深度学习算法在低剂量腰椎CT中同步降噪与边缘锐化的初步研究

Deep Learning Algorithm for Simultaneous Noise Reduction and Edge Sharpening in Low-Dose CT Images: A Pilot Study Using Lumbar Spine CT.
PubMed 文献 · Korean J Radiol · 2021-08-19

本研究回顾性纳入52例接受CT引导下腰椎骨活检的患者,比较50 mAs、经深度学习降噪的50 mAs与100 mAs腰椎CT图像。结果显示降噪后50 mAs图像噪声显著低于其他两组,SNR提高,且边缘锐度优于100 mAs图像,正常解剖结构可见性显著改善。提示深度学习重建可在低剂量腰椎CT中同步实现降噪与边缘锐化。

要点

  • 深度学习降噪后50 mAs图像噪声(36.38 HU)显著低于100 mAs(63.33 HU)和50 mAs(93.33 HU)。
  • 降噪后50 mAs图像SNR(1.46)高于100 mAs(0.99)和50 mAs(0.58)。
  • 降噪后50 mAs图像在椎体和腰大肌的边缘锐度均优于100 mAs图像。

关键影像表现/征象

  • 降噪后50 mAs图像噪声为36.38 ± 7.03 HU,低于50 mAs(93.33 ± 25.36 HU)和100 mAs(63.33 ± 16.09 HU)图像(p < 0.001)。
  • 降噪后50 mAs图像SNR为1.46 ± 0.54,高于100 mAs(0.99 ± 0.34)和50 mAs(0.58 ± 0.18)图像(p < 0.001)。
  • 降噪后50 mAs图像在椎体的边缘上升距离(ERD)为0.94 ± 0.2 mm,优于100 mAs图像的1.05 ± 0.24 mm(p = 0.036)。
  • 降噪后50 mAs图像在腰大肌的ERD为0.42 ± 0.09 mm,优于100 mAs图像的0.50 ± 0.12 mm(p = 0.002)。
  • 两位肌骨放射科医师评估显示降噪后50 mAs图像显著改善正常解剖结构的可见性(p < 0.001)。

临床应用与阅片要点

  • 在低剂量腰椎CT中可尝试使用深度学习降噪技术,以在降低辐射剂量的同时维持或改善图像质量。
  • 评估图像质量时需同时关注噪声、SNR和边缘锐度(ERD),避免仅以噪声降低作为唯一指标。
  • 注意该技术为不依赖厂商的深度学习模型,实际应用时需验证其在不同设备和协议下的表现。

局限性

  • 本研究为回顾性、单中心初步研究,样本量有限(52例)。
  • 仅评估了正常解剖结构的可见性,未涉及病变检出或诊断准确性的临床结局。
  • 需进一步研究更低的辐射剂量以及该技术的临床适用性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: The purpose of this study was to assess whether a deep learning (DL) algorithm could enable simultaneous noise reduction and edge sharpening in low-dose lumbar spine CT.

MATERIALS AND METHODS: This retrospective study included 52 patients (26 male and 26 female; median age, 60.5 years) who had undergone CT-guided lumbar bone biopsy between October 2015 and April 2020. Initial 100-mAs survey images and 50-mAs intraprocedural images were reconstructed by filtered back projection. Denoising was performed using a vendor-agnostic DL model (ClariCT.AI™, ClariPI) for the 50-mAS images, and the 50-mAs, denoised 50-mAs, and 100-mAs CT images were compared. Noise, signal-to-noise ratio (SNR), and edge rise distance (ERD) for image sharpness were measured. The data were summarized as the mean ± standard deviation for these parameters. Two musculoskeletal radiologists assessed the visibility of the normal anatomical structures.

RESULTS: Noise was lower in the denoised 50-mAs images (36.38 ± 7.03 Hounsfield unit [HU]) than the 50-mAs (93.33 ± 25.36 HU) and 100-mAs (63.33 ± 16.09 HU) images (p < 0.001). The SNRs for the images in descending order were as follows: denoised 50-mAs (1.46 ± 0.54), 100-mAs (0.99 ± 0.34), and 50-mAs (0.58 ± 0.18) images (p < 0.001). The denoised 50-mAs images had better edge sharpness than the 100-mAs images at the vertebral body (ERD; 0.94 ± 0.2 mm vs. 1.05 ± 0.24 mm, p = 0.036) and the psoas (ERD; 0.42 ± 0.09 mm vs. 0.50 ± 0.12 mm, p = 0.002). The denoised 50-mAs images significantly improved the visualization of the normal anatomical structures (p < 0.001).

CONCLUSION: DL-based reconstruction may enable simultaneous noise reduction and improvement in image quality with the preservation of edge sharpness on low-dose lumbar spine CT. Investigations on further radiation dose reduction and the clinical applicability of this technique are warranted.