肥胖患者光子计数CT与能量积分CT的辐射剂量及图像质量比较
本回顾性研究纳入51例肥胖患者,分别接受光子计数CT与能量积分CT胸腹部扫描,进行个体内比较。结果显示PCCT在血管、实质脏器及肌肉中信号更高,主观图像质量更优,且CTDIvol和DLP均低于EID CT。结论为PCCT在肥胖人群中以更低辐射剂量获得更优图像质量。
要点
- PCCT在肥胖患者中辐射剂量更低,可优先考虑用于需反复随访的胸腹部CT检查。
- PCCT主观图像质量、对比度及锐利度均优于EID CT,有利于提高肥胖患者诊断信心。
- 图像噪声在两种CT间相近,阅片时不应以噪声作为PCCT质量优势的主要判读依据。
核心结论
- 在肥胖患者中,个体内比较显示PCCT在更低辐射剂量下获得更优的定量与主观图像质量。
研究设计与样本
- 采用配对双侧t检验或Wilcoxon符号秩检验进行组间比较。
关键影像表现/征象
- PCCT在血管、实质脏器及肌肉中信号高于EID CT,皮下脂肪CT值更负,图像噪声相近。
- 升主动脉SNR PCCT为19.65±6.47,EID CT为13.22±4.83;肝脏SNR PCCT为5.53±1.69,EID CT为4.71±1.89。
- 两位阅片者均将PCCT的总体图像质量、对比度与锐利度评为更高,伪影相关干扰更低。
临床应用与阅片要点
- 肥胖患者行PCCT检查时可降低辐射剂量而不牺牲图像质量,适合纳入常规扫描方案评估。
- 报告时对PCCT图像的主观噪声评价与EID CT相近,不应将噪声列为PCCT不足。
- 对比度与锐利度的改善提示PCCT有助于肥胖患者血管与实质脏器细节的显示。
局限性
- 回顾性研究设计,存在选择偏倚可能。
- 样本量为51例,相对较小。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Background Maintaining diagnostic CT image quality in patients with obesity remains challenging because increased x-ray attenuation is often offset by higher radiation exposure. Photon-counting CT (PCCT) has demonstrated improved dose efficiency and image quality, but intraindividual data in this population are limited. Purpose To compare radiation dose, quantitative image quality, and subjective image quality between PCCT and energy-integrating detector (EID) CT intraindividually in patients with obesity. Materials and Methods This retrospective study included patients with obesity undergoing thoracic and/or abdominal CT at both PCCT and EID CT between December 1, 2022, and January 31, 2025. Signal, image noise, and signal to noise ratio (SNR) were measured in vascular, parenchymal, muscular, and adipose tissue. Two radiologists independently rated image quality on a six-point Likert scale. Radiation dose parameters were volume CT dose index (CTDIvol), dose-length product (DLP), and effective dose. The paired two-sided t test or Wilcoxon signed-rank test was used for group comparisons. Results This study included 51 patients with obesity (median age, 56 years [IQR, 51-66.5 years]; 33 male). Signal was higher with PCCT than EID CT in vascular structures, parenchymal organs, and muscle and more negative in subcutaneous fat (all P < .001), with similar image noise (P = .09-.84). SNR was higher with PCCT in vascular structures (ascending aorta: mean, 19.65 ± 6.47 [SD] vs 13.22 ± 4.83; P < .001) and most parenchymal organs (liver: 5.53 ± 1.69 vs 4.71 ± 1.89; P = .02). The difference in SNR increased with body mass index for vessels (r = 0.36) and parenchymal organs (r = 0.51) (both P < .001). Both readers rated overall image quality, contrast, and sharpness higher and artifact-related impairment lower for PCCT (all P < .001), whereas subjective image noise was similar (P = .16 and .14). CTDIvol (median, 11.99 vs 17.05 mGy) and DLP (median, 526.25 vs 709.32 mGy · cm) were lower with PCCT (both P < .001). Conclusion In patients with obesity, intraindividual comparisons demonstrated that PCCT had superior quantitative and subjective image quality at a lower radiation dose than EID CT. © RSNA, 2026 Supplemental material is available for this article.
引用
- Hegmanns, Zensen, Rosok, et al. Intraindividual Comparison of Radiation Dose and Image Quality: Photon-counting versus Energy-integrating Detector CT in Obesity. Radiology. 2026. PMID: 42899938. doi:10.1148/radiol.260573
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。