高螺距光子计数探测器CT主动脉血管成像:与等辐射剂量能量积分探测器CT的个体内比较
本研究对40例患者行高螺距双源光子计数探测器CT(PCD-CT)主动脉血管成像,与既往等辐射剂量的能量积分探测器CT(EID-CT)进行个体内比较。结果显示PCD-CT虚拟单能图像在40和45 keV时对比噪声比显著高于EID-CT,且超重患者获益更明显;主观分析提示45-50 keV为客观与主观图像质量的最佳平衡点。
要点
- PCD-CT高螺距主动脉CTA在40/45 keV虚拟单能图CNR显著优于等剂量EID-CT。
- 超重患者使用PCD-CT的CNR提升幅度大于正常体重患者。
- 日常阅片建议关注45-50 keV虚拟单能图像以平衡噪声与血管显示。
局限性
- 单中心、小样本(40例)个体内比较,仅纳入有既往EID-CT的患者。
- PCD-CT使用固定120 kV多能模式,而EID-CT采用自动管电压选择,扫描参数不完全一致。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: The aims of this study were to determine the objective and subjective image quality of high-pitch computed tomography (CT) angiography of the aorta in clinical dual-source photon-counting detector CT (PCD-CT) and to compare the image quality to conventional dual-source energy-integrating detector CT (EID-CT) in the same patients at equal radiation dose.
MATERIALS AND METHODS: Patients with prior CT angiography of the thoracoabdominal aorta acquired on third-generation dual-source EID-CT in the high-pitch mode and with automatic tube voltage selection (ATVS, reference tube voltage 100 kV) were included. Follow-up imaging was performed on a first-generation, clinical dual-source PCD-CT scanner in the high-pitch and multienergy (QuantumPlus) mode at 120 kV using the same contrast media protocol as with EID-CT. Radiation doses between scans were matched by adapting the tube current of PCD-CT. Polychromatic images for both EID-CT and PCD-CT (called T3D) and virtual monoenergetic images at 40, 45, 50, and 55 keV for PCD-CT were reconstructed. Computed tomography attenuation was measured in the aorta; noise was defined as the standard deviation of attenuation; contrast-to-noise ratio (CNR) was calculated. Subjective image quality (noise, vessel attenuation, vessel sharpness, and overall quality) was rated by 2 blinded, independent radiologists.
RESULTS: Forty patients were included (mean age, 63 years; 8 women; mean body mass index [BMI], 26 kg/m2). There was no significant difference in BMI, effective diameter, or radiation dose between scans (all P's > 0.05). The ATVS in EID-CT selected 70, 80, 90, 100, 110, and 120 kV in 2, 14, 14, 7, 2, and 1 patients, respectively. Mean CNR was 17 ± 8 for EID-CT and 22 ± 7, 20 ± 6, 18 ± 5, 16 ± 5, and 12 ± 4 for PCD-CT at 40, 45, 50, 55 keV, and T3D, respectively. Contrast-to-noise ratio was significantly higher for 40 and 45 keV of PCD-CT as compared with EID-CT (both P's < 0.05). The linear regression model (adjusted R2, 0.38; P < 0.001) revealed that PCD-CT reconstruction (P < 0.001), BMI group (P = 0.007), and kV of the EID-CT scan (P = 0.01) were significantly associated with CNR difference, with an increase by 34% with PCD-CT for overweight as compared with normal weight patients. Subjective image quality reading revealed slight differences between readers for subjective vessel attenuation and sharpness, whereas subjective noise was rated significantly higher for 40 and 45 keV (P < 0.001) and overall quality similar (P > 0.05) between scans.
CONCLUSIONS: High-pitch PCD-CT angiography of the aorta with VMI at 40 and 45 keV resulted in significantly increased CNR compared with EID-CT with ATVS at matched radiation dose. The CNR gain of PCD-CT increased in overweight patients. Taking into account the subjective analysis, VMI at 45 to 50 keV is proposed as the best trade-off between objective and subjective image quality.
引用
- Euler A, Higashigaito K, Mergen V, et al. High-Pitch Photon-Counting Detector Computed Tomography Angiography of the Aorta: Intraindividual Comparison to Energy-Integrating Detector Computed Tomography at Equal Radiation Dose. Invest Radiol. 2022. PMID: 34352805. doi:10.1097/RLI.0000000000000816
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。