文献速递头颈技术与安全MR

对比增强T1加权头颈部MRI:螺旋GRE、笛卡尔GRE与笛卡尔TSE序列的前瞻性个体内图像质量比较

Contrast-Enhanced T1-Weighted Head and Neck MRI: Prospective Intraindividual Image Quality Comparison of Spiral GRE, Cartesian GRE, and Cartesian TSE Sequences.
PubMed 文献 · AJR Am J Roentgenol · 2021-08-18

三位放射科医生独立评估显示,螺旋GRE在整体图像质量、搏动伪影、肌肉和血管轮廓、运动伪影及黏膜与咽肌区分方面显著优于笛卡尔序列,且对比噪声比更高、扫描时间更短,但牙科材料伪影更大。

要点

  • 螺旋GRE序列扫描时间更短且对比噪声比更高,整体图像质量优于笛卡尔序列。
  • 螺旋GRE在减少搏动伪影、运动伪影及改善肌肉和血管轮廓方面有优势。
  • 螺旋GRE的牙科材料伪影范围更大,是其主要劣势。

核心结论

  • 三维螺旋GRE可改善头颈部MRI的主观图像质量和对比噪声比,且扫描时间短于笛卡尔序列,但牙科材料伪影更大。

研究设计与样本

  • 所有患者均采集对比增强螺旋GRE(2分28秒)、笛卡尔GRE(4分27秒)和笛卡尔TSE(3分41秒)序列,采集顺序在患者间轮换。
  • 三位放射科医生独立使用5点Likert量表评估图像质量指标,包括预设病灶的显示清晰度;一位读者测量牙科材料伪影最大范围及对比噪声比。

关键影像表现/征象

  • 读者间对图像质量指标的一致性为中等至优秀(Krippendorff alpha,0.681-1.000)。
  • 螺旋GRE在整体图像质量(整个头颈部、仅颈部、仅头部)、搏动伪影、肌肉轮廓描绘、血管轮廓描绘、运动伪影及黏膜与咽肌区分方面的评分显著优于笛卡尔GRE和笛卡尔TSE(所有读者p < .05)。
  • 螺旋GRE与笛卡尔序列在病灶显示清晰度(整个头颈部、仅颈部、仅头部)、脂肪抑制质量、血流伪影和折叠伪影方面的评分无显著差异(所有读者p > .05)。
  • 螺旋GRE的牙科材料伪影评分显著差于其他序列(所有读者p < .05)。

诊断标准与关键数值

  • 牙科材料伪影平均最大范围:螺旋GRE 39.6 ± 25.5 mm,笛卡尔GRE 35.6 ± 24.3 mm,笛卡尔TSE 29.6 ± 21.4 mm。
  • 平均对比噪声比:螺旋GRE 221.1 ± 94.5,笛卡尔GRE 151.8 ± 85.7,笛卡尔TSE 153.0 ± 63.2(螺旋GRE与其他序列间p < .001)。

临床应用与阅片要点

  • 螺旋GRE序列可缩短扫描时间并提高整体图像质量,适合作为对比增强T1加权头颈部MRI的优选序列。
  • 对于有牙科材料的患者,需注意螺旋GRE可能产生更大的伪影,必要时可结合笛卡尔序列以减少伪影干扰。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND. Sequences with noncartesian k-space sampling may improve image quality of head and neck MRI. OBJECTIVE. The purpose of this study was to compare intraindividually the image quality of a spiral gradient-recalled echo (GRE) sequence and conventional cartesian GRE and cartesian turbo spin-echo (TSE) sequences for contrast-enhanced T1-weighted head and neck MRI. METHODS. This prospective study included patients referred for contrast-enhanced head and neck MRI from August 2020 to May 2021. Patients underwent 1.5-T MRI including contrast-enhanced spiral GRE (2 minutes 28 seconds), cartesian GRE (4 minutes 27 seconds), and cartesian TSE (3 minutes 41 seconds) sequences, acquired in rotating order across patients. Three radiologists independently assessed image quality measures, including conspicuity of prespecified lesions, using 5-point Likert scales. One reader measured maximal extent of dental material artifact and contrast-to-noise ratio (CNR). RESULTS. Thirty-one patients (13 men, 18 women; mean age, 63.8 years) were enrolled. Nineteen patients had a focal lesion; 22 had dental material. Interreader agreement for image quality measures was substantial to excellent (Krippendorff alpha, 0.681-1.000). Scores for overall image quality (whole head and neck, neck only, and head only), pulsation artifact, muscular contour delineation, vessel contour delineation, motion artifact, and differentiation between mucosa and pharyngeal muscles were significantly better for spiral GRE than for cartesian GRE and cartesian TSE for all readers (p < .05). Scores for lesion conspicuity (whole head and neck, neck only, and head only), quality of fat suppression, flow artifact, and foldover artifact were not significantly different between spiral GRE and the cartesian sequences for any reader (p > .05). Dental material artifact scores were significantly worse for spiral GRE than the other sequences for all readers (p < .05). The mean maximum extent of dental material artifact was 39.6 ± 25.5 (SD) mm for spiral GRE, 35.6 ± 24.3 mm for cartesian GRE, and 29.6 ± 21.4 mm for cartesian TSE; the mean CNR was 221.1 ± 94.5 for spiral GRE, 151.8 ± 85.7 for cartesian GRE, and 153.0 ± 63.2 for cartesian TSE (p < .001 between spiral GRE and other sequences for both measures). CONCLUSION. Three-dimensional spiral GRE improves subjective image quality and CNR of head and neck MRI with shorter scan time versus cartesian sequences, though it exhibits larger dental material artifact. CLINICAL IMPACT. A spiral sequence may help overcome certain challenges of conventional cartesian sequences for head and neck MRI.

来源:AJR Am J Roentgenol,查看原文。版权归原作者所有。

引用

  1. Sartoretti E, Sartoretti-Schefer S, van Smoorenburg L, et al. Contrast-Enhanced T1-Weighted Head and Neck MRI: Prospective Intraindividual Image Quality Comparison of Spiral GRE, Cartesian GRE, and Cartesian TSE Sequences. AJR Am J Roentgenol. 2021. PMID: 34406050. doi:10.2214/AJR.21.26413

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。