手持三维谐波运动成像用于乳腺肿瘤评估的临床可行性研究
该研究开发了一种基于行列阵列换能器的手持三维谐波运动成像系统(3-D-HMI),可同时对乳腺肿瘤进行谐波激励与位移追踪。先在含不同硬度包块的仿体中验证,再对5例乳腺肿瘤患者成像,结果显示系统可快速、准确地进行三维相对硬度容积成像,具备临床转化潜力。
要点
- 三维HMI可显示整个肿瘤体积内的相对硬度分布,避免二维采样遗漏局部硬质成分
- 位移图与包块-背景杨氏模量比呈反比,阅片时低位移区提示相对更硬的成分
- 对残留病灶亦可成像,或可用于新辅助化疗后硬度变化的容积评估
核心结论
- 三维谐波运动成像可手持式、快速且准确地对乳腺肿瘤全容积进行相对硬度分布成像
- 该系统在仿体实验中可检出最小2 mm的包块,并在人乳腺肿瘤中完成三维重建
研究设计与样本
- 仿体验证纳入12个含坚硬或柔软包块的模型,包块直径2.2至13.2 mm,背景杨氏模量分别为5.3 kPa和18.6 kPa
- 人体成像纳入5例乳腺肿瘤,包括1例纤维腺瘤和4例浸润性导管癌,其中2例为新辅助化疗后残留病灶
- 谐波辐射力频率为200或400 Hz,使用行列阵列换能器的行或列单元发射,正交单元接收并估算位移
关键影像表现/征象
- 三维归一化位移图可检出全部仿体包块,包括最小2 mm者,回读时低位移区对应相对高硬度包块
- 硬包块模拟恶性肿瘤,其杨氏模量为22、31、44和56 kPa,背景为5.3 kPa,位移与包块-背景模量比呈反比
临床应用与阅片要点
- 操作方式类似常规乳腺超声手持扫查,适合对肿瘤全容积进行硬度分布的补充评估
- 阅读三维位移图时可将低位移区域视为相对更硬区域,协助定位可疑病变范围
局限性
- 人体乳腺病例仅5例,样本量很小,且包含不同类型病灶,结果仅为初步临床可行性验证
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: Harmonic motion imaging (HMI) is a displacement-based ultrasound elastography that simultaneously induces and tracks oscillatory on-axis motion using focused ultrasound and imaging transducers. Single transducer (ST)-HMI was recently developed to perform harmonic excitation and tracking using 1-D imaging arrays and electronic steering to exclude redundant raster-scanning. However, ST-HMI is limited to 2D imaging. This study develops 3-D-HMI with a handheld system.
METHODS: Harmonic radiation force (200 or 400 Hz) was generated using row or column elements of a row-column array (RCA), with resulting displacements estimated using the same elements for focused transmission and the remaining orthogonal elements in receive, interleaved between excitation pulses.
RESULTS: The proposed method was first validated in phantoms with stiff (Young's moduli: 22, 31, 44 and 56 kPa in 5.3-kPa background) or soft (5.1 kPa in 18.6-kPa background) inclusions (n = 12, diameters: 2.2-13.2 mm) mimicking malignant and benign tumors. The 3-D estimated normalized displacement detected all inclusions, including the smallest at 2 mm, and was found to be inversely proportional to the Young's modulus ratio of inclusions to background. 3-D HMI displacement maps of human breast tumors (n = 5) were reconstructed, including fibroadenoma (FA, n = 1) and invasive ductal carcinoma (IDC, n = 4), among which two were residual diseases after up to 12 wk of neoadjuvant chemotherapy.
CONCLUSION: 3-D-HMI was hereby described and found to be feasible, accurate and time-efficient volumetric mapping of relative stiffness. The proposed method increases the clinical potential of HMI to assess the mechanical properties of entire tumor volumes in a typical handheld manner, as in conventional breast sonography.