容积超声定位显微镜肾小球成像的验证研究

A Validation of Spatially Compounded Volumetric Ultrasound Localization Microscopy for Glomerular Imaging With Light Sheet Microscopy.
PubMed 文献 · Ultrasound Med Biol · 2026-06-09

本研究对5周龄大鼠行全肾超声定位显微镜(ULM)成像,并与光片显微镜(LSM)金标准对比验证。ULM以500体积/秒采集容积数据,重建肾血管并测量肾小球密度。结果显示ULM可显示皮质血管和单个肾小球,肾小球密度与LSM及文献组织学值一致。

要点

  • ULM可显示活体全肾皮质血管和单个肾小球结构,分辨率达亚衍射水平。
  • ULM测得肾小球密度21.3±8.0个/mm³,与LSM及组织学报告值接近。
  • ULM可同时显示肾内微血管血流速度,未来或用于肾微结构及血流动力学评估。

核心结论

  • ULM作为活体肾小球定量评估的概念验证方法,以LSM为金标准获得验证,有望用于全肾肾小球水平微结构与血流动力学评估。

研究设计与样本

  • 研究类型为动物实验验证研究,对象为1只5周龄大鼠,行全肾ULM与LSM配对对比。
  • ULM使用定制1024通道Verasonics系统配32×32矩阵阵列换能器,以500体积/秒采集容积超声数据。
  • 微泡对比剂持续输注,经波束形成、奇异值分解滤波、微泡定位与追踪重建肾血管。
  • 超声后取肾固定,iDISCO+方案光学透明化后行LSM成像,对100个随机匹配的1mm³亚容积行非刚性配准对比。

关键影像表现/征象

  • ULM以亚衍射分辨率显示皮质血管和单个肾小球。
  • ULM显示肾小球密度21.3±8.0个/mm³,LSM为28.6±7.5个/mm³。
  • ULM显示肾内微血管血流速度分布。

诊断标准与关键数值

  • ULM肾小球密度21.3±8.0个/mm³,LSM肾小球密度28.6±7.5个/mm³(100个独立1mm³样本)。
  • 文献报告组织学肾小球密度为24.8±0.78个/mm³。

临床应用与阅片要点

  • ULM可用于活体全肾肾小球水平微结构成像,提供肾小球密度评估。
  • ULM可同时提供肾内微血管血流速度信息,有助评估肾功能与血流动力学。

局限性

  • 本研究为单只动物概念验证,未报告多动物重复性及变异。
  • ULM测得肾小球密度低于LSM,可能与成像原理及配准误差有关。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: Glomerular structure and microvascular function are central to kidney health, yet current clinical assessments rely largely on indirect serum markers or invasive biopsy. Imaging approaches capable of resolving glomeruli in vivo across the entire kidney remain limited by spatial and temporal resolution. In this study, we perform whole-kidney ultrasound localization microscopy (ULM) with quantification of glomerular microvascular structure and function, and evaluate its performance against light-sheet microscopy (LSM), an established optical gold standard.

METHODS: A 5-wk-old rat underwent kidney exposure and stabilization for high-frame-rate ULM imaging. Microbubble contrast agents were continuously infused, and volumetric ultrasound data were acquired at 500 volumes/s using a custom 1024-channel Verasonics system with a 32 × 32 matrix array transducer. ULM processing included beamforming, singular value decomposition filtering, microbubble localization, and tracking to reconstruct renal vasculature. Three spatially registered volumes were combined to form a whole-kidney dataset. Following ultrasound imaging, the kidney was fixed, optically cleared using the iDISCO+ protocol, and imaged with LSM. Non-rigid registration enabled comparison of glomerular density across 100 randomly selected, spatially matched 1 mm³ subvolumes.

RESULTS: ULM provided whole-organ imaging with sub-diffraction resolution, enabling visualization of cortical vasculature and individual glomeruli. Glomerular densities measured 21.3 ± 8.0 glomeruli/mm³ with ULM and 28.6 ± 7.5 glomeruli/mm³ with LSM across 100 independent samples, consistent with reported histological values (24.8 ± 0.78 glomeruli/mm³). ULM additionally visualized microvascular flow velocities throughout the kidney.

CONCLUSION: This work validates ULM as a proof-of-concept approach for quantitative in vivo assessment of glomeruli, using LSM as the gold standard. These results highlight ULM as a promising approach for whole-kidney, glomerulus-level assessment of renal microstructure and hemodynamics, with potential for future translational applications.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。