深度学习重建加速肩关节MRI:图像质量与诊断性能
回顾性研究纳入105例患者共110次3T肩关节MRI,比较标准序列、加速序列(无DLR)和加速序列(有DLR)的图像质量与诊断性能。加速序列联合DLR将扫描时间从9分23秒降至3分5秒(减少67%),主观图像质量、伪影和诊断性能与标准序列相似。
要点
- 加速序列联合DLR可减少67%扫描时间,图像质量与标准序列相当。
- DLR可改善加速序列的噪声和伪影,使其主观评分接近标准序列。
- 对肩袖、肱二头肌长头腱和盂唇撕裂的诊断准确性无显著差异。
关键影像表现/征象
- 标准序列主观图像质量评分读者1和2分别为10.6±1.2和10.5±1.4,加速序列无DLR为8.1±1.3和7.2±1.1,加速序列有DLR为10.7±1.2和10.5±1.6。
- 伪影评分标准序列为9.3±1.2和10.0±1.0,加速序列无DLR为7.3±1.3和9.1±0.8,加速序列有DLR为9.4±1.2和9.8±0.8。
- 评估肩胛下肌腱、冈上-冈下肌腱、肱二头肌长头腱和盂唇病变,图像集间一致性的kappa值为0.809-0.957(读者1冈上-冈下肌腱除外,为0.663-0.700)。
- 读者间一致性kappa值为0.813-0.951,加速序列无DLR评估冈上-冈下肌腱时kappa=0.673。
诊断标准与关键数值
- 在27例关节镜患者中,四种结构撕裂的准确性、敏感性和特异性在图像集间无差异(p > .05)。
临床应用与阅片要点
- DLR加速序列可作为标准肩关节MRI的替代方案,显著缩短扫描时间而不牺牲诊断性能。
- 阅片时需注意加速序列无DLR时图像噪声和伪影增加,可能影响冈上-冈下肌腱评估的读者间一致性。
局限性
- 回顾性研究设计。
- 仅27例患者有关节镜参考标准,样本量较小。
- 单中心研究,使用特定商业DLR流程,结果可能不可推广。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND. Shoulder MRI using standard multiplanar sequences requires long scan times. Accelerated sequences have tradeoffs in noise and resolution. Deep learning-based reconstruction (DLR) may allow reduced scan time with preserved image quality. OBJECTIVE. The purpose of this study was to compare standard shoulder MRI sequences and accelerated sequences without and with DLR in terms of image quality and diagnostic performance. METHODS. This retrospective study included 105 patients (45 men, 60 women; mean age, 57.6 ± 10.9 [SD] years) who underwent a total of 110 3-T shoulder MRI examinations. Examinations included standard sequences (scan time, 9 minutes 23 seconds) and accelerated sequences (3 minutes 5 seconds; 67% reduction), both including fast spin-echo sequences in three planes. Standard sequences were reconstructed using the conventional pipeline; accelerated sequences were reconstructed using both the conventional pipeline and a commercially available DLR pipeline. Two radiologists independently assessed three image sets (standard sequence, accelerated sequence without DLR, and accelerated sequence with DLR) for subjective image quality and artifacts using 4-point scales (4 = highest quality) and identified pathologies of the subscapularis tendon, supraspinatus-infraspinatus tendon, long head of the biceps brachii tendon, and glenoid labrum. Interobserver agreement and agreement between image sets for the evaluated pathologies were assessed using weighted kappa statistics. In 27 patients who underwent arthroscopy, diagnostic performance was calculated using arthroscopic findings as a reference standard. RESULTS. Mean subjective image quality scores for readers 1 and 2 were 10.6 ± 1.2 and 10.5 ± 1.4 for the standard sequence, 8.1 ± 1.3 and 7.2 ± 1.1 for the accelerated sequence without DLR, and 10.7 ± 1.2 and 10.5 ± 1.6 for the accelerated sequence with DLR. Mean artifact scores for readers 1 and 2 were 9.3 ± 1.2 and 10.0 ± 1.0 for the standard sequence, 7.3 ± 1.3 and 9.1 ± 0.8 for the accelerated sequence without DLR, and 9.4 ± 1.2 and 9.8 ± 0.8 for the accelerated sequence with DLR. Interobserver agreement ranged from kappa of 0.813-0.951 except for accelerated sequence without DLR for the supraspinatus-infraspinatus tendon (κ = 0.673). Agreement between image sets ranged from kappa of 0.809-0.957 except for reader 1 for supraspinatus-infraspinatus tendon (κ = 0.663-0.700). Accuracy, sensitivity, and specificity for tears of the four structures were not different (p > .05) among image sets. CONCLUSION. Accelerated sequences with DLR provide 67% scan time reduction with similar subjective image quality, artifacts, and diagnostic performance to standard sequences. CLINICAL IMPACT. Accelerated sequences with DLR may provide an alternative to standard sequences for clinical shoulder MRI.
引用
- Hahn S, Yi J, Lee HJ, et al. Image Quality and Diagnostic Performance of Accelerated Shoulder MRI With Deep Learning-Based Reconstruction. AJR Am J Roentgenol. 2021. PMID: 34523950. doi:10.2214/AJR.21.26577
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。