光子计数CT在不同辐射剂量下骨微结构定量与HR-pQCT对比
该研究对离体桡骨远端/胫骨标本及1例骨质疏松女性在体扫描,比较PCD-CT在2.5、5、10、20 mGy与HR-pQCT的小梁微结构测量。结果显示PCD-CT与HR-pQCT在桡骨远端和胫骨的骨小梁微结构参数呈强相关,且在低、临床可接受辐射剂量下即可实现。方法包括三维配准后对应立方体区域的参数对比。
要点
- PCD-CT在2.5 mGy低剂量下即可与HR-pQCT测量保持强相关。
- PCD-CT测得Tb.BV/TV、Tb.N、Tb.Th偏高,Tb.Sp及Tb.1/N.SD偏低。
- 模态间微结构值差异可能与图像分析参数设置有关。
核心结论
- PCD-CT在低且临床可接受的辐射剂量下,桡骨远端与胫骨小梁微结构测量与HR-pQCT呈强相关。
研究设计与样本
- 研究类型为离体标本与在体病例的方法对比研究;离体部分纳入1例88岁男性捐献者的2个完整腕关节和2个完整踝关节标本。
- 离体标本分别在PCD-CT 2.5、5、10、20 mGy剂量及HR-pQCT下扫描。在体部分纳入1例40岁骨质疏松女性,完成桡骨和胫骨的HR-pQCT与PCD-CT扫描。
- 对骨分割后的PCD-CT与HR-pQCT图像进行三维配准,在对应位置分别取边长6 mm(离体)和5 mm(在体)的立方体区域进行参数比较。
关键影像表现/征象
- 在桡骨远端和胫骨的对应立方体区域,PCD-CT与HR-pQCT对小梁微结构参数可进行三维配准后定量比较。
诊断标准与关键数值
- 离体标本中,PCD-CT与HR-pQCT的线性相关系数R²在2.5 mGy时为0.85-0.97,除Tb.1/N.SD外,随辐射剂量增至20 mGy仍保持稳定。
- Tb.1/N.SD的R²在2.5至5 mGy间有所升高,5 mGy以上各剂量保持稳定。
- 在每个辐射剂量下,PCD-CT测得Tb.BV/TV、Tb.N、Tb.Th高于HR-pQCT,Tb.Sp和Tb.1/N.SD低于HR-pQCT。
- 在体比较中,PCD-CT与HR-pQCT的R²在桡骨为0.89-0.95,在胫骨为0.82-0.97。
临床应用与阅片要点
- PCD-CT低剂量骨小梁微结构评估在日常阅片中可用于骨质疏松相关骨质量评价时的技术选择参考。
- 解读PCD-CT骨微结构参数时需注意其绝对值与HR-pQCT存在系统差异,不宜直接沿用HR-pQCT参考值。
局限性
- 离体和在体样本量均小,离体仅来自1例供体,在体仅1例患者,结果外推受限。
- 研究未报告与临床骨折风险或骨强度直接对应的诊断阈值,模态间差异仅提示可能相关。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To compare trabecular microarchitecture measurements at the distal radius and tibia from photon-counting detector CT (PCD-CT) at varying radiation doses with high-resolution peripheral quantitative CT (HR-pQCT).
METHODS: Two intact wrist and two intact ankle specimens from an 88-year-old man were scanned with PCD-CT at radiation doses of 2.5, 5, 10, and 20 mGy and with HR-pQCT. Additionally, clinical in-vivo HR-pQCT and PCD-CT scans at the radius and tibia were acquired of a 40-year-old woman with osteoporosis. After bone segmentation, the segmented PCD-CT and HR-pQCT scans were three-dimensionally registered. Cubic volumes (edge length: ex-vivo 6 mm, in-vivo 5 mm) were defined at corresponding locations in the PCD-CT and HR-pQCT scans based on the three-dimensional registration. For each cube, trabecular volume fraction (Tb.BV/TV), thickness (Tb.Th), number (Tb.N), separation (Tb.Sp), and heterogeneity (Tb.1/N.SD) were quantified and compared between corresponding PCD-CT and HR-pQCT cubes.
RESULTS: Ex-vivo, linear correlation coefficients (R2) between PCD-CT and HR-pQCT were 0.85-0.97 at 2.5 mGy and remained stable with increasing radiation dose for all parameters except Tb.1/N.SD. For Tb.1/N.SD, R2 increased between 2.5 and 5 mGy and remained stable at higher doses. At each radiation dose, Tb.BV/TV, Tb.N, and Tb.Th values were higher and Tb.Sp and Tb.1/N.SD lower on PCD-CT than on HR-pQCT. In-vivo, R2 was 0.89-0.95 (radius) and 0.82-0.97 (tibia).
CONCLUSIONS: PCD-CT strongly correlated with HR-pQCT in trabecular microarchitecture measurements at the distal radius and tibia at low, clinically acceptable, radiation dose. Between-modality differences in microarchitecture values are likely related to chosen image analysis settings.