双能CT肺灌注新型重建技术评估:58例患者初步经验
研究纳入58例患者(28例肺栓塞、30例无肺栓塞),比较虚拟单能图像减影生成的肺灌注血容量(Lung Mono)与三物质分解生成的肺灌注血容量(Lung PBV,参考标准)。Lung Mono在图像噪声、灌注缺损检出和叶间裂显示方面均优于Lung PBV。
要点
- Lung Mono图像主观噪声轻度者占58.6%(34/58),Lung PBV为25.9%(15/58),差异有统计学意义(p<0.0001)。
- 28例肺栓塞患者中,灌注区与缺损区衰减梯度Lung Mono中位数73.5 HU(QI=65.0;Q3=86.0),Lung PBV为24.5 HU(22.0;30.0),p<0.0001。
- 叶间裂在Lung Mono图像上77.6%(45/58)可精确识别,Lung PBV上模糊者51.7%(30/58)、不可见者48.3%(28/58)。
核心结论
- Lung Mono灌注成像可显著改善整体图像质量,并提高肺栓塞型灌注缺损的检出能力。
研究设计与样本
- 纳入58例患者(28例伴肺栓塞、30例不伴肺栓塞),重建Lung PBV图像(即参考标准)和Lung Mono图像,在患者水平和肺段水平进行技术间比较。
关键影像表现/征象
- 主观图像噪声评分分布在两种重建间显著不同(p<0.0001),Lung Mono轻度噪声占58.6%(34/58),Lung PBV占25.9%(15/58)。
- 肺段水平灌注缺损检出:1104个肺段一致(无缺损n=968;有缺损n=138),2个肺段不一致,仅Lung Mono显示肺栓塞相关缺损。
- 28例肺栓塞患者中,灌注区与缺损区衰减梯度Lung Mono中位数73.5 HU(QI=65.0;Q3=86.0),Lung PBV为24.5 HU(22.0;30.0),Lung Mono显著更高(p<0.0001)。
- 所有患者中,叶间裂在Lung Mono图像上77.6%(45/58)可精确识别,Lung PBV上模糊者51.7%(30/58)、不可见者48.3%(28/58)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To compare dual-energy (DE) lung perfused blood volume generated by subtraction of virtual monoenergetic images (Lung Mono) with images obtained by three-compartment decomposition (Lung PBV).
MATERIAL AND METHODS: The study included 58 patients (28 patients with and 30 patients without PE) with reconstruction of Lung PBV images (i.e., the reference standard) and Lung Mono images. The inter-technique comparison was undertaken at a patient and segment level.
RESULTS: The distribution of scores of subjective image noise (patient level) significantly differed between the two reconstructions (p<0.0001), with mild noise in 58.6% (34/58) of Lung Mono images vs 25.9% (15/58) of Lung PBV images. Detection of perfusion defects (segment level) was concordant in 1104 segments (no defect: n=968; defects present: n=138) and discordant in 2 segments with a PE-related defect only depicted on Lung Mono images. Among the 28 PE patients, the distribution of gradient of attenuation between perfused areas and defects was significantly higher on Lung Mono images compared to Lung PBV (median= 73.5 HU (QI=65.0; Q3=86.0) vs 24.5 HU (22.0; 30.0); p<0.0001). In all patients, fissures were precisely identified in 77.6% of patients (45/58) on Lung Mono images while blurred (30/58; 51.7%) or not detectable (28/58; 48.3%) on Lung PBV images.
CONCLUSION: Lung Mono perfusion imaging allows significant improvement in the overall image quality and improved detectability of PE-type perfusion defects.
引用
- Pinilo J, Hutt A, Labreuche J, et al. Evaluation Of a New Reconstruction Technique for Dual-Energy (DECT) Lung Perfusion: Preliminary Experience In 58 Patients. Acad Radiol. 2021. PMID: 34446359. doi:10.1016/j.acra.2021.07.023
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。