孕期MRI与经阴道超声测量宫颈长度的对照研究

Cervical length assessment using magnetic resonance imaging during pregnancy: correlation and agreement with transvaginal ultrasound.
PubMed 文献 · Ultrasound Obstet Gynecol · 2026-07-10

回顾性横断面研究纳入100例同日接受胎儿MRI与经阴道超声检查的孕妇,以经阴道超声为金标准。MRI与超声宫颈长度测量呈强相关,Bland-Altman分析示两种方法平均差为-0.97mm。孕中期以MRI宫颈长度≤25mm预测超声短宫颈的敏感度为100%,特异度为94.6%。

要点

  • MRI与超声宫颈长度强相关,但平均差达-0.97mm,报告时应注明测量方法差异
  • 孕中期MRI宫颈长度≤25mm对排除超声短宫颈的阴性预测值为100%,可用于筛选
  • 经阴道超声仍是孕期宫颈评估的金标准,MRI测量不能完全替代

核心结论

  • MRI测量的宫颈长度与经阴道超声测量结果呈强相关,r=0.775(P<0.001)
  • 孕中期MRI宫颈长度≤25mm预测超声短宫颈的敏感度为100%,特异度为94.6%
  • MRI测量宫颈长度可能有助于在因其他指征行MRI的特定患者中排除短宫颈,但经阴道超声仍是金标准

研究设计与样本

  • 共筛选132例,排除资料不全31例及宫颈环扎1例后,100例纳入分析
  • 以经阴道超声作为宫颈长度测量的金标准,采用Spearman相关及Bland-Altman法评价一致性
  • 对孕中期亚组,计算MRI宫颈长度≤25mm预测超声宫颈长度<25mm的诊断效能

关键影像表现/征象

  • MRI与经阴道超声来源的宫颈长度测量强相关,r=0.775
  • Bland-Altman分析显示两种方法宫颈长度测量的平均差为-0.97mm,95%一致性界限为-13.67至11.73mm
  • 无论孕妇体重指数、单胎或多胎、超声宫颈长度<25mm或≥25mm,MRI与超声宫颈长度相关性均保持显著

诊断标准与关键数值

  • 以经阴道超声宫颈长度<25mm定义为短宫颈
  • 孕中期MRI宫颈长度≤25mm预测超声短宫颈的敏感度为100%(95%CI:59.0-100%)
  • 孕中期MRI宫颈长度≤25mm预测超声短宫颈的特异度为94.6%(95%CI:86.9-98.5%)
  • 孕中期MRI宫颈长度≤25mm预测超声短宫颈的阳性预测值为63.6%(95%CI:30.8-89.1%),阴性预测值为100%(95%CI:94.9-100%)
  • Spearman相关系数r=0.775,P<0.001

临床应用与阅片要点

  • 因其他指征行胎儿MRI的孕妇,报告中可附带宫颈长度测量,用于排除短宫颈
  • 若孕中期MRI显示宫颈长度>25mm,排除超声短宫颈的把握较高
  • MRI与超声宫颈长度存在系统性差异,解读MRI宫颈长度时不宜直接套用超声阈值

局限性

  • 回顾性横断面设计,样本量较小,作者指出需要更大规模研究验证结果
  • 研究排除宫颈环扎患者,结论不适用于该类人群
  • 单中心转诊人群,存在选择偏倚可能
  • 经阴道超声仍被确认为孕期宫颈评估的金标准

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To assess the correlation between cervical length (CL) measurements obtained via fetal magnetic resonance imaging (MRI) and transvaginal ultrasound (TVS), and the accuracy of CL measurements obtained via MRI in diagnosing a second-trimester short cervix, using TVS as the gold standard.

METHODS: This was a retrospective cross-sectional study of pregnancies referred to our center between January 2022 and December 2024 that underwent fetal MRI and TVS examination on the same day. We estimated the correlation between MRI- and TVS-derived CL measurements using Spearman's rank correlation coefficient, with stratified analysis by pertinent clinical variables. Bland-Altman analysis was used to evaluate the agreement between the two imaging methods. Finally, for the subset of patients examined in the second trimester, we calculated the sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV) of CL ≤ 25 mm on fetal MRI for predicting a short cervix with CL < 25 mm on TVS.

RESULTS: We screened 132 patients, of whom 31 were excluded owing to incomplete data and one was excluded due to having a cervical cerclage in place. Therefore, 100 patients were included for analysis. There was a strong correlation between MRI- and TVS-derived CL measurements (r = 0.775; P < 0.001), which remained significant regardless of maternal body mass index, pregnancy type (singleton vs multiple) and CL on TVS (< 25 mm vs ≥ 25 mm). According to the Bland-Altman analysis, the mean difference in CL measurements between MRI and TVS was -0.97 (95% limits of agreement, -13.67 to 11.73) mm. The sensitivity, specificity, PPV and NPV of MRI-derived CL ≤ 25 mm for predicting a short cervix on TVS in the second trimester were 100% (95% CI, 59.0-100%), 94.6% (95% CI, 86.9-98.5%), 63.6% (95% CI, 30.8-89.1%) and 100% (95% CI, 94.9-100%), respectively.

CONCLUSIONS: CL measurements obtained using fetal MRI correlate with those obtained using TVS. MRI-derived CL estimates may help to exclude a short cervix during pregnancy in selected patients undergoing MRI for other indications. However, larger studies are needed to validate our findings, and TVS remains the gold standard for cervical assessment during pregnancy. © 2026 International Society of Ultrasound in Obstetrics and Gynecology.

来源:Ultrasound Obstet Gynecol,查看原文。版权归原作者所有。

引用

  1. Schenone, Fadiloglu, Sambatur, et al. Cervical length assessment using magnetic resonance imaging during pregnancy: correlation and agreement with transvaginal ultrasound. Ultrasound Obstet Gynecol. 2026. PMID: 42430210. doi:10.1002/uog.70269

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。