指南与共识骨肌CT

关节超高清CT采集协议优化的实用推荐

Ultra-high resolution computed tomography of joints: practical recommendations for acquisition protocol optimization.
指南 / 共识 · Quant Imaging Med Surg · 2021-01-01

本研究在超高清CT扫描仪上评估视野、管电压、管电流、焦点大小、层厚、重建矩阵及等中心偏移等参数对空间分辨率的影响,并使用CATPHAN 600体模和尸体膝关节进行分析。结果显示矩阵大小对空间分辨率影响最大,其次为管输出、重建算法和侧向位移。研究为肌肉骨骼成像的超高清CT空间分辨率优化提供了实用推荐。

要点

  • 重建矩阵从512×512增至2048×2048对空间分辨率提升最大,高剂量协议下TTF10%值增加28.2%。
  • 管输出是第二重要因素,可带来29.6%的TTF10%增益和额外5 lp/cm的可视化。
  • 层厚从0.5 mm减至0.25 mm可增加3 lp/cm(从17至20 lp/cm),但不改变平面内空间分辨率。

诊断标准与关键数值

  • 重建矩阵从512×512增至2048×2048,在高剂量协议下TTF10%值增加28.2%,低剂量协议下增加5.6%,但肉眼可分辨线对数无变化。
  • 管输出增加导致TTF10%值增益29.6%,并额外可视化5 lp/cm。
  • 重建算法选择和侧向位移均带来4 lp/cm的可视化增益。
  • 层厚从0.5 mm减至0.25 mm,线对数从17 lp/cm增至20 lp/cm(增加3 lp/cm),TTF10%值增加17.3%,平面内空间分辨率无变化。

局限性

  • 研究仅使用体模和尸体膝关节,未纳入临床患者;结果可能受扫描仪型号和特定参数设置限制。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: To assess the influence on the spatial resolution of various Ultra-high-resolution computed tomography (CT) parameters and provide practical recommendations for acquisition protocol optimization in musculoskeletal imaging.

METHODS: All acquisitions were performed with an Ultra-high resolution scanner, and variations of the following parameters were evaluated: field-of-view (150-300 mm), potential (80-140 KVp), current (25-250 mAs), focal spot size (0.4×0.5 to 0.8×1.3 mm2), slice thickness (0.25-0.5 mm), reconstruction matrix (512×512 to 2048×2048), and iso-centering (up to 85 mm off-center). Two different image reconstruction algorithms were evaluated: hybrid iterative reconstruction (HIR) and model-based iterative reconstruction (MBIR). CATPHAN 600 phantom images were analyzed to calculate the number of visible line pairs per centimeter (lp/cm). Task transfer function (TTF) curves were calculated to quantitatively evaluate spatial resolution. Cadaveric knee acquisitions were also performed.

RESULTS: Under the conditions studied, the factor that most intensely influenced spatial resolution was the matrix size (additional visualization of up to 8 lp/cm). Increasing the matrix from 512×512 to 2048×2048 led to a 28.2% increase in TTF10% values with a high-dose protocol and a 5.6% increase with a low-dose protocol with no change in the number of visually distinguishable line pairs. The second most important factor affecting spatial resolution was the tube output (29.6% TTF10% gain and 5 additional lp/cm visualized), followed by the reconstruction algorithm choice and lateral displacement (both with a 4 lp/cm gain). Decreasing the slice thickness from 0.5 to 0.25 mm, led to an increase of 3 lp/cm (from 17 to 20 lp/cm) and a 17.3% increase in TTF10% values with no change in the "in-plane" spatial resolution.

CONCLUSIONS: This study provides practical recommendations for spatial resolution optimization using Ultra-high-resolution CT.

来源:Quant Imaging Med Surg,查看原文。版权归原作者所有。

引用

  1. Gondim Teixeira PA, Villani N, Ait Idir M, et al. Ultra-high resolution computed tomography of joints: practical recommendations for acquisition protocol optimization. Quant Imaging Med Surg. 2021. PMID: 34603984. doi:10.21037/qims-21-217

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。