动态320排CT冠状动脉血流定量与13N-氨PET心肌血流储备验证
本研究在46例冠心病患者中,用320排低剂量动态冠状动脉CT血管成像估算静息冠状动脉血流指数(rCFI及改良rCFI),并以13N-氨PET测定的心肌血流储备(MFR)和缺血为参考标准进行验证。结果显示rCFI与负荷心肌血流量及MFR正相关,检测MFR<2.0和缺血的AUC分别为0.73~0.91,特异度高达93%~98%,但敏感度仅34%~53%。
要点
- rCFI检测MFR<2.0或缺血的特异度很高(93%~98%),但敏感度低(34%~53%)。
- rCFI与13N-氨PET的负荷心肌血流量及MFR呈显著正相关。
- rCFI可作为标准CCTA之外的附加功能学检查,但不能替代PET。
核心结论
- 两种rCFI均与NH3-PET定量值相关,检测功能性缺血时特异度高。
研究设计与样本
- 单中心研究,46例冠心病患者接受动态CCTA和NH3-PET,其中21例用原始rCFI、25例用改良rCFI。
- 使用320排低剂量动态CCTA,改良方案结合动态扫描与标准CCTA并采用剂量调制。
- 以NH3-PET测定的静息和负荷心肌血流量、MFR及缺血有无为参考标准。
诊断标准与关键数值
- rCFI和改良rCFI检测MFR<2.0的最佳截断值分别为0.39和0.61,AUC为0.75和0.73,敏感度48%和34%,特异度97%和98%。
- rCFI和改良rCFI区分缺血与非缺血节段的最佳截断值对应AUC为0.75和0.91,敏感度53%和50%,特异度93%和95%。
临床应用与阅片要点
- rCFI可在标准CCTA基础上提供附加功能学信息,高特异度有助于排除功能性缺血。
- 因敏感度较低,rCFI阴性不能完全排除缺血,需结合PET等功能学检查。
局限性
- 样本量较小(46例),且原始rCFI与改良rCFI分别仅21例和25例。
- 单中心研究,参考标准为NH3-PET而非有创血流储备分数。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: Resting coronary flow index (rCFI) estimated by 320-detector low-dose dynamic coronary CT angiography (CCTA) is a direct flow quantification using intracoronary attenuation. We propose modified-rCFI from new protocol combining dynamic scan and standard CCTA using dose-modulation, and validate its consistency with quantitative values and ischemia depicted by 13N-ammonia PET (NH3-PET).
METHODS: 46 patients who underwent dynamic CCTA and NH3-PET for coronary artery disease were evaluated using original rCFI in 21 patients and modified-rCFI in 25 patients. Two types of rCFI were calculated for three major coronary arteries. Myocardial blood flow (MBF) at rest and stress, myocardial flow reserve (MFR), and the presence or absence of ischemia for three major territories were depicted by NH3-PET. Coronary territories were categorized as territories with MFR <2.0, ≥2.0, or with and without ischemia. Receiver operating characteristic analysis was performed to determine the optimal cut-off of rCFI to distinguish territories with MFR <2.0 or the presence of ischemia.
RESULTS: rCFI and modified-rCFI had significant positive correlations with stress MBF and MFR. The optical cut-offs of rCFI and modified-rCFI of 0.39 and 0.61 could detect territories with MFR <2.0, with AUCs of 0.75 and 0.73, sensitivities of 48 and 34%, and specificities of 97 and 98%. Optimal cut-offs of rCFI and modified-rCFI distinguished ischemic segments from non-ischemic segments, with AUCs of 0.75 and 0.91, sensitivities of 53 and 50%, and specificities of 93 and 95%.
CONCLUSION: Two types of rCFI correlated with quantitative values from NH3-PET, and were consistent with a high specificity in detecting functional ischemia.
ADVANCES IN KNOWLEDGE: rCFI can contribute as additional functional test over standard CCTA in clinical work-up.