UHR-CT图像噪声更高但诊断置信度高,以侵入性造影为参考,对冠脉狭窄(≥70%)评估的敏感度为86%、特异度为88%。初步提示UHR-CT可能克服常规CT在严重钙化血管中评估狭窄的局限。

要点

  • 严重钙化冠脉狭窄评估中,UHR-CT以侵入性造影为参考的敏感度86%、特异度88%。
  • UHR-CT图像噪声明显高于常规CT重建,但诊断置信度评分仍高(4.3±0.9)。
  • 本组辐射剂量9.3 mSv,与扩大曝光范围(70%-99% R-R周期)的研究软件应用有关。

关键影像表现/征象

  • UHR-CT整体图像噪声为50.9±7.8,明显高于模拟常规CT重建图像的19.5±8.3(P < .01)。
  • UHR-CT诊断置信度评分较高,为4.3±0.9(1-5级)。
  • 无支架患者(n=6)平均钙化积分为1205;86支评估血管中22支(26%)侵入性造影示狭窄≥70%。

诊断标准与关键数值

  • 以侵入性冠状动脉造影为参考,UHR-CT评估冠脉狭窄(≥70%)的敏感度为86%(19/22),特异度为88%(56/64)。
  • 敏感度95%CI为65%-97%,特异度95%CI为77%-95%。

临床应用与阅片要点

  • 对严重钙化冠脉病变,UHR-CT可在高图像噪声下仍保持较高诊断置信度,阅片时需结合噪声水平判断。
  • 报告狭窄时应注意本组特异度88%,仍存在假阳性可能,必要时结合侵入性造影。
  • 辐射剂量9.3 mSv与扩大曝光范围(70%-99% R-R周期)有关,日常应用需关注剂量控制。

局限性

  • 单中心前瞻性研究,部分对比依赖可获得的临床常规CT结果。
  • 辐射剂量因研究软件应用扩大曝光范围而偏高(9.3 mSv)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: Conventional CT technology yields only modest accuracy of coronary artery stenosis assessment in severely calcified lesions. Reported herein are this study's initial observations on the potential of ultra-high-resolution CT (UHR-CT) for evaluating severely calcified coronary arterial lesions.

MATERIALS AND METHODS: Fifteen patients 45 years of age or older, with history of coronary artery disease, referred for invasive coronary angiography, were prospectively enrolled. Patients underwent UHR-CT within 30 days prior to cardiac catheterization. Image noise levels and diagnostic confidence (level 1-5) using UHR-CT were compared with reconstructed images simulating conventional CT technology. Stenosis assessment for the major coronary arteries and the left main coronary artery with UHR-CT and invasive angiography were compared. Results from clinically driven coronary CT using conventional technology were considered for comparison when available.

RESULTS: Mean patient age was 67 years (range, 53-79 years). Thirteen patients were men, nine had obesity. Radiation dose was 9.3 mSv owing to expanded x-ray exposure to accommodate research software application (70%-99% of R-R cycle). Overall image noise was considerably greater for UHR-CT (50.9 ± 7.8 [standard deviation]) versus conventional CT image reconstruction (19.5 ± 8.3, P < .01), yet diagnostic confidence scores for UHR-CT were high (4.3 ± 0.9). Average calcium score in patients without stents (n = 6) was 1205, and of 86 vessels evaluated, 22 had 70% or greater stenosis depicted with invasive angiography (26%). Stenosis comparison with invasive angiography yielded 86% (19 of 22) sensitivity and 88% (56 of 64) specificity (95% CI: 65%, 97%; and 77%, 95%, respectively).

CONCLUSION: Initial observations suggest UHR-CT may be effective in overcoming the limitation of conventional CT for accurately evaluating coronary artery stenoses in severely calcified vessels.Keywords: CT-Angiography, Coronary Arteries, ArteriosclerosisClinical trial registration no. NCT04272060See also commentary by Shanbhag and Chen in this issue.© RSNA, 2021.

来源:Radiol Cardiothorac Imaging,查看原文。版权归原作者所有。