文献速递儿科影像骨肌CT

儿童创伤患者胸腰椎CT辐射剂量水平及扫描参数优化评估

Radiation dose levels of thoracic-lumbar spine CT in pediatric trauma patients and assessment of scan parameters for dose optimization.
PubMed 文献 · Pediatr Radiol · 2021-10-14

该回顾性研究纳入192例接受脊柱CT的儿童创伤患者,按年龄分为0-10岁和11-17岁两组,记录扫描参数与辐射剂量指标。结果显示采用低管电压和高准直层宽可显著降低胸腰段及腰椎CT的CTDIvol和DLP值,提示在保持图像质量前提下可大幅降低辐射剂量。

要点

  • 儿童脊柱CT可采用低管电压和高准直层宽优化剂量
  • 0-10岁组胸腰段和11-17岁组腰椎CT剂量降低显著
  • 优化参数后辐射剂量可低于厂家默认协议

诊断标准与关键数值

  • 0-10岁组胸腰段和11-17岁组腰椎CT采用低管电压时CTDIvol和DLP显著更低(P<0.05)。
  • 两组胸腰段和11-17岁组腰椎CT采用高准直层宽时CTDIvol和DLP显著低于低准直层宽(P<0.05)。

临床应用与阅片要点

  • 儿童脊柱CT扫描应优先考虑低管电压和高准直层宽以降低剂量。
  • 需根据年龄和扫描部位调整参数,避免直接使用厂家默认协议。

局限性

  • 回顾性研究设计。
  • 单中心研究。
  • 未报告图像质量客观评价指标。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: CT is frequently used for assessing spinal trauma in children.

OBJECTIVE: To establish the local diagnostic reference levels of spine CT examinations in pediatric spinal trauma patients and analyze scan parameters to enable dose optimization.

MATERIALS AND METHODS: In this retrospective study, we included 192 pediatric spinal trauma patients who underwent spine CT. Children were divided into two age groups: 0-10 years (group 1) and 11-17 years (group 2). Each group was subdivided into thoracic, thoracolumbar and lumbar CT groups. CT acquisition parameters (tube potential, in kilovoltage [kV]; mean tube current-time product, in milliamperes [mAs]; reference mAs; collimated slice width; tube rotation time; pitch; scan length) and radiation dose descriptors (volume CT dose index [CTDIvol] and dose-length product [DLP]) were recorded. The CTDIvol and DLP values of spine CTs obtained with different tube potential and collimated slice width values were compared for each group.

RESULTS: CTDIvol and DLP values of thoracolumbar spine CTs in group 1 and lumbar spine CTs in group 2 were significantly lower in CTs acquired with low tube potential levels (P<0.05). CTDIvol and DLP values of thoracolumbar spine CTs in both groups and lumbar spine CTs in group 2 acquired with high collimated slice width values were significantly lower than in corresponding CTs acquired with low collimated slice width values (P<0.05).

CONCLUSION: Pediatric spine CT radiation doses can be notably reduced from the manufacturers' default protocols while preserving image quality.

来源:Pediatr Radiol,查看原文。版权归原作者所有。