文献速递神经外周血管

双能CTA豆纹动脉成像的能级与卷积核优化

Dual-energy computed tomography angiography for lenticulostriate arteries: Stepwise optimization of virtual monochromatic imaging and kernel setting.
PubMed 文献 · Eur J Radiol · 2026-01-26

本回顾性研究纳入556例头颈双能CTA患者,分两步在40–90 keV虚拟单能量与5种卷积核中优化重建参数。发现40 keV联合Hv49可显著提高豆纹动脉边缘锐利度与可见分支数,优于常规重建。

要点

  • 40 keV联合Hv49重建可增加豆纹动脉可见分支数,适用于头颈CTA常规评估。
  • 更锐利的卷积核可提高豆纹动脉边缘锐利度,阅片时优先选择Hv49-40 keV重建。
  • 低keV虚拟单能量图像对细小穿支动脉显示更佳,有助于豆纹动脉相关疾病评价。

核心结论

  • 头颈双能CTA采用Hv49-40 keV重建能显著改善不同管径动脉的显示,尤其提高豆纹动脉的可视化。

研究设计与样本

  • 回顾性研究,纳入556例患者(年龄59 [46-72]岁;男性310例 [55.8%]),2023年5月至2024年7月行头颈双能CTA。
  • 图像在40-90 keV虚拟单能量及Qr40、Hv36、Hv40、Hv44、Hv49五种卷积核下重建,分两步比较图像质量与最优核。

关键影像表现/征象

  • 40 keV重建获得最高CT值与CNR,对动脉分支显示的主观评分最佳。
  • Hv49-40 keV重建边缘锐利度最高,ERS为10374.4 [7407.0-13341.9],ERD为0.10 [0.08-0.13]。
  • Hv49-40 keV重建的豆纹动脉可见分支数最高,为6.5 [2.5-10.5](所有校正后p<0.001)。

诊断标准与关键数值

  • 不同keV水平间CT值与CNR存在显著差异(所有p<0.001),40 keV的CT值与CNR最高。
  • 使用更锐利的卷积核可显著提高ERS并降低ERD,Hv49-40 keV的边缘锐利度最佳。
  • Hv49-40 keV重建的豆纹动脉可见分支数显著高于其他重建组合(所有校正后p<0.001)。

临床应用与阅片要点

  • 在头颈双能CTA阅片中,可优先调取Hv49-40 keV重建序列评估豆纹动脉。
  • 评价豆纹动脉等细小穿支动脉时,应使用低keV虚拟单能量图像以提高显示率。

局限性

  • 本研究为回顾性设计,可能引入选择偏倚。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: The lenticulostriate arteries (LSAs) are critical perforating vessels, but conventional CTA poorly depicts them. Dual-Energy CT offers Virtual Monochromatic Imaging and kernel adjustment, yet the optimal strategy for LSAs visualization remains unclear.

MATERIALS AND METHODS: This retrospective study included 556 patients (59 [46-72] years; 310 males [55.8%]) who underwent DE-CTA between May 2023 to July 2024. Images were reconstructed Virtual Monochromatic Imaging at 40-90 KeV and multiple convolution kernels (Qr40, Hv36, Hv40, Hv44, and Hv49). Step 1 compared image quality across KeV levels, and Step 2 optimized kernels on the basis of the best KeV and compared these optimized reconstructions again. Quantitative metrics included CT value, SNR, CNR, edge rise distance, and edge rise slope; qualitative assessment and visible LSAs branch counts were performed.

RESULTS: Step 1 demonstrated significant differences in CT value, and CNR across KeV levels (all p < 0.001), with 40 KeV yielding the highest CT value and CNR and the best subjective scores for artery branches visualization. Step 2 showed that sharper kernels significantly improved ERS and reduced ERD, with Hv49-40 KeV achieving the greatest edge sharpness (ERS, 10374.4 [7407.0-13341.9]; ERD, 0.10 [0.08-0.13]) and the highest number of visible LSAs branches (6.5 [2.5-10.5]; all adjusted p < . 001).

CONCLUSION: Hv49-40KeV on DE‑CTA of the head and neck substantially improves the visualization of arteries across different calibers, with a marked enhancement in the depiction of the LSAs, providing a solid technical foundation for the evaluation of LSAs with DE‑CT.

RELEVANCE STATEMENT: This optimized Dual Energy-CT angiography reconstruction increases the utility of head and neck CTA by improving lenticulostriate artery visualization, thereby supporting routine evaluation of both large and small cerebral vessels in clinical practice.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Yin, Wang, Xu, et al. Dual-energy computed tomography angiography for lenticulostriate arteries: Stepwise optimization of virtual monochromatic imaging and kernel setting. Eur J Radiol. 2026. PMID: 41616759. doi:10.1016/j.ejrad.2026.112694

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。