体环姿态相关MRI射频致热风险:观察、表征与建议
基于12个可摆位成人男性人体模型,在1.5T MR系统中数值模拟11个解剖标志点的射频致热,并用定制体模实验验证。体环姿态致热可达原始姿态的11倍,增大体环尺寸和减小接触面积使致热增加;在体环中留空气隙或使偏振磁场平行于体环至少降低致热3倍。建议使用绝缘材料防止环路形成,无法避免时将体环置于射频发射线圈外。
要点
- 体环姿态可致射频致热较原始姿态高达11倍,阅片时需关注患者摆位。
- 体环越大、接触面积越小,射频致热越严重。
- 体环中心靠近射频线圈等中心时磁通量增加,致热增加。
关键影像表现/征象
- 体环姿态所致射频致热可达原始姿态的11倍。
- 射频致热随体环尺寸增大和接触面积减小而增加。
- 体环中心与射频线圈等中心靠近时磁通量增加,导致射频致热增加。
- 体环中留空气隙或产生平行于体环的偏振磁场至少降低致热3倍。
临床应用与阅片要点
- 摆位时使用绝缘材料防止体环形成,以规避高射频致热。
- 若体环无法避免,应将体环置于射频发射线圈外。
- 可采用磁场平行于体环的线偏振以规避高射频致热。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To assess RF-induced heating hazards in 1.5T MR systems caused by body-loop postures.
METHODS: Twelve advanced high-resolution anatomically correct human body models with different body-loop postures are created based on poseable human adult male models. Numerical simulations are performed to assess the radiofrequency (RF)-induced heating of these 12 models at 11 landmarks. A customized phantom is developed to validate the numerical simulations and quantitatively analyze factors affecting the RF-induced heating, eg, the contact area, the loop size, and the loading position. The RF-induced heating inside three differently posed phantoms is measured.
RESULTS: The RF-induced heating from the body-loop postures can be up to 11 times higher than that from the original posture. The RF-induced heating increases with increasing body-loop size and decreasing contact area. The magnetic flux increases when the body-loop center and the RF coil isocenter are close to each other, leading to increased RF-induced heating. An air gap created in the body loop or generating a polarized magnetic field parallel to the body loop can reduce the heating by a factor of three at least. Experimental measurements are provided, validating the correctness of the numerical results.
CONCLUSION: Safe patient posture during MR examinations is recommended with the use of insulation materials to prevent loop formation and consequently avoiding high RF-induced heating. If body loops cannot be avoided, the body loop should be placed outside the RF transmitting coil. In addition, linear polarization with magnetic fields parallel to the body loop can be used to circumvent high RF-induced heating.
引用
- Yang X, Zheng J, Wang Y, et al. Body-loop related MRI radiofrequency-induced heating hazards: Observations, characterizations, and recommendations. Magn Reson Med. 2021. PMID: 34355817. doi:10.1002/mrm.28954
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。