双能量甲状旁腺4D-CT:40-keV虚拟单能图像改善甲状旁腺病变与甲状腺组织的鉴别
本研究回顾性纳入20例双能量甲状旁腺CT,通过ROI分析比较标准70-keV与40-keV非增强图像上甲状腺、甲状旁腺病变及胸锁乳突肌的衰减和对比噪声比。结果显示40-keV下甲状腺衰减绝对值和百分比增加显著大于甲状旁腺病变和胸锁乳突肌(P<.001),甲状腺相对于甲状旁腺病变的对比噪声比中位增加0.8(P<.001)。结论认为40-keV虚拟单能图像通过显著增加甲状腺衰减及对比噪声比,有助于区分甲状旁腺病变与甲状腺组织。
要点
- 40-keV虚拟单能图像可增大甲状腺与甲状旁腺病变的衰减差异,辅助鉴别。
- 该技术对动脉期和静脉期与甲状腺等密度的甲状旁腺病变尤其有价值。
- 非增强期结合40-keV重建可减少因等密度而漏诊的甲状旁腺病变。
核心结论
- 40-keV虚拟单能图像通过显著增加甲状腺衰减及对比噪声比,促进甲状旁腺病变与甲状腺组织的区分。
研究设计与样本
- 回顾性研究,纳入20例双能量甲状旁腺CT,采用ROI分析比较70-keV与40-keV非增强图像。
关键影像表现/征象
- 40-keV下甲状腺衰减的绝对值和百分比增加显著大于甲状旁腺病变和胸锁乳突肌(P<.001)。
- 40-keV下甲状腺相对于甲状旁腺病变的对比噪声比中位增加0.8(P<.001)。
- 40-keV下甲状腺相对于胸锁乳突肌的对比噪声比中位增加1.3(P<.001)。
临床应用与阅片要点
- 在非增强期加用40-keV虚拟单能图像,可突出甲状腺碘含量,帮助识别与甲状腺等密度的甲状旁腺病变。
- 对于动脉期和静脉期与甲状腺等密度的甲状旁腺病变,非增强40-keV图像可减少漏诊。
局限性
- 回顾性研究,样本量较小(20例),未报告多中心验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND PURPOSE: In parathyroid CT, a noncontrast phase aids discrimination of parathyroid lesions (not iodine-containing) from thyroid tissue (iodine-containing). When thyroid iodine is pathologically diminished, this differentiation is difficult with standard CT. Because the attenuation of an element is maximal near its K-edge (iodine = 33.2 keV), we hypothesized that dual-energy CT 40-keV virtual monoenergetic images will accentuate thyroid iodine relative to standard images, improving the differentiation of thyroid from parathyroid lesions. Our purpose was to test this hypothesis through quantitative assessment of Hounsfield unit attenuation and contrast-to-noise on dual-energy CT standard (70-keV) and 40-keV noncontrast images.
MATERIALS AND METHODS: For this retrospective study including 20 dual-energy parathyroid CTs, we used an ROI-based analysis to assess the attenuation of thyroid tissue, parathyroid lesions, and sternocleidomastoid muscle as well as corresponding contrast-to-noise on standard and 40- keV noncontrast images. Wilcoxon signed rank tests were performed to compare differences between 70 and 40 keV.
RESULTS: Absolute and percentage increases in attenuation at 40 keV were significantly greater for thyroid gland than for parathyroid lesions and sternocleidomastoid muscle (P < .001 for all). Significant increases in the contrast-to-noise of thyroid relative to parathyroid lesions (median increase, 0.8; P < .001) and relative to sternocleidomastoid muscle (median increase, 1.3; P < .001) were observed at 40 keV relative to 70 keV.
CONCLUSIONS: Forty-kiloelectron volt virtual monoenergetic images facilitate discrimination of parathyroid lesions from thyroid tissue by significantly increasing thyroid attenuation and associated contrast-to-noise. These findings are particularly relevant for parathyroid lesions that exhibit isoattenuation to the thyroid on parathyroid CT arterial and venous phases and could, therefore, be missed without the noncontrast phase.
引用
- Bunch PM, Pavlina AA, Lipford ME, et al. Dual-Energy Parathyroid 4D-CT: Improved Discrimination of Parathyroid Lesions from Thyroid Tissue Using Noncontrast 40-keV Virtual Monoenergetic Images. AJNR Am J Neuroradiol. 2021. PMID: 34475194. doi:10.3174/ajnr.A7265
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。