使用kVp无关重建算法在超高螺距Sn150 kVp CT上准确测量Agatston评分
前瞻性研究纳入114例患者,比较标准120 kVp与超高螺距Sn150 kVp扫描及不同重建算法获得的Agatston评分。Sn150 kVp联合kVp无关重建算法(协议C)与标准120 kVp(协议A)评分无显著差异(p=0.060),ICC=0.997;而Sn150 kVp联合标准重建算法(协议B)显著低估评分(p<0.001)。UHP Sn150 kVp模式有效辐射剂量降低约30%(0.58 vs. 0.82 mSv,p<0.001)。
要点
- 协议C(Sn150 kVp + Sa36d)与协议A(120 kVp + Qr36d)的Agatston评分无显著差异(p=0.060),ICC=0.997。
- 协议B(Sn150 kVp + Qr36d)较协议A显著低估Agatston评分(p<0.001),不宜用于钙化评分。
- UHP Sn150 kVp模式较标准120 kVp模式有效辐射剂量降低约30%(0.58 vs. 0.82 mSv,p<0.001)。
核心结论
- 采用kVp无关重建算法的UHP Sn150 kVp模式与标准120 kVp模式的Agatston评分高度一致(平均差异小、一致性界限窄),且UHP Sn150 kVp模式可显著降低辐射剂量。
研究设计与样本
- 前瞻性研究,纳入114例患者(平均年龄60.3±9.8岁;74例男性),均行标准120 kVp扫描和UHP Sn150 kVp扫描用于冠状动脉钙化评分(CACS)。
诊断标准与关键数值
- Agatston评分:协议A中位数63.05(IQR 0-232.28)vs 协议C中位数60.25(IQR 0-195.20),p=0.060;协议A与C平均差异-7.82,一致性界限-65.20至49.56,ICC=0.997;协议B中位数34.85(IQR 0-120.73)较协议A显著低估,p<0.001。
临床应用与阅片要点
- 使用UHP Sn150 kVp模式行CACS时,应采用kVp无关重建算法(Sa36d)以避免Agatston评分低估。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To investigate the accuracy of the Agatston score obtained with the ultra-high-pitch (UHP) acquisition mode using tin-filter spectral shaping (Sn150 kVp) and a kVp-independent reconstruction algorithm to reduce the radiation dose.
MATERIALS AND METHODS: This prospective study included 114 patients (mean ± standard deviation, 60.3 ± 9.8 years; 74 male) who underwent a standard 120 kVp scan and an additional UHP Sn150 kVp scan for coronary artery calcification scoring (CACS). These two datasets were reconstructed using a standard reconstruction algorithm (120 kVp + Qr36d, protocol A; Sn150 kVp + Qr36d, protocol B). In addition, the Sn150 kVp dataset was reconstructed using a kVp-independent reconstruction algorithm (Sn150 kVp + Sa36d, protocol C). The Agatston scores for protocols A and B, as well as protocols A and C, were compared. The agreement between the scores was assessed using the intraclass correlation coefficient (ICC) and the Bland-Altman plot. The radiation doses for the 120 kVp and UHP Sn150 kVp acquisition modes were also compared.
RESULTS: No significant difference was observed in the Agatston score for protocols A (median, 63.05; interquartile range [IQR], 0-232.28) and C (median, 60.25; IQR, 0-195.20) (p = 0.060). The mean difference in the Agatston score for protocols A and C was relatively small (-7.82) and with the limits of agreement from -65.20 to 49.56 (ICC = 0.997). The Agatston score for protocol B (median, 34.85; IQR, 0-120.73) was significantly underestimated compared with that for protocol A (p < 0.001). The UHP Sn150 kVp mode facilitated an effective radiation dose reduction by approximately 30% (0.58 vs. 0.82 mSv, p < 0.001) from that associated with the standard 120 kVp mode.
CONCLUSION: The Agatston scores for CACS with the UHP Sn150 kVp mode with a kVp-independent reconstruction algorithm and the standard 120 kVp demonstrated excellent agreement with a small mean difference and narrow agreement limits. The UHP Sn150 kVp mode allowed a significant reduction in the radiation dose.