商业二维斑点追踪超声心动图节段应变值的质量控制:基于计算机仿真模型

Quality Assurance of Segmental Strain Values Provided by Commercial 2-D Speckle Tracking Echocardiography Using in Silico Models: A Report from the EACVI-ASE Strain Standardization Task Force.
PubMed 文献 · Ultrasound Med Biol · 2021-08-12

本研究利用计算机仿真生成的左心室二维超声灰阶图像,评估不同厂商节段应变值的准确性与可重复性。五种模型中有四种模拟不同血管区域透壁梗死导致的收缩期节段拉伸。结果显示,节段应变分析识别节段拉伸的AUC为0.80至0.96,但单个节段峰值应变的绝对误差在不同厂商间差异显著(3.3±3.1%至11.2±5.9%)。结论是节段应变分析可半定量评估区域功能障碍,但单个节段峰值应变值需谨慎解读。

要点

  • 节段应变分析识别节段拉伸的AUC为0.80至0.96,准确性良好。
  • 单个节段峰值应变的绝对误差在不同厂商间差异显著,范围3.3±3.1%至11.2±5.9%。
  • 建议将节段应变作为半定量工具评估区域功能障碍,避免过度依赖单个节段峰值。

核心结论

  • 节段应变分析能较准确地识别节段拉伸区域,但单个节段峰值应变值不准确,应谨慎解读。
  • 半定量应变评估可用于检测区域性功能障碍。

研究设计与样本

  • 研究使用计算机仿真生成的左心室二维超声灰阶图像,基于已知节段纵向应变值的电机械左心室模型。
  • 五种模型中四种模拟不同血管区域的透壁梗死伴收缩期节段拉伸,在三个心尖切面以四种噪声水平合成动态图像。
  • 所有仿真图像由一名研究者重复分析,部分由另一名研究者分析。

诊断标准与关键数值

  • 识别节段拉伸的AUC范围为0.80(95%CI 0.77-0.83)至0.96(95%CI 0.95-0.98)。
  • 不同厂商的绝对误差范围为3.3±3.1%至11.2±5.9%。
  • 观察者内变异的一致性界限范围为-3.0%至2.9%和-5.2%至4.8%。
  • 观察者间变异的一致性界限范围为-3.6%至3.5%和-14.5%至8.5%。

临床应用与阅片要点

  • 在日常阅片中,可利用节段应变半定量识别区域功能障碍,但报告单个节段峰值应变时应注明可能不准确。
  • 注意不同厂商设备间节段应变值存在显著差异,跨厂商比较需谨慎。

局限性

  • 研究基于计算机仿真模型,非真实患者数据,结果外推需谨慎。
  • 部分图像仅由单一研究者分析,观察者间变异评估不完全。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

The aim of this study was to determine the accuracy and reproducibility of vendor-specific regional strain values by echocardiography using in silico data. Synthetic 2-D ultrasound gray-scale images of the left ventricle (LV) were generated with knowledge of the longitudinal segmental strain values from the underlying electromechanical LV model. Four of five models mimicked transmural infarctions with systolic segmental stretching in different vascular areas. Cine loops in the three apical views were synthetically generated at four noise levels. All in silico images were repeatedly analyzed by a single investigator and some by another investigator. The absolute errors varied significantly between vendors from 3.3 ± 3.1% to 11.2 ± 5.9%. The area under the curve for the identification of segmental stretching ranged from 0.80 (confidence interval: 0.77-0.83) to 0.96 (0.95-0.98). The levels of agreement for intra-investigator variability varied between -3.0% to 2.9% and -5.2% to 4.8%, and for inter-investigator variability, between -3.6% to 3.5% and -14.5% to 8.5%. Segmental strain analysis allows the identification of areas with segmental stretching with good accuracy. However, single segmental peak-strain values are not accurate and should be interpreted with caution. Nevertheless, our results indicate the usefulness of semiquantitative strain assessment for the detection of regional dysfunction.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。

引用

  1. Røsner A, Alessandrini M, Kjønås D, et al. Quality Assurance of Segmental Strain Values Provided by Commercial 2-D Speckle Tracking Echocardiography Using in Silico Models: A Report from the EACVI-ASE Strain Standardization Task Force. Ultrasound Med Biol. 2021. PMID: 34392996. doi:10.1016/j.ultrasmedbio.2021.07.011