HD5数据。

核心结论

  • Caresono PadScan HD5膀胱容积测量可靠但存在负性系统偏差,平均低估34.39 mL。
  • 膀胱容积≥300 mL时低估幅度显著增大,年龄小于65岁者低估斜率更陡。
  • HD5应用后初始CBCT重复率由48.7%下降至22.8%(P<.001),提示其可优化盆腔放疗膀胱准备流程。

研究设计与样本

  • 回顾性研究,纳入114例使用HD5评估的患者和111例历史对照,共两个队列。
  • 以盆腔放疗定位CT的膀胱容积为参考标准,评估HD5测量值与定位CT的一致性。
  • 采用相关性分析、回归分析、Bland-Altman分析和配对检验评价两种测量方法的一致性。
  • 通过亚组分析(膀胱容积<300 mL vs ≥300 mL;年龄<65岁 vs ≥65岁)、性别和多变量线性回归探索测量误差的预测因素。

关键影像表现/征象

  • HD5测量膀胱容积与定位CT容积呈中度正相关(r=0.723,P<.001)。
  • Bland-Altman分析显示平均偏差为-34.39 mL,HD5呈系统性低估。
  • 定位CT膀胱容积≥300 mL时,HD5低估幅度为-71.46 mL,明显大于<300 mL组的-18.64 mL(P<.001)。
  • 多变量回归显示膀胱容积是测量误差的主要驱动因素,年龄与膀胱容积存在交互作用。

诊断标准与关键数值

  • HD5与定位CT容积相关系数r=0.723(P<.001)。
  • HD5平均偏差为-34.39 mL,提示存在负性系统误差。
  • 膀胱容积≥300 mL时低估-71.46 mL,膀胱容积<300 mL时低估-18.64 mL,组间差异P<.001。
  • 年龄<65岁患者较年长患者低估斜率更陡(交互作用P<.001)。

临床应用与阅片要点

  • 日常阅片使用HD5数据时应意识到其系统性低估,膀胱越充盈低估越明显。
  • 对年轻患者(<65岁)的HD5读数应更谨慎,因其低估程度随容积增大而更为显著。
  • 建议采用年龄分层的中等充盈流程,利用HD5测量指导盆腔放疗患者膀胱准备。
  • 将HD5测量嵌入常规盆腔放疗流程可标准化膀胱准备,并降低因膀胱充盈不足导致的初始CBCT重复扫描。

局限性

  • 研究为回顾性设计,存在选择偏倚和信息偏倚的潜在风险。
  • 历史对照组的可比性受限于时间推移及流程改变等混杂因素的可能影响。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: Validate the accuracy of a portable bladder scanner (Caresono PadScan HD5) against planning CT and demonstrate its clinical application to optimize bladder preparation in pelvic radiotherapy.

METHODS: Two retrospective cohorts were analyzed: 114 patients assessed with HD5 and 111 historical controls. Agreement between HD5-measured volume (VHD5) and planning CT (VCT) was evaluated with correlation, regression, Bland-Altman, and paired tests. Initial CBCT repetition rates were compared between cohorts. Subgroup analyses (bladder volume [BV]<300 vs ≥300 mL; age<65 vs ≥ 65 years); sex and multivariable linear regression with interaction terms explored predictors of measurement error.

RESULTS: V HD5 correlated with VCT (r = 0.723, P < .001) but underestimated volumes (mean bias -34.39 mL). Underestimation was markedly greater at VCT ≥300 mL (-71.46 vs -18.64 mL, P < .001). Multivariable regression confirmed BV as the primary driver, with an age-dependent interaction showing steeper underestimation in younger (<65 years) patients (P < .001). Following HD5 implementation, initial CBCT repetition rates decreased from 48.7% to 22.8% (P < .001).

CONCLUSIONS: HD5 enables reliable, though negatively biased, BV estimates. Embedding an age-stratified, moderate-fill workflow reduces repeat initial CBCT and standardizes bladder preparation.

ADVANCES IN KNOWLEDGE: We quantify volume-dependent underestimation and an age interaction, and translate them into a predictive, age-stratified protocol that operationalizes HD5 measurements for routine pelvic RT bladder management.

来源:Br J Radiol,查看原文。版权归原作者所有。

引用

  1. Huang, Fu, Zhang, et al. Validation and predictive modeling of a portable bladder scanner for precision bladder management in pelvic radiotherapy. Br J Radiol. 2026. PMID: 42438151. doi:10.1093/bjr/tqag138

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。