这项前瞻性研究纳入105个膝关节(66例患儿),评估腔内凸阵探头(ICAP)显示外侧半月板及诊断儿童盘状半月板(DLM)的价值。ICAP显示半月板内缘显著优于线阵探头(p<0.01),<8岁患儿显示更佳(p<0.001);ICAP与MRI评估半月板形态高度一致(κ=0.849)。ICAP超声是儿童DLM无创、便捷、低成本的诊断或筛查工具,尤其适用于<8岁儿童。
要点
- ICAP显示外侧半月板内缘显著优于线阵探头,<8岁儿童显示效果更佳。
- ICAP与MRI评估半月板形态高度一致(κ=0.849),可作为儿童DLM筛查工具。
- 半月板越宽,超声与MRI测量偏差越大,两者呈正相关(r=0.612)。
局限性
- 原文未明确报告研究局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
This prospective study aimed to assess the usefulness of an intracavitary convex array probe (ICAP) in visualizing the lateral meniscus (LM) and improving the diagnostic utility of ultrasound (US) when diagnosing or screening for discoid lateral meniscus (DLM) in children. We included 105 knees (66 patients) that had symptomatic or asymptomatic DLM. We extracted and retrospectively reviewed data regarding patient demographics, medical records, magnetic resonance imaging (MRI), ultrasonographic features and arthroscopic findings. The inner edge of the LM visualized using an ICAP was significantly clearer than that visualized using a linear array probe, and the difference was significant (p < 0.01). The edges were better visualized in patients aged <8 y than in those aged >8 y, and the difference was significant (p < 0.001). The average widths of the LM body using an ICAP and MRI were 19.85 ± 3.63 and 24.46 ± 4.94 mm, respectively, and the wider the meniscal width, the greater was the deviation between the US and MRI measurements, which were positively correlated (r = 0.612, p < 0.001). With the use of MRI measurements and an ICAP, meniscal widths in poorly visualized LMs were greater than those in clearly visualized LMs, but this difference was not significant (p = 0.161). US scans using an ICAP and MRI were highly consistent in assessing the shape of the menisci (κ = 0.849, p < 0.001). US scan using an ICAP is a non-invasive, convenient and low-cost modality for diagnosing or screening for DLM in the pediatric population, especially in children aged <8 y.
引用
- Yang H, Li Q, Liang Z, et al. Diagnostic Value of Ultrasound in Children with Discoid Lateral Meniscus Using Either an Intracavitary Convex Array Probe or a Linear Array Probe. Ultrasound Med Biol. 2021. PMID: 34229908. doi:10.1016/j.ultrasmedbio.2021.05.024
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。