文献速递骨肌技术与安全MR

3D打印解剖模型优化骨科植入物旁膝关节MRI

Optimizing knee MRI near orthopedic hardware using a 3D-printed anatomical phantom.
PubMed 文献 · Skeletal Radiol · 2026-01-27

本研究开发了整合钛螺钉与不锈钢固定板的3D打印膝关节模型,在1.5T MRI上迭代优化T1w和STIR序列,并应用于3名无症状金属植入物志愿者。结果发现模型优化序列在空间分辨率、伪影减少和整体图像质量上均显著优于常规方案,且不增加采集时间。

要点

  • 骨科金属植入物旁MRI可通过模型优化序列,显著提升T1w和STIR图像质量且不延长扫描时间
  • 模型优化序列在空间分辨率、伪影减少和整体图像质量的评分提升幅度均具有统计学意义
  • 3D打印解剖模型为金属植入物条件下MRI序列调试提供了可重复、不依赖患者的平台

核心结论

  • 基于模型的序列优化为骨科植入物旁MRI协议调试提供了可重复且不依赖患者的框架

研究设计与样本

  • 探索性概念验证研究,开发3D打印膝关节模型,内含钛螺钉和不锈钢固定板
  • 模型成像在临床1.5T MRI系统上完成,T1w和STIR序列经模型迭代优化后与厂商默认及常规临床方案比较
  • 优化后的序列在3名无症状金属膝关节植入物志愿者中验证,由4名设盲阅片者采用5分Likert量表独立评估

关键影像表现/征象

  • 3D打印膝关节模型整合了钛螺钉和不锈钢固定板,用于模拟骨科植入物旁的金属条件

诊断标准与关键数值

  • 空间分辨率评分平均提升+0.70分,P<0.001
  • 伪影减少评分平均提升+0.65分,P<0.001
  • 整体图像质量评分平均提升+0.78分,P<0.001
  • T1w和STIR序列均获得改善,且采集时间未延长

临床应用与阅片要点

  • 3D打印解剖模型可用于在受控临床相关条件下开发、测试和标准化金属植入物旁MRI序列
  • 模型引导的序列优化可应用于临床协议调整,改善骨科植入物患者的图像质量

局限性

  • 探索性概念验证研究,样本量有限,志愿者仅3名

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To evaluate whether a dedicated, anatomically realistic knee phantom with orthopedic implants can be used to optimize clinical MRI sequences under realistic metallic conditions and to assess whether phantom-optimized sequences improve image quality compared with routine protocols.

MATERIALS AND METHODS: A 3D-printed knee phantom integrating a titanium screw and a stainless-steel fixation plate was developed. Phantom imaging was performed on a clinical 1.5 T MRI system. In this exploratory proof-of-concept study, T1w and STIR sequences were iteratively optimized using the phantom and compared with vendor-default and routine clinical protocols. The phantom-optimized sequences were subsequently applied in three asymptomatic volunteers with metallic knee implants and compared with routine sequences. Image quality was independently assessed by four blinded readers using a 5-point Likert scale across spatial resolution, artifact reduction, and overall image quality.

RESULTS: Phantom-optimized sequences achieved significantly higher image quality scores compared with routine protocols. Improvements were consistent across all three predefined criteria: spatial resolution (mean increase +0.70 points, p < 0.001), artifact reduction (+0.65, p < 0.001), and overall image quality (+0.78, p < 0.001). These gains were observed for both T1-weighted and STIR acquisitions without extending acquisition times.

CONCLUSION: Phantom-guided optimization provides a reproducible, patient-independent framework for tuning MRI protocols near orthopedic hardware. Anatomically realistic phantoms represent a promising methodological tool for developing, testing, and standardizing MRI sequences under controlled and clinically relevant conditions.

来源:Skeletal Radiol,查看原文。版权归原作者所有。

引用

  1. Tracqui, Yusuff, Hausser, et al. Optimizing knee MRI near orthopedic hardware using a 3D-printed anatomical phantom. Skeletal Radiol. 2026. PMID: 41591485. doi:10.1007/s00256-026-05136-z

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。