英国肺癌筛查试验CT扫描协议的建立

Establishing scanning protocols for a CT lung cancer screening trial in the UK.
PubMed 文献 · Br J Radiol · 2021-09-24

本研究使用体模在社区移动CT上开发低剂量、高质量肺癌筛查CT协议。替代协议剂量较原协议降低达72%,CTDIvol为0.6mGy(55kg等效体模)和1.3mGy(附加6cm组织等效材料)。调整重建参数后图像质量满足全部六项QIBA建议,放射科医生认为适用于肺癌筛查试验。

要点

  • 替代协议剂量降低达72%,CTDIvol为0.6mGy和1.3mGy。
  • 原协议未通过QIBA边缘增强和3D分辨率纵横比两项指标。
  • 调整重建参数后满足全部六项QIBA图像质量建议。

关键影像表现/征象

  • 替代协议CTDIvol为0.6mGy(55kg等效体模)和1.3mGy(附加6cm组织等效材料)。
  • 原协议未通过QIBA边缘增强和3D分辨率纵横比两项指标。
  • 调整重建参数后图像质量满足全部六项QIBA建议。

诊断标准与关键数值

  • 以QIBA建议作为图像质量客观指标,包括边缘增强和3D分辨率纵横比等六项指标。
  • 替代协议剂量较原协议降低达72%。

临床应用与阅片要点

  • 建立肺癌筛查CT协议时应使用拟人胸部体模和图像质量体模进行剂量与客观图像质量验证。
  • 应参照QIBA建议评估图像质量,确保协议满足推荐指标。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To develop a CT scanning protocol for lung cancer screening which achieved low radiation dose and a high level of objectively assessed image quality.

METHODS: An anthropomorphic chest phantom and a commercially available lung screening image quality phantom were scanned on a series of scan protocols from a previous UK lung screening pilot and on an alternative protocol. The chest phantom scans were used to assess the CT dose metrics on community-based mobile CT scanners and comparisons were made with published recommended doses. Scans of the image quality phantom were objectively assessed against the RSNA Quantitative Imaging Biomarkers Alliance (QIBA) recommendations. Protocol adjustments were made to ensure that the recommended dose and image quality levels were both achieved.

RESULTS: The alternative scan protocol yielded doses up to 72% lower than on the previously used protocols with a CTDIvol of 0.6mGy for the 55 kg equivalent phantom and 1.3mGy with an additional 6 cm of tissue equivalent material in place. Scans on the existing protocols failed on two of the QIBA image quality metrics (edge enhancement and 3D resolution aspect ratio). Following adjustments to the reconstruction parameters of the resulting image quality met all six QIBA recommendations. Radiologist review of phantom images with this scan protocol deemed them suitable for a lung screening trial.

CONCLUSIONS: Scan protocols yielding low radiation doses and high levels of objectively assessed image quality which meet published criteria can be established through the use of specific anthropomorphic and image quality phantoms, and are deliverable in community-based lung cancer screening.

ADVANCES IN KNOWLEDGE: Development of a standard methodology for establishing CT lung screening scanning protocolsUse of QIBA recommendations as objective image quality metricsStandardised lung phantoms are essential tools for setting up lung screening protocols.