文献速递心血管CT

人工智能全自动冠状动脉钙化积分软件:基于三个CT队列的验证研究

Fully Automatic Coronary Calcium Score Software Empowered by Artificial Intelligence Technology: Validation Study Using Three CT Cohorts.
PubMed 文献 · Korean J Radiol · 2021-07-26

本研究利用三个已发表的CT队列(共2985例,涵盖无症状、有症状及瓣膜病患者,四种CT机型)验证深度学习全自动冠状动脉钙化积分系统(CAC_auto),以手工分割积分(CAC_hand)为参考标准。结果显示CAC_auto检测钙化病灶的敏感度为93.3%,Agatston积分ICC达0.99,风险分类加权kappa为0.94,假阳性主要源于图像噪声、主动脉壁钙化和心包钙化。

要点

  • 全自动钙化积分与手工测量一致性极高,Agatston积分ICC为0.99,风险分类kappa为0.94。
  • 病灶水平敏感度93.3%,假阳性率0.11个/患者,主要原因为噪声、主动脉壁及心包钙化。
  • 该系统有望简化冠状动脉钙化成像流程,但需注意上述假阳性来源。

关键影像表现/征象

  • 手工分割共识别6218个冠状动脉钙化病灶。
  • 全自动系统检测钙化的病灶水平敏感度为93.3%(5800/6218),假阳性率为每患者0.11个病灶。
  • 假阳性主要原因为图像噪声(29.1%,97/333病灶)、主动脉壁钙化(25.5%,85/333病灶)和心包钙化(24.3%,81/333病灶)。

诊断标准与关键数值

  • 全自动与手工Agatston积分在所有血管的ICC为0.99(左主干0.91,左前降支0.99,左回旋支0.96,右冠状动脉0.99)。
  • 全自动与手工Agatston积分的Bland-Altman一致性界限为1.6 ± 52.2。
  • 基于Agatston积分分类(0、1-10、11-100、101-400、>400)的线性加权kappa值为0.94。

临床应用与阅片要点

  • 全自动钙化积分可用于日常冠状动脉钙化评估,但需警惕图像噪声、主动脉壁钙化和心包钙化导致的假阳性。
  • 在报告自动积分结果时,建议对上述假阳性来源区域进行复核,确保钙化定位准确。

局限性

  • 验证队列来自已发表数据,可能存在选择偏倚。
  • 假阳性原因中图像噪声占比较高,提示图像质量对自动积分的影响。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: This study aimed to validate a deep learning-based fully automatic calcium scoring (coronary artery calcium [CAC]_auto) system using previously published cardiac computed tomography (CT) cohort data with the manually segmented coronary calcium scoring (CAC_hand) system as the reference standard.

MATERIALS AND METHODS: We developed the CAC_auto system using 100 co-registered, non-enhanced and contrast-enhanced CT scans. For the validation of the CAC_auto system, three previously published CT cohorts (n = 2985) were chosen to represent different clinical scenarios (i.e., 2647 asymptomatic, 220 symptomatic, 118 valve disease) and four CT models. The performance of the CAC_auto system in detecting coronary calcium was determined. The reliability of the system in measuring the Agatston score as compared with CAC_hand was also evaluated per vessel and per patient using intraclass correlation coefficients (ICCs) and Bland-Altman analysis. The agreement between CAC_auto and CAC_hand based on the cardiovascular risk stratification categories (Agatston score: 0, 1-10, 11-100, 101-400, > 400) was evaluated.

RESULTS: In 2985 patients, 6218 coronary calcium lesions were identified using CAC_hand. The per-lesion sensitivity and false-positive rate of the CAC_auto system in detecting coronary calcium were 93.3% (5800 of 6218) and 0.11 false-positive lesions per patient, respectively. The CAC_auto system, in measuring the Agatston score, yielded ICCs of 0.99 for all the vessels (left main 0.91, left anterior descending 0.99, left circumflex 0.96, right coronary 0.99). The limits of agreement between CAC_auto and CAC_hand were 1.6 ± 52.2. The linearly weighted kappa value for the Agatston score categorization was 0.94. The main causes of false-positive results were image noise (29.1%, 97/333 lesions), aortic wall calcification (25.5%, 85/333 lesions), and pericardial calcification (24.3%, 81/333 lesions).

CONCLUSION: The atlas-based CAC_auto empowered by deep learning provided accurate calcium score measurement as compared with manual method and risk category classification, which could potentially streamline CAC imaging workflows.

来源:Korean J Radiol,查看原文。版权归原作者所有。

引用

  1. Lee JG, Kim H, Kang H, et al. Fully Automatic Coronary Calcium Score Software Empowered by Artificial Intelligence Technology: Validation Study Using Three CT Cohorts. Korean J Radiol. 2021. PMID: 34402248. doi:10.3348/kjr.2021.0148

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。