缩短透视时间警报对儿童改良钡餐检查辐射剂量与透视时间的影响

Effect of shortening the fluoroscopy time audible alarm on total fluoroscopy time and reference air kerma for pediatric modified barium swallow studies.
PubMed 文献 · Pediatr Radiol · 2026-01-08

本研究回顾性分析3875例18岁以下儿童改良钡餐检查,比较透视时间警报从5分钟改为2分钟前后的参考点空气比释动能和透视时间。结果显示主院区及卫星点平均与中位参考点空气比释动能、透视时间均显著下降。研究提示缩短警报设置可降低儿童辐射暴露,建议法规考虑灵活设置。

要点

  • 透视警报从5分钟缩至2分钟后,儿童改良钡餐检查辐射剂量与透视时间显著降低。
  • 主院区平均参考点空气比释动能下降14.9%,平均透视时间下降15.5%。
  • 儿童透视检查中可考虑采用更短时间警报以控制辐射剂量。

核心结论

  • 缩短透视时间警报可显著降低儿童改良钡餐检查的参考点空气比释动能和透视时间。
  • 研究建议州和联邦法规考虑建立灵活的透视警报设置,尤其针对儿童人群。

研究设计与样本

  • 回顾性分析研究,纳入3875例18岁以下患者的改良钡餐检查。
  • 比较透视时间警报从每5分钟改为每2分钟前后的参考点空气比释动能和透视时间。
  • 数据来源于主院区及卫星点,研究未报告设备、序列或参考标准细节。

诊断标准与关键数值

  • 主院区平均参考点空气比释动能下降14.9%,中位值下降19.6%,P<0.001。
  • 主院区平均透视时间下降15.5%,中位透视时间下降16.2%,P<0.001。
  • 卫星点平均参考点空气比释动能下降14.3%,中位值下降9.1%,P<0.001。
  • 卫星点平均透视时间下降13.9%,中位透视时间下降8.3%,P<0.001。

临床应用与阅片要点

  • 在日常儿童改良钡餐检查中,可采用短于5分钟的透视时间警报以强化时间意识。
  • 读片或操作时应注意累积透视时间警报是控制儿童辐射剂量的可行干预手段。

局限性

  • 研究为回顾性设计,未报告机构具体设备、操作者差异及病例适应证分布。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: Federal regulations require an audible alarm for each passage of 5 min of cumulative x-ray irradiation time during a diagnostic fluoroscopy procedure. In contrast to United States federal requirements, regulations vary amongst individual states.

OBJECTIVE: Determine whether a shortened fluoroscopy time alarm reduces pediatric patient ionizing radiation exposure for modified barium swallow studies (MBSS).

MATERIALS AND METHODS: Fluoroscopy time alarms were changed from the passage of every 5 min to 2 min. A retrospective analysis of MBSS compared patient reference point air kerma and fluoroscopy time before and after the alarm time was shortened.

RESULTS: A total of 3,875 MBSS performed on patients under the age of 18 years were analyzed. Reductions of 14.9% and 19.6% for average and median reference point air kerma (P<0.001) and 15.5% and 16.2% for average and median fluoroscopy time (P<0.001) were observed at the main campus for MBSS performed before and after changing the fluoroscopy time alarm from 5 min to 2 min. Reductions of 14.3% and 9.1% for average and median reference point air kerma (P<0.001) and 13.9% and 8.3% for average and median fluoroscopy time (P<0.001) were observed at the satellite locations for MBSS performed before and after changing the fluoroscopy time alarm.

CONCLUSION: This retrospective study suggests a shortened fluoroscopy time alarm can significantly lower pediatric patient dose and fluoroscopy time in MBSS. State and federal regulations should consider establishing flexible fluoroscopic alarm settings, especially when used for the pediatric population.

来源:Pediatr Radiol,查看原文。版权归原作者所有。