文献速递胸部CT

CT图像采集参数对不同大小肺结节放射组学恶性预测性能的影响

Effect of CT image acquisition parameters on diagnostic performance of radiomics in predicting malignancy of pulmonary nodules of different sizes.
PubMed 文献 · Eur Radiol · 2021-09-21

该回顾性研究纳入696例肺结节患者,按直径分为T1a、T1b、T1c三组,并比较四种CT图像设置(层厚与卷积核组合)下放射组学模型预测恶性的AUC。结果显示,对于直径≤1.0 cm的结节,薄层联合锐利卷积核重建的图像所建模型AUC最高(0.84),且显著优于其他设置;对于直径>1.0 cm的结节,重建参数对诊断性能无显著影响。

要点

  • 对于直径≤1.0 cm的肺结节,推荐使用薄层(1.25 mm)联合锐利卷积核重建图像进行放射组学分析。
  • 直径>1.0 cm的肺结节,放射组学模型性能受CT重建参数影响不大,可灵活选择。
  • 不同大小肺结节需匹配不同CT采集参数以优化放射组学诊断效能。

核心结论

  • 对于直径≤1.0 cm的肺结节,CT图像采集参数显著影响放射组学预测恶性的诊断性能,薄层联合锐利卷积核重建的模型AUC最高。
  • 对于直径>1.0 cm的肺结节,CT重建参数对放射组学诊断性能无显著影响。

研究设计与样本

  • 回顾性研究,纳入2015年3月至2018年3月共696例肺结节患者的CT图像,按结节直径分为T1a(≤1.0 cm)、T1b(1.0 cm<直径≤2.0 cm)、T1c(2.0 cm<直径≤3.0 cm)三组。

诊断标准与关键数值

  • 预测肺结节恶性的AUC:T1a组在设置1、设置2、设置3及随机设置下分别为0.84、0.64、0.68、0.68。
  • 预测肺结节恶性的AUC:T1b组在设置1、设置2、设置3及随机设置下分别为0.68、0.74、0.76、0.70。
  • 预测肺结节恶性的AUC:T1c组在设置1、设置2、设置3及随机设置下分别为0.66、0.64、0.63、0.70。
  • T1a组中设置1的AUC显著高于其他所有设置;T1b组中设置3的AUC显著高于部分设置;T1c组中各模型间无统计学显著差异。

临床应用与阅片要点

  • 对于直径≤1.0 cm的肺结节,放射组学分析应优先选用薄层(1.25 mm)联合锐利卷积核重建的CT图像,以获得最佳恶性预测性能。
  • 对于直径>1.0 cm的肺结节,重建参数选择对放射组学模型性能影响不大,可依据临床常规选择。
  • 在临床应用中,需注意不同CT采集参数可能引入放射组学特征变异,尤其对小肺结节应规范重建方案。

局限性

  • 本研究为回顾性单中心研究,样本量696例,可能存在选择偏倚。
  • 放射组学模型仅在内部训练集上构建和验证,缺乏外部独立验证队列。
  • 研究未报告敏感度、特异度及阈值等具体诊断效能指标,仅以AUC衡量模型性能。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To investigate the effect of CT image acquisition parameters on the performance of radiomics in classifying benign and malignant pulmonary nodules (PNs) with respect to nodule size.

METHODS: We retrospectively collected CT images of 696 patients with PNs from March 2015 to March 2018. PNs were grouped by nodule diameter: T1a (diameter ≤ 1.0 cm), T1b (1.0 cm < diameter ≤ 2.0 cm), and T1c (2.0 cm < diameter ≤ 3.0 cm). CT images were divided into four settings according to slice-thickness-convolution-kernels: setting 1 (slice thickness/reconstruction type: 1.25 mm sharp), setting 2 (5 mm sharp), setting 3 (5 mm smooth), and random setting. We created twelve groups from two interacting conditions. Each PN was segmented and had 1160 radiomics features extracted. Non-redundant features with high predictive ability in training were selected to build a distinct model under each of the twelve subsets.

RESULTS: The performance (AUCs) on predicting PN malignancy were as follows: T1a group: 0.84, 0.64, 0.68, and 0.68; T1b group: 0.68, 0.74, 0.76, and 0.70; T1c group: 0.66, 0.64, 0.63, and 0.70, for the setting 1, setting 2, setting 3, and random setting, respectively. In the T1a group, the AUC of radiomics model in setting 1 was statistically significantly higher than all others; In the T1b group, AUCs of radiomics models in setting 3 were statistically significantly higher than some; and in the T1c group, there were no statistically significant differences among models.

CONCLUSIONS: For PNs less than 1 cm, CT image acquisition parameters have a significant influence on diagnostic performance of radiomics in predicting malignancy, and a model created using images reconstructed with thin section and a sharp kernel algorithm achieved the best performance. For PNs larger than 1 cm, CT reconstruction parameters did not affect diagnostic performance substantially.

KEY POINTS: • CT image acquisition parameters have a significant influence on the diagnostic performance of radiomics in pulmonary nodules less than 1 cm. • In pulmonary nodules less than 1 cm, a radiomics model created by using images reconstructed with thin section and a sharp kernel algorithm achieved the best diagnostic performance. • For PNs larger than 1 cm, CT image acquisition parameters do not affect diagnostic performance substantially.

来源:Eur Radiol,查看原文。版权归原作者所有。

引用

  1. Xu Y, Lu L, Sun SH, et al. Effect of CT image acquisition parameters on diagnostic performance of radiomics in predicting malignancy of pulmonary nodules of different sizes. Eur Radiol. 2021. PMID: 34549324. doi:10.1007/s00330-021-08274-1

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。