文献速递神经介入DSAMR

优化TES-MEP在脑动脉瘤血管内治疗中的监测价值

Clinical impact of transcranial unilateral motor evoked potentials monitoring in cerebral aneurysm endovascular treatment.
PubMed 文献 · Neuroradiology · 2026-01-08

回顾性分析62个接受血管内弹簧圈栓塞的脑动脉瘤病例,在全静脉麻醉下采用优化单侧经颅电刺激运动诱发电位方案。该方案在所有病例中均成功获得MEP,未出现因患者体动中断手术的情况,而传统C3(±)-C4(∓)刺激在29.3%病例中引起不可接受的体动。术中2例出现MEP波幅下降,均与缺血事件相关并促成及时干预。

要点

  • 传统C3(±)-C4(∓)刺激导致29.3%病例出现不可接受体动,需警惕DSA配准错误风险
  • 优化单侧刺激方案可在不中断手术的前提下稳定获得MEP,避免重复DSA成像
  • 术中MEP波幅下降提示缺血事件,应作为血管内治疗中实时神经功能监测的核心观察指标

核心结论

  • 优化TES-MEP方案在神经血管内操作中可行且安全,减少干扰性体动,同时保持缺血检测能力
  • 该方法改善工作流程安全性,对提高患者预后具有潜在价值,但受回顾性设计限制,需前瞻性研究验证

研究设计与样本

  • 回顾性研究,纳入62个接受血管内弹簧圈栓塞治疗的脑动脉瘤病例,均在全身静脉麻醉下进行
  • 采用多种TES-MEP刺激导联组合:C3(±)-C4(∓)、C1/C2(+)-C4/C3(-)、C4/C3(+)-C1/C2(-)或C1(±)-C2(∓),专门优化以诱出单侧MEP并减少体动

关键影像表现/征象

  • 传统C3(±)-C4(∓)刺激在29.3%病例中引起不可接受的体动,导致DSA配准错误并需重复成像
  • 优化单侧导联方案在所有病例中成功获得稳定MEP,无手术因体动中断
  • 术中2例患者出现MEP波幅下降,均与缺血事件相关

诊断标准与关键数值

  • 传统C3(±)-C4(∓)刺激引起不可接受体动的发生率为29.3%

临床应用与阅片要点

  • 神经血管内治疗中宜优先采用优化单侧TES-MEP导联组合,减少体动以保障DSA配准质量
  • 术中应持续观察MEP波幅变化,波幅下降需立即结合DSA评估有无血管闭塞或栓塞相关缺血
  • 使用传统C3(±)-C4(∓)导联时需警惕体动导致的图像配准错误和重复曝光

局限性

  • 回顾性研究设计存在固有局限
  • 需要前瞻性研究验证结果并进一步优化刺激方案

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: Neurosurgery commonly employs motor evoked potential (MEP) monitoring to assess intraoperative motor function. However, its utility in neuroendovascular procedures has been sporadically reported, widespread adoption remains limited, largely due to challenges associated with stimulus-induced patient movement and subsequent image misregistration. This study aimed to address these limitations by evaluating the feasibility of an optimized transcranial electrical stimulation MEP (TES-MEP) protocol designed to minimize body movement while preserving monitoring reliability. METHODS: We performed a retrospective review of 62 aneurysms treated with endovascular coil embolization for cerebral aneurysms under total intravenous anesthesia (TIVA). We employed various TES-MEP montages, C3(±)-C4(∓), C1/C2(+)-C4/C3(-), C4/C3(+)-C1/C2(-), or C1(±)-C2(∓), specifically optimized to elicit a unilateral MEP and minimize body movement. We assessed stimulus-induced movement, the quality of MEP recordings, and intraoperative MEP changes. RESULTS: MEPs were successfully obtained in all cases, with no procedures interrupted by patient movement. In contrast, conventional C3(±)-C4(∓) stimulation caused unacceptable movement in 29.3% of cases. Our modified montage facilitated stable monitoring, eliminating digital subtraction angiography (DSA) misregistration and the need for repeated imaging. Intraoperative MEP amplitude reductions occurred in two patients, both linked to ischemic events and prompting timely intervention. CONCLUSIONS: Our TES-MEP monitoring method is feasible and safe in neuroendovascular procedures. By reducing procedure-disrupting movement while maintaining ischemia detection, the protocol improves workflow safety and holds meaningful potential to enhance patient outcomes. Given its retrospective design, the study is subject to inherent limitations, and prospective studies are warranted to validate these findings and further refine stimulation protocols.

来源:Neuroradiology,查看原文。版权归原作者所有。

引用

  1. Kamada, Yamada, Kim, et al. Clinical impact of transcranial unilateral motor evoked potentials monitoring in cerebral aneurysm endovascular treatment. Neuroradiology. 2026. PMID: 41504788. doi:10.1007/s00234-025-03890-w

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。