光谱光子计数CT冠状动脉支架评估的首例人体CT血管成像结果
该前瞻性研究纳入8例患者共16枚冠状动脉支架,对比双层探测器能量积分CT(EID-DLCT)与光谱光子计数CT(SPCCT)的体内支架成像质量。SPCCT辐射剂量更低(固定CTDI 25.7 mGy vs 中位35.7 mGy,P=0.02),支架外径更小、内径更大,开花伪影减少,支架内腔、支架结构、周围钙化及束硬化伪影的主观评分均更高(P≤0.05)。
要点
- SPCCT较EID-DLCT显著减少支架开花伪影,支架内径测量更大、外径更小(P<0.05)。
- SPCCT在支架内腔、支架结构、周围钙化及束硬化伪影的主观评分均优于EID-DLCT(P≤0.05)。
- SPCCT-Sharp重建对支架结构和束硬化评估优于SPCCT-Detailed 2,且辐射剂量更低(P=0.02)。
关键影像表现/征象
- SPCCT上支架外径测量值更小、内径更大(均P<0.05),开花伪影减少(P<0.05)。
- 冠状动脉内CT值两扫描仪相似,但Δ S-C在SPCCT-Sharp上低于EID-DLCT-XCD和SPCCT-Detailed 2(P<0.05)。
- SPCCT对支架内腔、支架结构、周围钙化和束硬化伪影的主观评分均高于EID-DLCT(均P≤0.05)。
- SPCCT-Sharp在支架结构和束硬化评估方面主观评分优于SPCCT-Detailed 2。
诊断标准与关键数值
- 主观图像质量采用4分法评估冠状动脉支架前血管壁、支架内腔、支架结构、支架周围钙化及束硬化伪影。
- 客观开花伪影通过测量支架近端和远端的外径与内径进行量化,并计算支架内腔与上游冠状动脉的CT值差值(Δ S-C)。
- 辐射剂量:SPCCT固定CT剂量指数25.7 mGy,EID-DLCT中位CT剂量指数35.7 mGy(IQR 13.6),P=0.02。
临床应用与阅片要点
- SPCCT可减少支架开花伪影,有助于更准确评估支架内腔和支架结构,但需注意5枚支架因运动伪影在SPCCT上不可评估。
- SPCCT-Sharp重建在评估支架结构和束硬化伪影方面可能优于Detailed 2重建,阅片时可优先选择。
局限性
- 样本量小(8例患者,16枚支架),5枚支架因运动伪影在SPCCT上不可评估,1枚支架内径在EID-DLCT上无法评估,1枚支架信息缺失。
- 单中心前瞻性研究,所有患者均为男性,结论需更大样本验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: The aim of this study is to compare the image quality of in vivo coronary stents between an energy integrating detectors dual-layer computed tomography (EID-DLCT) and a clinical prototype of spectral photon counting computed tomography (SPCCT).
MATERIALS AND METHODS: In January to June 2021, consecutive patients with coronary stents were prospectively enrolled to undergo a coronary computed tomography (CT) with an EID-DLCT (IQon, Philips) and an SPCCT (Philips). The study was approved by the local ethical committee and patients signed an informed consent. A retrospectively electrocardiogram-gated acquisition was performed with optimized matching parameters on the 2 scanners (EID-DLCT: collimation, 64 × 0.625 mm; kVp, 120, automatic exposure control with target current at 255 mAs; rotation time, 0.27 seconds; SPCCT: collimation, 64 × 0.275 mm; kVp, 120; mAs, 255; rotation time, 0.33 seconds). The injection protocol was the same on both scanners: 65 to 75 mL of Iomeron (Bracco) at 5 mL/s. Images were reconstructed with slice thickness of 0.67 mm, 512 matrix, XCB (Xres cardiac standard) and XCD (Xres cardiac detailed) kernel, iDose 3 for EID-DLCT and 0.25-mm slice thickness, 1024 matrix, Detailed 2 and Sharp kernel, and iDose 6 for SPCCT. Two experienced observers measured the proximal and distal external and internal diameters of the stents to quantify blooming artifacts. Regions of interest were drawn in the lumen of the stent and of the upstream coronary artery. The difference (Δ S-C) between the respective attenuation values was calculated as a quantification of stent-induced artifacts on intrastent image quality. For subjective image quality, 3 experienced observers graded with a 4-point scale the image quality of different parameters: coronary wall before the stent, stent lumen, stent structure, calcifications surrounding the stent, and beam-hardening artifacts.
RESULTS: Eight patients (age, 68 years [interquartile range, 8]; all men; body mass index, 26.2 kg/m2 [interquartile range, 4.2]) with 16 stents were scanned. Five stents were not evaluable owing to motion artifacts on the SPCCT. Of the remaining, all were drug eluting stents, of which 6 were platinum-chromium, 3 were cobalt-platinum-iridium, and 1 was stainless steel. For 1 stent, no information could be retrieved. Radiation dose was lower with the SPCCT (fixed CT dose index of 25.7 mGy for SPCCT vs median CT dose index of 35.7 [IQ = 13.6] mGy; P = 0.02). For 1 stent, the internal diameter was not assessable on EID-DLCT. External diameters were smaller and internal diameters were larger with SPCCT (all P < 0.05). Consequently, blooming artifacts were reduced on SPCCT (P < 0.05). Whereas Hounsfield unit values within the coronary arteries on the 2 scanners were similar, the Δ S-C was lower for SPCCT-Sharp as compared with EID-DLCT-XCD and SPCCT-Detailed 2 (P < 0.05). The SPCCT received higher subjective scores than EID-DLCT for stent lumen, stent structure, surrounding calcifications and beam-hardening for both Detailed 2 and Sharp (all P ≤ 0.05). The SPCCT-Sharp was judged better for stent structure and beam-hardening assessment as compared with SPCCT-Detailed 2.
CONCLUSION: Spectral photon counting CT demonstrated improved objective and subjective image quality as compared with EID-DLCT for the evaluation of coronary stents even with a reduced radiation dose.
引用
- Boccalini S, Si-Mohamed SA, Lacombe H, et al. First In-Human Results of Computed Tomography Angiography for Coronary Stent Assessment With a Spectral Photon Counting Computed Tomography. Invest Radiol. 2022. PMID: 34711766. doi:10.1097/RLI.0000000000000835
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。