文献速递外周血管CT

下肢外周动脉钙化定量:不同CT扫描仪与评分平台的比较

Quantification of Calcium in Peripheral Arteries of the Lower Extremities: Comparison of Different CT Scanners and Scoring Platforms.
PubMed 文献 · Invest Radiol · 2022-01-01

不同CT扫描仪间Agatston和体积评分差异显著,而不同评分平台间变异有限,体积评分尤为稳定。钙化定量的变异性主要来源于CT扫描仪,且Agatston评分更明显。

要点

  • 不同CT扫描仪间钙化评分差异显著,Agatston评分变异大于体积评分。
  • 不同评分平台间体积评分变异小,Agatston评分仅OsiriX与其他平台有差异。
  • 随访比较钙化评分时需注意扫描仪型号对Agatston评分的影响。

局限性

  • 仅比较3种CT扫描仪和4种评分平台,未涉及活体或临床结局。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: The aim of this study was to investigate the interscanner and interscoring platform variability of calcium quantification in peripheral arteries of the lower extremities.

MATERIALS AND METHODS: Twenty human fresh-frozen legs were scanned using 3 different computed tomography (CT) scanners. The radiation dose (CTDIvol) was kept similar for all scanners. The calcium scores (Agatston and volume scores) were quantified using 4 semiautomatic scoring platforms. Comparative analysis of the calcium scores between scanners and scoring platforms was performed by using the Friedman test; post hoc analysis was performed by using the Wilcoxon signed rank test with Bonferroni correction.

RESULTS: Sixteen legs had calcifications and were used for data analysis. Agatston and volume scores ranged from 12.1 to 6580 Agatston units and 18.2 to 5579 mm3. Calcium scores differed significantly between Philips IQon and Philips Brilliance 64 (Agatston: 19.5% [P = 0.001]; volume: 14.5% [P = 0.001]) and Siemens Somatom Force (Agatston: 18.1% [P = 0.001]; volume: 17.5% [P = 0.001]). The difference between Brilliance 64 and Somatom Force was smaller (Agatston: 5.6% [P = 0.778]; volume: 7.7% [P = 0.003]). With respect to the interscoring platform variability, OsiriX produced significantly different Agatston scores compared with the other 3 scoring platforms (OsiriX vs IntelliSpace: 14.8% [P = 0.001] vs Syngo CaScore: 13.9% [P = 0.001] vs iX viewer: 13.2% [P < 0.001]). For the volume score, the differences between all scoring platforms were small ranging from 2.9% to 4.0%. Post hoc analysis showed a significant difference between OsiriX and IntelliSpace (3.8% [P = 0.001]).

CONCLUSIONS: The use of different CT scanners resulted in notably different Agatston and volume scores, whereas the use of different scoring platforms resulted in limited variability especially for the volume score. In conclusion, the variability in calcium quantification was most evident between different CT scanners and for the Agatston score.

来源:Invest Radiol,查看原文。版权归原作者所有。