最新一代锥形束CT用于胸部成像的可行性初步研究

Feasibility of Latest-Generation Cone-Beam CT for Thoracic Imaging: A Pilot Study.
PubMed 文献 · Acad Radiol · 2026-01-10

本研究回顾性纳入88例吸烟者,采用最新一代CBCT行胸部扫描,其中低剂量组40例、标准剂量组48例。低剂量组中位累积DLP为42.8 mGy·cm,较标准剂量组107.7 mGy·cm降低2.5倍。两组诊断信心及观察者间一致性相当,提示低剂量CBCT有望用于肺癌筛查场景。

要点

  • 低剂量CBCT较标准剂量辐射降低2.5倍,诊断信心无显著下降
  • 肺结节总体检出率为41%,观察者间一致性κ=0.73–0.83
  • 低剂量组空间分辨率略低于标准剂量组,阅片时需注意

核心结论

  • 低剂量CBCT在保证诊断信心和观察者间可靠性的前提下可显著降低辐射剂量
  • CBCT有望成为低剂量CT的有效替代或补充手段用于预防性肺部影像项目

研究设计与样本

  • 回顾性研究,纳入88例当前或既往吸烟者,单中心预防保健中心完成
  • 使用NewTom 7 G双能CBCT系统,标准剂量组48例,低剂量组40例
  • 两名认证放射科医师独立评估肺内发现和图像质量,统计采用混合效应模型

关键影像表现/征象

  • 肺结节总体检出率为41%
  • 低剂量组空间分辨率略低于标准剂量组,OR=0.51,p=0.04

诊断标准与关键数值

  • 图像质量评为良好或优秀的比例:低剂量组85%,标准剂量组83%,RD=+2% [IC95% -12; +16],RR=1.03
  • 诊断信心高的比例:低剂量组82%,标准剂量组83%,RD=-1% [-14; +12],RR=0.99
  • 低剂量组中位累积DLP为42.8 mGy·cm,标准剂量组为107.7 mGy·cm,p<0.001
  • 肺结节检出观察者间一致性:κ=0.73–0.83

临床应用与阅片要点

  • 低剂量CBCT可用于吸烟人群胸部筛查,辐射仅为标准剂量CBCT的约40%
  • 阅片时需注意低剂量方案下空间分辨率有所下降,可能影响细微结构评估
  • 两名医师独立阅片一致性较高,提示该技术在不同阅片者间可重复

局限性

  • 回顾性研究设计
  • 未与低剂量CT进行直接前瞻性对比

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: Lung cancer remains the leading cause of cancer-related death worldwide. Low-dose computed tomography (LDCT) is the current gold standard for screening, but radiation exposure remains a concern. Cone-Beam Computed Tomography (CBCT) offers ultra-low-dose acquisition but has not yet been tested for thoracic screening.

PURPOSE: To evaluate the feasibility of using a latest-generation CBCT system for thoracic imaging in a lung cancer screening context.

MATERIALS AND METHODS: 88 patients, current or former smokers, were retrospectively included. They underwent thoracic CBCT (7 G Dual Energy system, NewTom, Italy) between July 2024 and April 2025 at a preventive health center, with either a standard-dose protocol (n = 48) or a low-dose protocol (n = 40). Two board-certified radiologists independently assessed pulmonary findings and image quality. Radiation doses were compared. Statistical analysis included mixed-effects models and interobserver agreement.

RESULTS: The radiation exposure was significantly lower in the low-dose group, with a median cumulative DLP of 42.8 mGy·cm, compared to 107.7 mGy·cm, representing a 2.5-fold reduction in radiation (p<0.001). Nodule detection occurred in 41% of patients overall. Overall image quality was rated good/excellent in 85% (low-dose) vs 83% (standard-dose); Risk Difference (RD) = +2% [IC95% -12; +16], Risk Ratio (RR) = 1.03. Diagnostic confidence was high in 82% vs 83%; RD = -1% [-14; +12], RR = 0.99. Spatial resolution was slightly lower in the low-dose group (OR = 0.51, p = 0.04). Interobserver agreement for nodule detection was substantial across both groups (κ = 0.73-0.83).

CONCLUSION: Low-dose CBCT offers substantial radiation dose reduction without compromising diagnostic confidence or interobserver reliability. CBCT may represent a promising, cost-effective alternative or complement to low-dose CT in preventive lung imaging programs, but prospective studies comparing CBCT to LDCT are warranted.

来源:Acad Radiol,查看原文。版权归原作者所有。