TGSE-BLADE DWI用于颅内肿瘤术后早期MRI评估
本研究回顾性纳入33例颅内肿瘤切除术后48小时内行1.5T MRI的患者,比较RESOLVE DWI与TGSE-BLADE DWI的图像质量、伪影及对切除腔测量的准确性。结果显示TGSE-BLADE DWI在切除区附近几何变形和磁敏感伪影显著减轻,图像质量和诊断信心更优;其切除腔测量与T1加权像接近一致,而RESOLVE DWI存在显著差异。该技术适合作为术后早期检测围手术期缺血改变的可靠DWI替代方案。
要点
- 术后早期颅内积气明显时,可优先选择TGSE-BLADE DWI以减少切除区附近变形和磁敏感伪影。
- TGSE-BLADE DWI对切除腔大小的测量与T1加权像一致性更好,报告测量时可优先参考。
- RESOLVE DWI信噪比和对比噪声比略高,但几何变形可导致切除腔测量明显偏差。
核心结论
- TGSE-BLADE DWI可显著减少颅内术后成像中的伪影和几何变形,提高围手术期缺血改变的检出能力。
- 在B0不均匀性明显的术后环境中,TGSE-BLADE DWI可作为传统EPI类DWI的有效替代序列。
研究设计与样本
- 回顾性、机构审查委员会批准的研究,共纳入33例患者,平均年龄52.3±25.1岁。
- 患者均在颅内肿瘤切除术后48小时内接受1.5T MRI检查。
- RESOLVE DWI与TGSE-BLADE DWI采用相同扫描方位和5mm层厚采集。
- 由两名神经放射科医师盲法评估几何变形、磁敏感伪影、整体图像质量和诊断信心。
- 定量分析包括信噪比、对比噪声比、ADC值、图像质量指标及以T1加权像为参照的切除缺损尺寸测量。
关键影像表现/征象
- TGSE-BLADE DWI在切除区附近显示几何变形和磁敏感伪影明显少于RESOLVE DWI。
- TGSE-BLADE DWI对切除缺损的尺寸测量与T1加权像接近,RESOLVE DWI则呈明显偏差。
诊断标准与关键数值
- TGSE-BLADE DWI的定性图像质量和诊断信心显著优于RESOLVE DWI,p<0.001。
- TGSE-BLADE DWI切除缺损测量与T1加权像相比差异无统计学意义,p=0.974。
- RESOLVE DWI切除缺损测量与T1加权像相比差异有统计学意义,p<0.001。
- RESOLVE DWI的信噪比和对比噪声比略高于TGSE-BLADE DWI。
临床应用与阅片要点
- 术后48小时内需要判断有无急性梗死与残留/强化肿瘤鉴别时,可加扫TGSE-BLADE DWI。
- 因颅内积气或手术区B0不均匀导致常规DWI明显变形时,TGSE-BLADE DWI可提供更可靠图像。
- 测量切除腔大小时应注意RESOLVE DWI的几何变形可能导致高估或低估。
局限性
- 本研究为回顾性设计,样本量为33例,可能存在选择偏倚。
- 研究为单中心1.5T设备结果,未涉及3T及其他厂商平台的验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: Postoperative MRI plays a pivotal role in assessing residual disease and planning therapy after intracranial tumor resection. As acute perioperative infarction may mimic enhancing and non-enhancing tumor tissue in follow-up imaging, the initial postoperative MRI examination is essential for this purpose. However, conventional DWI techniques like RESOLVE are often impaired by artifacts from intracranial air. This study compares RESOLVE DWI with TGSE-BLADE DWI within early postoperative scans to improve diagnostic confidence in identifying ischemic changes.
METHOD: In a retrospective, institutional review board-approved study, 33 patients (mean age 52.3 ± 25.1 years) underwent a 1.5-Tesla MRI within 48 h of resection. RESOLVE and TGSE-BLADE DWI sequences were acquired in the same orientation and slice thickness of 5 mm. Two neuroradiologists blindly evaluated geometric distortion, susceptibility artifacts, overall image quality, and diagnostic confidence. Quantitative analysis included measuring signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), apparent diffusion coefficient (ADC) values, image quality metrics and resection defect dimensions against T1-weighted images.
RESULTS: TGSE-BLADE DWI exhibited significantly reduced geometric distortions and susceptibility artifacts near the resection site compared with RESOLVE DWI, producing superior qualitative image quality and diagnostic confidence (p < 0.001). Conversely, RESOLVE DWI yielded slightly higher SNR and CNR. Measurements of the resection defect on TGSE-BLADE DWI closely matched those on T1-weighted images (p = 0.974), while RESOLVE DWI showed a significant discrepancy (p < 0.001).
CONCLUSIONS: TGSE-BLADE DWI significantly minimizes artifacts and geometric distortions in intracranial postoperative imaging, thereby improving the detection of perioperative ischemic changes. This robust technique serves as an effective alternative to conventional EPI-based sequences in challenging settings with pronounced B0 inhomogeneities, ultimately enhancing diagnostic confidence and patient care.
引用
- Ruff, Hauser, Bombach, et al. A comprehensive quantitative and qualitative assessment of TGSE-BLADE DWI in postoperative imaging following intracranial tumor resection. Eur J Radiol. 2026. PMID: 41534273. doi:10.1016/j.ejrad.2026.112659
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。