文献速递神经介入DSA

意大利介入神经放射诊断参考水平更新:纳入复杂度的全国调查

Establishing updated diagnostic reference levels for interventional radiology: a national Italian survey incorporating procedure complexity indices - Part II: interventional neuroradiology.
PubMed 文献 · Eur J Radiol · 2026-01-07

意大利国家卫生研究院协调20家医院,收集脑血管造影、脑动脉瘤栓塞和机械取栓的辐射剂量数据。通过非参数检验和广义线性回归识别影响剂量的因素,建立复杂度评分并对操作分层。提出按复杂度分层的全国诊断参考水平,强调定期更新DRL的必要性。

要点

  • 脑血管造影总DRL为191 Gy·cm²,复杂操作可高达198 Gy·cm²
  • 机械取栓总DRL为169 Gy·cm²,简单操作仅71 Gy·cm²
  • 报告辐射剂量时应参考操作复杂度分层而非单一数值

核心结论

  • 本研究通过复杂度分层方法建立了意大利介入神经放射学更新的全国诊断参考水平
  • 研究证明定期更新DRL对反映实践模式演变和技术进步至关重要

研究设计与样本

  • 多中心数据收集研究,纳入意大利20家医院
  • 涉及脑血管造影、脑动脉瘤栓塞和机械血栓切除术三类介入神经放射操作
  • 记录空气比释动能Ka,r、比释动能面积积PKA和透视时间FT
  • 统计学采用非参数检验和广义线性回归模型分析PKA分布
  • 通过对数转换构建复杂度评分系统对操作进行分类

关键影像表现/征象

  • 脑血管造影记录动脉评估数量、诊断优先级(择期或急诊)和研究类型
  • 脑动脉瘤栓塞记录动脉瘤状态、瘤颈比、大小、位置和形态
  • 机械取栓记录闭塞类型、血栓位置和是否使用支架取栓器
  • 同时登记患者年龄、性别和放射设备信息

诊断标准与关键数值

  • 脑血管造影新NDRL为191 Gy·cm²,简单71 Gy·cm²,中等117 Gy·cm²,复杂198 Gy·cm²
  • 脑动脉瘤栓塞新NDRL为252 Gy·cm²,简单136 Gy·cm²,中等174 Gy·cm²,复杂268 Gy·cm²
  • 机械取栓新NDRL为169 Gy·cm²,简单71 Gy·cm²,中等132 Gy·cm²,复杂169 Gy·cm²

临床应用与阅片要点

  • 介入神经放射辐射剂量管理应结合操作复杂度分层进行评价
  • 脑血管造影中动脉评估数量、急诊与否和研究类型影响辐射剂量
  • 脑动脉瘤栓塞中动脉瘤大小、位置、形态和瘤颈比与剂量相关
  • 机械取栓中闭塞类型、血栓位置和支架取栓器使用影响剂量水平

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

INTRODUCTION: To update the Italian Diagnostic Reference Levels (NDRLs) in interventional neuroradiology, the Istituto Superiore di Sanità coordinated a working group with consolidated expertise in this field. Establishing DRLs in interventional procedures is particularly challenging due to the complexity influenced by clinical and technical factors. For this reason, aim of this work is to provide DRLs that also take into account procedure complexity.

MATERIAL AND METHODS: Data were collected from 20 Italian hospitals concerning cerebral angiography(CA), embolization of cerebral aneurysms (ECA), and Mechanical Thrombectomy (MT). For each procedure, Air-Kerma (Ka,r), Kerma-Area-Product(PKA), and Fluoroscopy Time(FT) were recorded. Data related to patient age, sex, and radiological equipment were registered. Additionally, for CA number of arteries evaluated, diagnostic priority (elective or emergency), and study type were documented. For ECA, aneurysm status, dome-to-neck ratio, aneurysm size, aneurysm location, and aneurysm morphology were recorded. For MT, collected data included occlusion type, thrombus location, and use of a stent retriever.

RESULTS: PKA distributions were analysed with non-parametric tests and generalized linear regression models to identify clinical and technical factors affecting dosimetry. Log transformations helped develop a complexity score to classify procedures. The new NDRLs are 191 Gy.cm2 for CA (71 Gy.cm2 for simple, 117 Gy.cm2 for medium, 198 Gy.cm2 for complex), 252 Gy.cm2 for ECA (136 Gy.cm2 for simple, 174 Gy.cm2 medium, 268 Gy.cm2 complex) and 169 Gy.cm2 for MT (71 Gy.cm2 for simple, 132 Gy.cm2 for medium, 169 Gy.cm2 for complex).

CONCLUSION: This study establishes updated national diagnostic reference levels for interventional neuroradiology in Italy through a comprehensive complexity-stratified approach. The research demonstrates the critical importance of regular DRL updates to reflect evolving practice patterns and technological advances, offering a more nuanced framework that recognizes the inherent variability of neuro interventional procedures across different complexity tiers.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Cavallari, D'Ercole, Grande, et al. Establishing updated diagnostic reference levels for interventional radiology: a national Italian survey incorporating procedure complexity indices - Part II: interventional neuroradiology. Eur J Radiol. 2026. PMID: 41558397. doi:10.1016/j.ejrad.2026.112662

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。