文献速递头颈技术与安全CT

基于单能CT学习合成双能CT用于质子放疗停止本领比计算

Learning-based synthetic dual energy CT imaging from single energy CT for stopping power ratio calculation in proton radiation therapy.
PubMed 文献 · Br J Radiol · 2021-10-28

本研究提出一种残差注意力生成对抗网络方法,从单能CT合成双能CT图像,在70例头颈癌患者中回顾性评估。合成高低能CT图像的平均绝对误差分别为36.9 HU和35.8 HU,归一化互相关均为0.96。基于合成双能CT生成的停止本领比图与真实双能CT相比平均归一化均方偏差约1%,剂量体积直方图指标差异在1%以内。

要点

  • 合成双能CT图像平均绝对误差约36 HU,归一化互相关0.96。
  • 合成双能CT生成的停止本领比图与真实双能CT偏差约1%。
  • 基于合成与真实双能CT计算的剂量体积直方图指标差异在1%以内。

局限性

  • 回顾性研究,样本量70例,仅针对头颈癌患者。
  • 采用留一交叉验证,未在独立外部数据集验证。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: Dual energy CT (DECT) has been shown to estimate stopping power ratio (SPR) map with a higher accuracy than conventional single energy CT (SECT) by obtaining the energy dependence of photon interactions. This work presents a learning-based method to synthesize DECT images from SECT image for proton radiotherapy.

METHODS: The proposed method uses a residual attention generative adversarial network. Residual blocks with attention gates were used to force the model to focus on the difference between DECT images and SECT images. To evaluate the accuracy of the method, we retrospectively investigated 70 head-and-neck cancer patients whose DECT and SECT scans were acquired simultaneously. The model was trained to generate both a high and low energy DECT image based on a SECT image. The generated synthetic low and high DECT images were evaluated against the true DECT images using leave-one-out cross-validation. To evaluate our method in the context of a practical application, we generated SPR maps from synthetic DECT (sDECT) using a dual-energy based stoichiometric method and compared the SPR maps to those generated from DECT. A dosimetric comparison for dose obtained from DECT was performed against that derived from sDECT.

RESULTS: The mean of mean absolute error, peak signal-to-noise ratio and normalized cross-correlation for the synthetic high and low energy CT images was 36.9 HU, 29.3 dB, 0.96 and 35.8 HU, 29.2 dB, and 0.96, respectively. The corresponding SPR maps generated from synthetic DECT showed an average normalized mean square deviation of about 1% with reduced noise level and artifacts than those from original DECT. Dose-volume histogram (DVH) metrics for the clinical target volume agree within 1% between the DECT and sDECT calculated dose.

CONCLUSION: Our method synthesized accurate DECT images and showed a potential feasibility for proton SPR map generation.

ADVANCES IN KNOWLEDGE: This study investigated a learning-based method to synthesize DECT images from SECT image for proton radiotherapy.

来源:Br J Radiol,查看原文。版权归原作者所有。