超声弹性成像中剪切波速度与频散量化变异性的体模研究
该体模研究评估了操作者相关采集参数对剪切波速度(SWS)和剪切波频散(SWD)测量变异性的影响。使用Canon Aplio i800系统和四种硬度认证体模(3.7–44 kPa),发现SWD变异性是SWS的2–3倍,采集深度是主要变异因素,且SWD对ROI横向位置也敏感。结果提示多参数弹性成像指南应纳入深度和横向依赖性变异。
要点
- SWD测量变异性显著高于SWS,解读时需更谨慎。
- 采集深度增加导致SWS和SWD均显著衰减,应固定深度。
- ROI横向位置对SWD变异影响与深度相当,需标准化。
局限性
- 仅使用单一品牌超声系统和体外均匀体模,结果外推至临床需谨慎。
- 未评估其他操作者因素或生理运动影响。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
There is a growing interest in quantifying shear-wave dispersion (SWD) with ultrasound shear-wave elastography (SWE). Recent studies suggest that SWD complements shear-wave speed (SWS) in diffuse liver disease diagnosis. To accurately interpret these metrics in clinical practice, we analyzed the impact of operator-dependent acquisition parameters on SWD and SWS measurements. Considered parameters were the acquisition depth, lateral position and size of the region of interest (ROI), as well as the size of the SWE acquisition box. Measurements were performed using the Canon Aplio i800 system (Canon Medical Systems, Otawara, Tochigi, Japan) and four homogeneous elasticity phantoms with certified stiffness values ranging from 3.7 to 44 kPa. In general, SWD exhibited two to three times greater variability than SWS. The acquisition depth was the main variance-contributing factor for both SWS and SWD, which decayed significantly with depth. The lateral ROI position contributed as much as the acquisition depth to the total variance in SWD. Locations close to the initial shear-wave excitation pulse were more robust to biases because of inaccurate probe-phantom coupling. The size of the ROI and acquisition box did not introduce significant variations. These results suggest that future guidelines on multiparametric elastography should account for the depth- and lateral-dependent variability of measurements.
引用
- Korta Martiartu N, Nambiar S, Nascimento Kirchner I, et al. Sources of Variability in Shear Wave Speed and Dispersion Quantification with Ultrasound Elastography: A Phantom Study. Ultrasound Med Biol. 2021. PMID: 34548187. doi:10.1016/j.ultrasmedbio.2021.08.013
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。