双能CT肠道蠕动伪影:不同平台与肠道对比剂的体外研究
该体外研究模拟肠道蠕动,在四种双能CT平台上比较120-kVp-like与碘图重建的蠕动伪影严重程度。结果显示,光谱探测器与快速千伏切换平台碘图伪影更轻,双源与分滤器平台则120-kVp-like图更优;实验性暗对比剂较碘剂和水可减轻伪影。
要点
- 光谱探测器和快速kV切换平台优先用碘图评估肠道,伪影更轻。
- 双源和分滤器平台优先用120-kVp-like图,伪影评分更低。
- 实验性暗对比剂可减轻蠕动伪影,优于碘对比剂和水。
关键影像表现/征象
- 蠕动伪影在120-kVp-like图上,分滤器(1.47)和双源(1.86)评分低于光谱探测器(2.58)和快速kV切换(2.74),均p<.001。
- 碘图上,光谱探测器(1.03)和快速kV切换(2.09)伪影较120-kVp-like图轻,p<.001;双源(2.77)和分滤器(2.62)则更重,p<.001。
- 实验性暗肠道对比剂伪影评分(1.79)低于其他肠道对比剂(均p<.001)。
诊断标准与关键数值
- 伪影严重程度评分0(无)到3(严重),读片者间ICC为0.82(95%CI 0.79-0.84)。
- 碘图帮助识别120-kVp-like图上肠壁高密度为伪影的比例:光谱探测器94.7%,快速kV切换40.7%。
临床应用与阅片要点
- 根据双能CT平台选择重建图像:光谱探测器和快速kV切换看碘图,双源和分滤器看120-kVp-like图。
- 碘图可用于鉴别肠壁高密度是否为蠕动伪影,光谱探测器平台尤其有效。
- 暗肠道对比剂未来临床应用后可能进一步减少蠕动伪影。
局限性
- 体外模拟研究,使用波纹管而非真实肠道。
- 实验性暗对比剂尚未临床可用。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND. The value of dual-energy CT (DECT) for bowel wall assessment is increasingly recognized. Although technical improvements reduce peristalsis artifact in conventional CT, the effects of peristalsis on DECT image reconstructions remain poorly studied. OBJECTIVE. The purpose of this study was to evaluate the influence of different DECT scanners and enteric contrast agents on the severity of bowel peristalsis artifact in vitro. METHODS. To simulate bowel peristalsis, a 3-cm-diameter corrugated hollow tube representing the bowel was oscillated constantly in the z-axis within a larger water-filled cylinder. The bowel was serially filled with air, water, and iodinated or experimental dark contrast material and scanned on four different DECT platforms (spectral detector, rapid peak kilovoltage switching, split filter, and dual source) to reconstruct 120-kVp-like and iodine images. Two readers rated each image reconstruction for artifact severity from 0 (none) to 3 (severe) and recorded the degree to which iodine images depicted bowel wall hyperattenuation on 120-kVp-like images as artifactual. Artifact severity scores were compared by ANOVA with Bonferroni correction. RESULTS. Interrater agreement on artifact scores was excellent (intraclass correlation coefficient, 0.82 [95% CI, 0.79-0.84]). For 120-kVp-like images, mean peristalsis artifact scores were lower (all p < .001) for split-filter (1.47) and dual-source (1.86) scanners than for spectral-detector (2.58) and rapid-kilovoltage-switching (2.74) scanners. Compared with those on 120-kVp images, peristalsis artifacts on iodine images were less severe for spectral-detector (score, 1.03; p < .001) and rapid-kilovoltage-switching (2.09; p < .001) systems but more severe for dual-source (2.77; p < .001) and split-filter (2.62; p < .001) systems. Peristalsis artifact was rated less severe with experimental dark bowel contrast medium (score, 1.79) than with other bowel contrast agents (all p < .001). Iodine images helped identify bowel wall hyperattenuation as artifactual in 94.7% of reviewed cases for spectral-detector and 40.7% of cases for rapid-kilovoltage-switching scanners. CONCLUSION. For spectral-detector and rapid-kilovoltage-switching DECT, iodine images minimize peristalsis artifact, but for dual-source and split-filter DECT, mixed 120-kVp-like images are preferred. Compared with iodinated contrast material and water, experimental dark bowel contrast material reduces peristalsis artifact. CLINICAL IMPACT. Knowledge of the preferred images for reducing peristalsis artifact can lessen the effect of peristalsis on clinical DECT interpretation. Dark enteric contrast agents, when they become clinically available, may further reduce the effects of peristalsis.
引用
- Obmann MM, Sun Y, An C, et al. Bowel Peristalsis Artifact on Dual-Energy CT: In Vitro Study on the Influence of Different Dual-Energy CT Platforms and Enteric Contrast Agents. AJR Am J Roentgenol. 2021. PMID: 34406059. doi:10.2214/AJR.21.26345
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。