呼吸触发与屏气压缩感知3D MRCP在1.5T和3T的图像质量比较及个体因素影响
该回顾性研究在1.5T和3T各200例患者中比较了屏气压缩感知3D MRCP(BH-CS-MRCP)与呼吸触发标准3D MRCP(STD-MRCP)的图像质量。结果显示3T时BH-CS-MRCP整体质量、锐利度及胆胰管显示优于STD-MRCP,但1.5T时除锐利度外均劣于STD-MRCP;1.5T时腹水量与BH-CS-MRCP质量负相关,BMI与STD-MRCP质量负相关,屏气质量与BH-CS-MRCP质量相关。
要点
- 3T时BH-CS-MRCP整体质量及胆胰管显示优于STD-MRCP,可优先考虑。
- 1.5T时BH-CS-MRCP整体质量及多数胆胰管显示劣于STD-MRCP,但锐利度更佳。
- 1.5T屏气序列受腹水量影响,呼吸触发序列受BMI影响,报告需注意。
核心结论
- BH-CS-MRCP在1.5T和3T临床常规中可行,3T时图像质量显著优于STD-MRCP,1.5T时则不然。
- 1.5T时BH-CS-MRCP图像质量受腹水量影响,STD-MRCP受BMI影响;屏气质量与BH-CS-MRCP质量相关。
研究设计与样本
- 回顾性分析,1.5T和3T各纳入200例患者,均行BH-CS-MRCP和STD-MRCP序列。
- 图像质量采用5分制(整体质量及胆胰管显示)和4分制(锐利度及背景抑制)评分,50例双读评估观察者间可重复性。
- 研究个体因素包括性别、年龄、BMI、腹水、腹围和屏气质量。
关键影像表现/征象
- 1.5T时BH-CS-MRCP整体质量、背景抑制、胆囊管、左右胆管及主胰管显示均劣于STD-MRCP,但锐利度更优。
- 3T时BH-CS-MRCP整体质量、锐利度及胆胰管显示均优于STD-MRCP,但背景抑制劣于STD-MRCP。
- 1.5T时腹水量与BH-CS-MRCP图像质量呈负相关,BMI与STD-MRCP图像质量呈负相关。
- 屏气质量与BH-CS-MRCP图像质量在1.5T和3T均相关。
诊断标准与关键数值
- 整体图像质量及胆胰管显示采用5分制评分,锐利度和背景抑制采用4分制评分。
- 1.5T时BH-CS-MRCP vs STD-MRCP:整体质量p=0.0046,背景抑制p<0.0001,胆囊管p<0.0001,右胆管p=0.0008,左胆管p=0.0152,主胰管p<0.0001,锐利度p=0.028。
- 3T时BH-CS-MRCP vs STD-MRCP:整体质量p<0.0001,锐利度p<0.0001,胆胰管显示p<0.0001,背景抑制p<0.0001。
临床应用与阅片要点
- 3T可优先选用BH-CS-MRCP以缩短检查时间并获更优整体及胆胰管图像。
- 1.5T若需显示胆囊管、左右胆管及主胰管细节,STD-MRCP可能更可靠。
- 1.5T使用BH-CS-MRCP时需关注腹水量对图像质量的影响,大量腹水可能降低质量。
- 1.5T使用STD-MRCP时需考虑BMI对图像质量的影响,高BMI可能降低质量。
- 屏气质量直接影响BH-CS-MRCP图像质量,检查前需充分训练患者屏气。
局限性
- 回顾性分析设计。
- 仅50例进行双读评估观察者间可重复性。
- 原文未报告观察者间可重复性具体结果。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To evaluate the image quality of an accelerated compressed-sensing single-breath-hold 3D magnetic resonance cholangiopancreatography (BH-CS-MRCP) prototype sequence compared to the standard 3D sequence with respiratory triggering (STD-MRCP) at 1.5 T and 3 T. To assess the individual factors that can affect image quality.
METHOD: This is a retrospective analysis. Both sequences (BH-CS-MRCP and STD-MRCP) were performed in 200 patients at 1.5 T and 200 patients at 3 T. Overall image quality and the visualization of the bilio-pancreatic ducts were rated on a 5-point scale. Image sharpness and background suppression were rated on a 4-point scale. A double reading was performed in 50 patients to assess the inter-observer reproducibility. Individual characteristics studied were gender, age, BMI, ascites, abdominal surface and breath-hold quality.
RESULTS: At 1.5 T, BH-CS-MRCP was inferior to STD-MRCP in terms of overall quality (p = 0.0046), background suppression (p < 0.0001), visualization of the cystic duct (p < 0.0001), the right bile duct (p = 0.0008), the left bile duct (p = 0.0152), and the main pancreatic duct (p < 0.0001). However, BH-CS-MRCP was sharper than STD-MRCP (p = 0.028). At 3 T, BH-CS-MRCP was superior to STD-MRCP for overall quality (p < 0.0001), sharpness (p < 0.0001), and visualization of the bilio-pancreatic ducts (p < 0.0001). Background signal was conversely better suppressed in STD-MRCP (p < 0.0001). At 1.5 T, the volume of ascites was inversely correlated with image quality for BH-CS-MRCP while BMI was inversely correlated with image quality for STD-MRCP. Breath-hold quality was correlated with image quality for BH-CS-MRCP at 1.5 T and 3 T.
CONCLUSION: BH-CS-MRCP is feasible in clinical routine at 1.5 and 3 T, yielding significantly better perceived image quality at 3 T but not at 1.5 T. BH-CS-MRCP appears to be influenced by ascites whereas STD-MRCP is influenced by BMI at 1.5 T. This study was approved by the Ethics Review Board for Research in Medical Imaging (IRB: CRM-2003-065).
引用
- Blaise H, Remen T, Ambarki K, et al. Comparison of respiratory-triggered 3D MR cholangiopancreatography and breath-hold compressed-sensing 3D MR cholangiopancreatography at 1.5 T and 3 T and impact of individual factors on image quality. Eur J Radiol. 2021. PMID: 34371309. doi:10.1016/j.ejrad.2021.109873