临床光子计数探测器CT增强腹部CT:图像质量评估及与能量积分探测器CT的比较

Contrast-Enhanced Abdominal CT with Clinical Photon-Counting Detector CT: Assessment of Image Quality and Comparison with Energy-Integrating Detector CT.
PubMed 文献 · Acad Radiol · 2021-08-11

回顾性纳入39例患者,对比增强腹部门静脉期光子计数探测器CT(PCD-CT)与既往能量积分探测器CT(EID-CT)图像质量。PCD-CT重建40-90 keV虚拟单能图像(VMI),40和50 keV VMI的实质和血管CNR显著高于EID-CT;50 keV VMI主观图像质量与EID-CT相当,且辐射剂量相同。

要点

  • 50 keV VMI在相同辐射剂量下CNR显著高于EID-CT,主观图像质量与EID-CT相当。
  • 40-60 keV VMI主观对比度显著高于EID-CT,但40-50 keV主观噪声也显著更高。
  • 病灶显示清晰度在各重建图像间评分相似。

核心结论

  • 个体内分析显示,腹部PCD-CT的50 keV VMI在相同辐射剂量下CNR显著高于EID-CT,主观图像质量与EID-CT相似。

研究设计与样本

  • 回顾性个体内研究,纳入39例患者(平均年龄63±10岁,女性10例,平均BMI 26.0±5.7 kg/m2),均行临床指征的增强腹部门静脉期CT,先后使用第一代双源PCD-CT和EID-CT,对比剂方案相同;PCD-CT采用QuantumPlus模式(获取全光谱信息)120 kVp,EID-CT采用自动管电压选择(参考管电压100 kVp);PCD-CT重建40-90 keV、间隔10 keV的VMI;每位患者调整PCD-CT管电流-时间乘积以使CTDIvol与EID-CT相同;两名独立盲法放射科医师采用5点Likert量表评估主观图像质量。

诊断标准与关键数值

  • PCD-CT的40和50 keV VMI的实质CNR和血管CNR均显著高于EID-CT(均p<0.05)。
  • 50和60 keV VMI的总体图像质量与EID-CT评分相似(均p>0.05)。
  • 40-50 keV VMI的主观噪声显著高于EID-CT(p<0.05)。
  • 40-60 keV VMI的主观对比度显著高于EID-CT(p<0.05)。
  • 各图像上病灶显示清晰度评分相似。

临床应用与阅片要点

  • 在相同辐射剂量下,50 keV VMI可提供比EID-CT更高的CNR,同时保持相似的主观图像质量。
  • 40-60 keV VMI主观对比度更高,但40-50 keV主观噪声也更高。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVES: To determine quantitative and qualitative image quality of contrast-enhanced abdominal photon-counting detector CT (PCD-CT) compared to energy-integrating detector CT (EID-CT) in the same patients.

MATERIAL AND METHODS: Thirty-nine patients (mean age 63 ± 10 years, 10 females, mean BMI 26.0 ± 5.7 kg/m2) were retrospectively included who underwent clinically indicated, contrast-enhanced abdominal CT in portal-venous phase with first-generation dual-source PCD-CT and who underwent previous abdominal CT with EID-CT. For both scan, same contrast media protocol was used. PCD-CT was performed in QuantumPlus mode (obtaining full spectral information) at 120kVp. EID-CT was performed using automated tube voltage selection (reference tube voltage 100kVp). In PCD-CT, virtual monoenergetic images (VMI) were reconstructed in 10keV intervals (40-90 keV). Tube current-time product in PCD-CT was modified in each patient to obtain same volume CT-dose-index (CTDIvol) as with EID-CT. Attenuation of organs and vascular structures were measured, noise quantified, and contrast-to-noise ratio (CNR) calculated. Two independent, blinded radiologists assessed subjective image quality using a 5-point Likert scale (overall image quality, image noise, contrast, and liver lesion conspicuity).

RESULTS: Median time interval between the scan was 12 months. BMI (p = 0.905) and CTDIvol (p = 0.984) were similar between scans. CNRparenchymal and CNRvascular of VMI from PCD-CT at 40 and 50keV were significantly higher than EID-CT (all, p < 0.05). Overall, inter-reader agreement for all subjective image quality readings was substantial (Krippendorff's alpha = 0.773). Overall image quality of VMI was rated similar at 50 and 60 keV compared to EID-CT (all, p > 0.05). Subjective image noise was significantly higher at 40-50 keV, contrast significantly higher at 40-60 keV (all, p < 0.05). Lesion conspicuity was rated similar on all images.

CONCLUSION: Our intra-individual analysis of abdominal PCD-CT indicates that VMI at 50 keV shows significantly higher CNR at similar subjective image quality as compared to EID-CT at identical radiation dose.

来源:Acad Radiol,查看原文。版权归原作者所有。