护士主导胃准备在光子计数CT虚拟胃镜成像的应用
本研究为单中心前瞻性随机试验,纳入120例行腹部增强CT的患者,分为PCCT-3g、PCCT-6g、EID-6g及PCCT-3g+W四组。评估护士主导的五步胃准备方案对气体阴性对比容积再现虚拟内镜成像质量的影响。结果显示6g产气粉优于3g,PCCT较EID-CT噪声更低、SNR/CNR更高、胃壁分层显示更清晰。
要点
- 气体阴性对比虚拟内镜成像推荐使用6g产气粉,3g仅适用于常规胃评估
- PCCT在相同产气剂量下较EID-CT噪声更低、胃壁分层显示更清晰
- 饮水辅助可增加胃窦扩张评分,但不改善气体阴性对比虚拟内镜成像质量
核心结论
- 护士主导的PCCT胃准备可根据不同成像目的进行定制,6g产气粉更适合气体阴性对比虚拟内镜成像,3g对常规胃评估可能已足够
- 在相同6g产气粉方案下,PCCT较EID-CT具有更低的图像噪声、更高的SNR和CNR以及更清晰的胃壁分层显示
- 饮水辅助可增加胃窦横截面评估价值,但未能改善气体阴性对比虚拟内镜成像质量
研究设计与样本
- 单中心前瞻性随机试验,纳入120例行对比增强腹部CT的患者,平均分为四组(每组30例)
- 四组分别为PCCT-3g、PCCT-6g、EID-6g和PCCT-3g+W
- 所有胃准备均遵循标准化护士培训和护士主导的五步准备方案,包括患者指导、方案特异性胃准备、屏气指导、体位摆放和扫描时机控制
- 主要终点为气体阴性对比容积再现虚拟内镜成像质量评分,次要终点包括胃内气/液体积、整体及胃窦扩张程度、胃壁分层可见性、定量图像指标、患者耐受性和不良事件
关键影像表现/征象
- PCCT-6g组胃内气体体积较PCCT-3g组增加201.85 cm³,胃扩张和气体阴性对比虚拟内镜质量更优
- 相同6g产气粉剂量下,PCCT噪声为5.37 ± 0.72 HU,显著低于EID-CT的10.90 ± 1.98 HU
- PCCT的SNR为12.56 ± 1.92,CNR为18.89 ± 4.16,均显著高于EID-CT(SNR 5.08 ± 0.88,CNR 4.97 ± 1.80)
- 饮水辅助使PCCT-3g组胃内液体体积增加279.68 cm³,胃窦扩张评分提高2.13分
诊断标准与关键数值
- PCCT-6g与PCCT-3g相比,胃内气体体积、胃扩张及气体阴性对比虚拟内镜质量评分的差异均有统计学意义(均P<0.001)
- PCCT-6g组数字评分量表不适感评分为0.63 ± 0.85,高于PCCT-3g组的0.07 ± 0.25
- 相同6g剂量下PCCT与EID-CT在噪声、SNR、CNR及胃壁分层显示方面的差异均有统计学意义(均P<0.001)
- 饮水辅助未改善气体阴性对比虚拟内镜成像质量
临床应用与阅片要点
- 进行气体阴性对比虚拟内镜成像时优先选择6g产气粉方案,以获取充分的胃腔气体充盈和成像质量
- 常规胃部评估可考虑使用3g产气粉,在保证基本诊断需求的同时减少患者不适感
- PCCT平台在相同胃准备条件下可提供比EID-CT更低的噪声和更高的SNR/CNR,有利于胃壁细节观察
- 需要重点评估胃窦横截面时,可辅以饮水以增加胃窦扩张程度
局限性
- 本研究为单中心设计,样本量为每组30例,结果的推广性有限
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
RATIONALE AND OBJECTIVES: To evaluate nurse-led gastric preparation for photon-counting computed tomography (PCCT)-based gas-negative-contrast volume-rendered (gas-VRT) virtual endoscopic gastric imaging, focusing on effervescent powder dose, PCCT-related image-quality advantages, and water-assisted preparation.
MATERIALS AND METHODS: This single-center prospective randomized trial enrolled 120 patients undergoing contrast-enhanced abdominal CT. Participants were allocated into four groups (n = 30 each): PCCT-3g, PCCT-6g, energy-integrating detector (EID)-6g, and PCCT-3g+W. All gastric preparations followed standardized nurse training and a nurse-led five-step preparation protocol, including patient instruction, protocol-specific gastric preparation, breath-holding guidance, positioning, and scan-timing control. The primary endpoint was the gas-VRT virtual endoscopy-like imaging quality score. Secondary endpoints included intragastric gas/liquid volume, overall and antral gastric distension, gastric wall layer visibility, quantitative image metrics, patient tolerance, and adverse events.
RESULTS: All patients finished examinations without severe adverse events. PCCT-6g achieved a larger intragastric gas volume than PCCT-3g, with an increase of 201.85 cm³, superior gastric distension and gas-VRT quality (all P < 0.001), accompanied by higher discomfort (numeric rating scale score, 0.63 ± 0.85 vs 0.07 ± 0.25). At an identical 6 g effervescent dose, PCCT and EID-CT yielded comparable distension and gas-VRT images; PCCT provided lower noise (5.37 ± 0.72 HU vs 10.90 ± 1.98 HU) and improved signal-to-noise ratio (SNR) (12.56 ± 1.92 vs 5.08 ± 0.88), contrast-to-noise ratio (CNR) (18.89 ± 4.16 vs 4.97 ± 1.80), and wall layer visualization (all P < 0.001). Water supplementation increased intragastric liquid volume by 279.68 cm³ and antral distension score by 2.13 points in the PCCT-3g group, but failed to enhance gas-VRT quality.
CONCLUSION: Nurse-led PCCT gastric preparation can be customized for different imaging objectives. Six grams of effervescent powder is preferable for gas-VRT virtual endoscopic imaging, while 3 g may be sufficient for routine gastric evaluation. Under the same 6 g preparation protocol, PCCT provided lower image noise, higher SNR/CNR, and clearer gastric wall layer visualization than EID-CT, suggesting platform-specific value beyond gastric distension alone. Water assistance may benefits antral cross-sectional assessment but did not improve gas-VRT imaging.