指南与共识儿科影像技术与安全CT

儿科图像引导放疗中的低剂量锥形束CT:图像质量与实践建议

Low dose cone beam CT for paediatric image-guided radiotherapy: Image quality and practical recommendations.
指南 / 共识 · Radiother Oncol · 2021-07-31

该指南基于7例6-13岁患儿共192次锥形束CT(0.5–12.8 mGy)分析噪声来源,发现解剖噪声主导图像质量,量子噪声贡献<20%;剂量≥1 mGy时90%以上病例CNR>3且软组织配准准确,剂量>1 mGy无额外获益。

要点

  • 儿科CBCT中解剖噪声(如运动伪影)是图像质量的主要决定因素,量子噪声贡献<20%。
  • 剂量≥1 mGy时,超过90%的病例CNR>3,且软组织配准准确;0.5 mGy时仅57%病例CNR>3。
  • 剂量>1 mGy对改善图像质量或配准精度无额外获益,建议采用低剂量协议。

核心结论

  • 在儿科CBCT中,解剖噪声主导量子噪声;剂量低至1 mGy即可获得适当的软组织对比度和配准精度;由于解剖噪声的存在,剂量>1 mGy对改善图像质量或配准精度无获益。

研究设计与样本

  • 指南类文件,基于7例6-13岁患儿的60次胸部或腹部/盆腔CBCT(模拟低剂量扫描共192次,剂量范围0.5–12.8 mGy),并与10岁体模对比,建模总噪声和量子噪声随剂量的变化。

关键影像表现/征象

  • 在所有病例中,量子噪声对总噪声的贡献<20%,提示解剖噪声是腹部/盆腔区域图像质量的最大决定因素。
  • 在剂量≥1 mGy时,超过90%的病例CNR>3;在0.5 mGy时,57%的病例CNR>3。
  • 剂量>1 mGy时软组织配准准确。

诊断标准与关键数值

  • CNR在脂肪/肌肉之间测量;剂量≥1 mGy时CNR>3的病例超过90%,0.5 mGy时CNR>3的病例为57%。
  • 软组织配准在肾脏上进行,剂量>1 mGy时配准准确。

临床应用与阅片要点

  • 推荐儿科CBCT成像协议采用低至1 mGy的剂量,以实现适当的软组织对比度和配准精度。
  • 剂量>1 mGy不改善图像质量或配准精度,无需增加剂量。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: Cone beam CT (CBCT) is used in paediatric image-guided radiotherapy (IGRT) for patient setup and internal anatomy assessment. Adult CBCT protocols lead to excessive doses in children, increasing the risk of radiation-induced malignancies. Reducing imaging dose increases quantum noise, degrading image quality. Patient CBCTs also include 'anatomical noise' (e.g. motion artefacts), further degrading quality. We determine noise contributions in paediatric CBCT, recommending practical imaging protocols and thresholds above which increasing dose yields no improvement in image quality.

METHODS AND MATERIALS: Sixty CBCTs including the thorax or abdomen/pelvis from 7 paediatric patients (aged 6-13 years) were acquired at a range of doses and used to simulate lower dose scans, totalling 192 scans (0.5-12.8 mGy). Noise measured in corresponding regions of each patient and a 10-year-old phantom were compared, modelling total (including anatomical) noise, and quantum noise contributions as a function of dose. Contrast-to-noise ratio (CNR) was measured between fat/muscle. Soft tissue registration was performed on the kidneys, comparing accuracy to the highest dose scans.

RESULTS: Quantum noise contributed <20% to total noise in all cases, suggesting anatomical noise is the largest determinant of image quality in the abdominal/pelvic region. CNR exceeded 3 in over 90% of cases ≥ 1 mGy, and 57% of cases at 0.5 mGy. Soft tissue registration was accurate for doses > 1 mGy.

CONCLUSION: Anatomical noise dominates quantum noise in paediatric CBCT. Appropriate soft tissue contrast and registration accuracy can be achieved for doses as low as 1 mGy. Increasing dose above 1 mGy holds no benefit in improving image quality or registration accuracy due to the presence of anatomical noise.

来源:Radiother Oncol,查看原文。版权归原作者所有。

引用

  1. Bryce-Atkinson A, De Jong R, Marchant T, et al. Low dose cone beam CT for paediatric image-guided radiotherapy: Image quality and practical recommendations. Radiother Oncol. 2021. PMID: 34343544. doi:10.1016/j.radonc.2021.07.027

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。