支架取栓器长度不足相对血栓长度的风险:急性缺血性卒中患者研究
该回顾性研究纳入418例急性缺血性卒中行支架取栓患者,采用SWI测量血栓长度,分析血栓长度/支架取栓器长度比值与再通及栓塞的关系。结果示该比值越高,首次再通及总体再通成功率越低,新区域栓塞风险越高;建议根据血栓长度选择支架取栓器以避免长度不足。
要点
- 血栓长度/支架取栓器长度比值是再通失败的独立预测因素。
- 比值越高,首次再通及总体再通率越低,新区域栓塞风险越高。
- 选择支架取栓器时应尽量使其长度不短于血栓长度。
关键影像表现/征象
- SWI上磁敏感血管征用于测量血栓长度。
- 支架取栓器长度增加与首次再通相关。
- 血栓长度缩短及血栓长度/支架取栓器长度比值降低与首次再通和总体再通相关。
临床应用与阅片要点
- 术前应通过SWI测量血栓长度,并据此选择足够长度的支架取栓器。
- 避免使用长度短于血栓的支架取栓器,以降低再通失败和新区域栓塞风险。
局限性
- 单中心回顾性研究。
- 血栓长度通过SWI磁敏感血管征测量,可能受成像条件影响。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND PURPOSE: Results regarding the association of thrombus length, stent retriever length, and recanalization success in patients with acute ischemic stroke are inconsistent. We hypothesized that the ratio of thrombus length to stent retriever length may be of particular relevance.
MATERIALS AND METHODS: Patients with acute ischemic stroke undergoing stent retriever thrombectomy at our institution between January 2010 and December 2018 were reviewed retrospectively. Thrombus length was assessed by measuring the susceptibility vessel sign on SWI using a 1.5T or 3T MR imaging scanner. Multivariable logistic regression models were used to determine the association between thrombus length, stent retriever length, and thrombus length/stent retriever length ratio with first-pass recanalization, overall recanalization, and embolization in new territories. Results are shown as adjusted ORs with 95% CIs. Additional mediation analyses were performed to test for indirect effects on first-pass recanalization and overall recanalization success.
RESULTS: The main analysis included 418 patients (mean age, 74.9 years). Increasing stent retriever length was associated with first-pass recanalization. Decreasing thrombus length and lower thrombus length/stent retriever length ratios were associated with first-pass recanalization and overall recanalization. Thrombus length and stent retriever length showed no association with first-pass recanalization or overall recanalization once thrombus length/stent retriever length ratio was factored in, while thrombus length/stent retriever length ratio remained a significant factor in both models (adjusted OR, 0.316 [95% CI, 0.112-0.892]; P = .030 and adjusted OR, = 0.366 [95% CI, 0.194-0.689]; P = .002). Mediation analyses showed that decreasing thrombus length and increasing stent retriever length had a significant indirect effect on first-pass recanalization mediated through thrombus length/stent retriever length ratio. The only parameter associated with embolization in new territories was an increasing thrombus length/stent retriever length ratio (adjusted OR, 5.079 [95% CI, 1.332-19.362]; P = .017).
CONCLUSIONS: Information about thrombus and stent length is more valuable when combined. High thrombus length/stent retriever length ratios, which may raise the risk of unsuccessful recanalization and embolization in new territories, should be avoided by adapting stent retriever selection to thrombus length whenever possible.
引用
- Belachew NF, Dobrocky T, Meinel TR, et al. Risks of Undersizing Stent Retriever Length Relative to Thrombus Length in Patients with Acute Ischemic Stroke. AJNR Am J Neuroradiol. 2021. PMID: 34649917. doi:10.3174/ajnr.A7313
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。